| | Class 2 Device Recall ADVIA 2120i with Single Aspirate Autosampler |  |
| Date Initiated by Firm | March 05, 2021 |
| Create Date | April 09, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1400-2021 |
| Recall Event ID |
87488 |
| 510(K)Number | K102644 K162977 |
| Product Classification |
Counter, differential cell - Product Code GKZ
|
| Product | ADVIA 2120i with Single Aspirate Autosampler (SMN 10488923) - OUS, automated hematology analyzer
Software Versions 6.10 and 6.11 |
| Code Information |
Software Versions 6.10 and 6.11 |
| FEI Number |
2432235
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown
NY
10591-5005
|
| For Additional Information Contact | SAME 914-631-8000 |
Manufacturer Reason for Recall | Potential Sample Identification (SID) Mismatch with14-Character Barcodes |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter HSW21-01.A.US to US customers on March 5, 2021 via FedEx overnight delivery. An Urgent Field Safety Notice (UFSN) HSW21-01.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities.
Letter states reason for recall, health risk and action to take:
with your Medical Director.
If you are using 14-character Sample Identification (SID) barcodes:
o Ensure that the Barcode Encoding Selectivity feature is enabled.
Open the Sample Identification screen to determine if the checkbox to turn the Barcode Encoded Selectivity is on
o If Barcode Encoding Selectivity is disabled, please perform the following steps to enable Barcode Encoding Selectivity:
Open the Sample Identification screen
Select the checkbox to turn the Barcode Encoded Selectivity on
Whether or not you have the Barcode Encode Selectivity feature disabled, please complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product
If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. |
| Quantity in Commerce | 969 units |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Albania, Algeria, Angola, Argentina, Australia,
Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium,
Benin, Bosnia Herzeg., Botswana, Brazil, Bulgaria, Burkina, Faso, Cambodia, Canada, Chile, Colombia, Croatia, Cura¿ao
Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland,
France, Fren.Polynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq,
Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait,
Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Mali
Malta, Martinique, Mexico, Namibia, Netherlands, Norway,
Oman, P.R. China, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania,
Russian Fed., San Marino, Saudi Arabia, Senegal, Serbia,
Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden,
Switzerland, Syria, Taiwan, Thailand, Trinidad,Tobago, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Uruguay,
Vatikancity, Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|