 Class 2 Device Recall System 6 Aseptic Housingsee related information
Date Initiated by FirmJanuary 07, 2015
Date PostedFebruary 09, 2015
Recall Status1 Terminated 3 on March 09, 2016
Recall NumberZ-1066-2015
Recall Event ID 70298
Product Classification Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment - Product Code HWE
ProductSystem 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.
Code Information Part No. 6126-120-000, Lots 13205, 13209, 13210, 13212
FEI Number 1811755
Recalling Firm/
Manufacturer
Stryker Instruments Div. of Stryker Corporation
4100 E Milham Ave
Portage MI 49002-9704
For Additional Information ContactKara Spath
269-289-4518
Manufacturer Reason
for Recall
Stryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.
FDA Determined
Cause 2
Under Investigation by firm
ActionStryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or kara.spath@stryker.com.
Quantity in Commerce399
DistributionWorldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.
Total Product Life CycleTPLC Device Report