| Date Initiated by Firm | June 04, 2021 |
| Date Posted | July 12, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2063-2021 |
| Recall Event ID |
88147 |
| 510(K)Number | K142085 K160618 K163331 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | Optima MR450w 1.5T, Nuclear Magnetic Resonance Imaging System |
| Code Information |
N/A |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE Healthcare Service 800-437-1171 |
Manufacturer Reason for Recall | GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images. |
FDA Determined Cause 2 | Software design |
| Action | Deployment of the original customer letter to consignees in the U.S. began on June 4, 2021. Dissemination of the updated customer letter to consignees in the U.S. began on June 11, 2021. The letters discuss an issue on the affected products listed below where the Patient Orientation button may inadvertently be clicked when intending to click on the Save RX button. Users are directed to review and confirm that the actual patient orientation matches the prescribed patient orientation before initiating the scan. A correction is planned. |
| Quantity in Commerce | 1519 units |
| Distribution | Worldwide distributions. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|