|
Class 2 Device Recall Mobile Column TruSystem 7500, Material Number 1717023; Mobile Column TruSystem 5500, Material Number |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
July 10, 2016 |
Date Posted |
October 05, 2016 |
Recall Status1 |
Terminated 3 on March 02, 2017 |
Recall Number |
Z-0040-2017 |
Recall Event ID |
75115 |
Product Classification |
Table, surgical with orthopedic accessories, ac-powered - Product Code JEA
|
Product |
Mobile Column TruSystem 7500, Material Number 1717023; Mobile Column TruSystem 5500, Material Number 1607822 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recover from the anesthesia. Also, patient transport on the operating table top from a patient transfer system to the operating theater of from the operating theater to a patient transfer system |
Code Information |
Serial Numbers - 101281354, 101856082, 101858789, 101858794, 101861232, 101871990, 101871991, 101871994, 101871995, 101871998, 101872002, 101872003, 101875823, 101884649, 101884662, 101890660, 101890661, 101890683, 101895634, 101912771, 101912832, 101912833, 101912837, 101912838, 101916174, 101924149, 101863428, 101875818, 101882033, 101882034, 101882037, 101882038, 101901949, 101901951, 101904490, 101904492, 101907204, 101907205, 101909370, 101909371, 101909372, 101909464, 192052657 |
Recalling Firm/ Manufacturer |
Trumpf Medical Systems, Inc. 1046 Legrand Blvd Charleston SC 29492-7672
|
For Additional Information Contact |
Chandler Sprinkles 843-329-0543
|
Manufacturer Reason for Recall |
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543. |
Quantity in Commerce |
84 units |
Distribution |
Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland, Russia, Slovakia, and Switzerland |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|