• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall HA FlexTrak

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall HA FlexTraksee related information
Date Initiated by FirmMay 03, 2024
Date PostedJune 04, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1948-2024
Recall Event ID 94603
510(K)NumberK213583 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductHA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MR system, Product Number: 989710006411
Code Information UDI: N/A Serial Numbers: 3 5 6 9 11 12 13 14 15 16 17 18 19 20 22 23 25 26 29 30 31 32 33 34 35 37 38 39 41 42 43 44 45 49 50 51 52 53 55 56 58 61 63 64 66 67 68 70 71 72 73 75 77 79 80 81 82 83 84 86 87 88 89 91 93 94 95 97 98 100 101 102 103 104 105 107 108 109 110 111 112 113 114 115 116 117 118 119 120 122 123 124 125 126 129 130 132 133 135 136 137 138 139 140 141 142 143 146 148 149 150 151 152 153 154 155 156 158 159 160 161 162 163 164 165 167 169 170 171 172 173 174 175 177 180 181 183 184 185 186 188 189 190 191 192 193 195 196 197 198 200 201 202 203 204 205 206 207 208 209 211 212 213 215 216 218 219 220 221 222 223 224 225 226 228 229 232 233 234 235 236 237 238 240 241 242 243 244 246 248 250 251 253 254 255 257 258 259 260 262 263 264 265 267 269 270 272 273 274 275 278 279 281 282 284 286 287 288 289 290 291 292 293 294 295 296 297 298 301 302 303 304 305 307 308 309 310 311 312 313 314 315 317 318 319 320 322 323 324 325 326 328 331 332 334 335 336 337 338 339 340 341 342 343 344 345 346 347 350 351 352 355 356 358 359 360 361 362 363 364 366 367 368 369 370 371 372 373 374 375 379 380 381 382 384 386 387 388 390 391 392 394 395 396 397 399 400 401 402 403 404 405 406 408 409 411 412 413 414 415 416 417 419 420 421 422 423 424 426 427 428 429 431 432 434 435 436 437 438 439 440 441 442 444 445 446 448 449 450 451 452 453 454 456 457 460 461 462 463 464 465 466 467 468 469 472 473 474 475 476 477 478 479 480 481 482 483 485 487 489 491 492 493 494 496 497 498 499 500 501 502 503 504 506 507 508 509 510 511 512 515 516 517 518 520 522 523 524 525 528 529 531 532 534 535 536 537 538 541 542 543 544 545 546 547 548 549 550 552 553 554 555 556 558 559 560 561 562 564 565 566 567 568 569 571 574 575 577 579 582 583 584 585 587 588 591 592 593 594 596 597 598 599 600 601 603 605 607 608 609 610 611 613 614 618 619 620 621 623 625 627 628 629 630 631 633 634 635 639 640 641 642 643 645 646 647 648 650 651 652 653 656 658 659 660 661 662 663 664 665 666 667 668 670 671 672 673 674 675 676 677 679 680 681 682 683 684 686 687 688 689 692 694 696 697 698 700 702 705 706 708 709 710 711 712 716 717 718 719 723 724 725 727 728 729 730 731 732 733 734 735 737 738 739 740 742 743 744 745 746 747 748 749 750 752 753 754 755 756 757 758 759 760 762 763 764 765 766 768 769 770 771 772 773 774 775 777 778 779 780 781 782 783 784 787 788 789 790 791 792 793 794 795 796 797 798 799 800 801 802 803 804 807 808 809 810 811 813 814 815 816 817 819 820 821 822 823 825 826 827 828 829 830 831 832 834 835 836 837 838 839 840 841 842 843 844 845 847 848 849 850 852 853 854 856 857 858 859 860 862 863 865 867 868 869 870 871 872 873 874 875 876 877 880 881 882 883 884 885 886 887 888 889 890 891 892 893 894 895 896 897 898 899 901 903 904 905 906 908 909 910 911 912 914 915 916 918 919 921 922 924 925 926 927 928 929 930 931 932 933 934 935 936 937 938 939 940 941 942 943 944 945 947 948 949 950 953 954 955 956 957 958 961 962 963 968 969 970 971 973 974 975 977 979 980 981 982 984 985 986 987 988 989 990 991 992 993 994 995 996 997 1000 1001 1002 1003 1004 1005 1006 1007 1008 1010 1012 1013 1015 1016 1017 1018 1019 1020 1021 1022 1023 1024 1025 1026 1027 1028 1031 1032 1033 1034 1035 1036 1037 1038 1039 1040 1043 1044 1045 1046 1047 1048 1049 1050 1051 1053 1054 1055 1056 1057 1059 1060 1061 1062 1063 1064 1065 1066 1068 1069 1071 1073 1074 1076 1077 1078 1079 1080 1083 1084 1085 1086 1088 