| | Class 2 Device Recall HA FlexTrak |  |
| Date Initiated by Firm | May 03, 2024 |
| Date Posted | June 04, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1949-2024 |
| Recall Event ID |
94603 |
| 510(K)Number | K213583 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | HA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MR system,
Product Number: 989710006412 |
| Code Information |
UDI: (01)00884838099470(21)
Serial Numbers:
1538
1539
1540
1541
1542
1543
1544
1545
1546
1548
1550
1552
1553
1554
1556
1557
1558
1559
1560
1561
1563
1564
1565
1566
1567
1568
1569
1634
1635
1636
1637
1638
1639
1640
1641
1642
1643
1644
1646
1647
1648
1649
1669
1670
1672
1673
1674
1675
1676
1677
1678
1679
1680
1681
1682
101747
101749
101750
101751
101753
101754
101756
101757
101758
101759
101762
101779
101780
101781
101782
101783
101784
101785
101787
101789
101790
101791
101792
101793
101794
101795
101797
101798
101799
101800
101801
101802
101803
101804
101806
101807
101809
101810
101811
101812
101813
101814
101850
101851
SN101852
|
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
| For Additional Information Contact | Customer Care Solution Center 800-722-9377 |
Manufacturer Reason for Recall | Oil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Philips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US.
Letter states reason for recall, health risk and action to take:
A. If you see hydraulic fluid (oil) leakage:
" First remove fluid from the floor using a paper towel. Then remove any remaining oil film
from the floor using a cleaning agent (e.g. soap and water) and paper towel.
" Contact your local Philips service representative.
B. You may continue to use your HA FlexTrak Trolley in accordance with the intended use.
C. Circulate this notice to all users of this device so that they are aware of the issue and
associated hazard/harm until this issue has been resolved.
D. Please post this letter near the associated MR system(s) for ease of reference.
E. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the URGENT Medical Device Correction, understanding of the issue, and required actions to be taken.
A Philips representative will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site to inspect the HA FlexTrack Trolley and correct if required (reference FCO78100580).
If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377, 8AM-8PM EST, Monday-Friday). |
| Quantity in Commerce | 100 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain,
Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark,
Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Jordan, Lithuania, Martinique, Mexico, Morocco,
Netherlands, Norway, Oman, Peru, Poland, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|