1090 1091 1092 1094 1095 1096 1097 1098 1099 1100 1101 1103 1104 1105 1106 1107 1108 1109 1110 1113 1114 1117 1118 1119 1120 1121 1123 1124 1125 1126 1130 1131 1132 1133 1135 1136 1138 1139 1140 1141 1142 1143 1145 1146 1147 1148 1149 1150 1151 1153 1154 1155 1156 1157 1159 1160 1161 1162 1163 1164 1165 1166 1167 1168 1169 1170 1172 1174 1175 1177 1178 1179 1180 1181 1182 1183 1184 1185 1186 1188 1189 1190 1191 1194 1197 1198 1199 1200 1201 1203 1204 1205 1206 1207 1208 1209 1210 1211 1212 1214 1215 1216 1217 1218 1219 1220 1221 1222 1223 1224 1226 1227 1228 1229 1231 1233 1234 1235 1236 1237 1238 1239 1240 1241 1242 1243 1244 1247 1248 1249 1250 1252 1253 1254 1256 1258 1260 1261 1262 1263 1264 1265 1268 1270 1272 1273 1274 1275 1276 1277 1278 1280 1281 1283 1285 1286 1287 1288 1290 1291 1293 1295 1296 1297 1298 1299 1300 1303 1304 1305 1306 1307 1308 1309 1311 1313 1314 1315 1316 1317 1318 1319 1320 1321 1322 1323 1324 1325 1326 1327 1328 1329 1330 1331 1332 1334 1335 1336 1337 1338 1339 1341 1342 1343 1347 1348 1349 1351 1352 1353 1354 1355 1356 1357 1358 1359 1360 1361 1362 1363 1364 1365 1366 1367 1368 1369 1370 1371 1372 1373 1374 1375 1377 1378 1379 1380 1381 1382 1383 1385 1386 1387 1388 1390 1393 1394 1396 1397 1398 1399 1400 1402 1403 1404 1405 1406 1407 1410 1411 1412 1413 1414 1415 1417 1418 1419 1421 1422 1423 1424 1425 1426 1427 1428 1429 1430 1431 1432 1433 1436 1437 1439 1440 1441 1442 1443 1444 1445 1446 1447 1448 1449 1450 1451 1452 1454 1455 1457 1458 1460 1461 1462 1463 1464 1465 1466 1467 1468 1469 1470 1471 1472 1473 1474 1475 1476 1477 1478 1479 1481 1482 1483 1484 1485 1486 1487 1488 1489 1490 1491 1492 1493 1494 1495 1496 1497 1499 1500 1501 1502 1503 1504 1507 1508 1509 1510 1511 1512 1513 1514 1515 1516 1517 1518 1519 1520 1521 1522 1523 1524 1525 1526 1527 1529 1530 1531 1532 1533 1534 1535 1536 1537 1570 1571 1572 1573 1574 1575 1577 1578 1579 1580 1581 1582 1583 1586 1589 1590 1592 1594 1595 1596 1597 1598 1600 1601 1602 1606 1607 1608 1609 1610 1611 1613 1614 1615 1616 1617 1619 1620 1621 1622 1623 1624 1625 1626 1627 1628 1629 1630 1632 1633 1651 1652 1656 1657 1658 1659 1660 1662 1664 1665 1666 1699 1703 1704 1705 1706 1708 1712 1713 1714 1715 1717 1718 1720 1721 1722 1723 1726 1727 1772 2060 101683 101684 101685 101686 101688 101689 101690 101691 101692 101693 101694 101696 101701 101707 101711 101719 101725 101728 101729 101730 101731 101733 101734 101738 101739 101740 101741 101742 101743 101744 101745 101763 101767 101768 101769 101770 101771 101773 101775 101777 101778
Recalling Firm/
Manufacturer
Philips North America Llc
222 Jacobs St
Cambridge MA 02141-2289
For Additional Information ContactCustomer Care Solution Center
800-722-9377
Manufacturer Reason
for Recall
Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard
FDA Determined
Cause 2
Under Investigation by firm
ActionPhilips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: A. If you see hydraulic fluid (oil) leakage: " First remove fluid from the floor using a paper towel. Then remove any remaining oil film from the floor using a cleaning agent (e.g. soap and water) and paper towel. " Contact your local Philips service representative. B. You may continue to use your HA FlexTrak Trolley in accordance with the intended use. C. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm until this issue has been resolved. D. Please post this letter near the associated MR system(s) for ease of reference. E. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction, understanding of the issue, and required actions to be taken. A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the HA FlexTrack Trolley and correct if required (reference FCO78100580). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday).
Quantity in Commerce1348 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, Oman, Peru, Poland, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LNH
-
-