• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Venue Go

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Venue Gosee related information
Date Initiated by FirmMarch 06, 2025
Date PostedApril 11, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1555-2025
Recall Event ID 96538
510(K)NumberK183362 K202233 K220800 K240053 
Product Classification System, imaging, pulsed doppler, ultrasonic - Product Code IYN
ProductVenue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for China (Model Number H45113VGSU) diagnostic ultrasound systems
Code Information UDI-DI: Venue Go R2 system (GTIN 00840682138840); Venue Go R3 system (GTIN 00840682190503 / GTIN 00195278358011); Venue Go R4 system (GTIN 00195278516510 / GTIN 00195278516527); Venue Go R3 Surgery for China (GTIN 00195278547866) Venue Go R3 PeriOP for China (GTIN 00195278548344) Venue Go R3 Expert for China (GTIN 00195278547262) Venue Go R3 Pro for China (GTIN 00195278547828) Venue Go R3 Performance for China (GTIN 00195278547842) Venue Go R3 Intervention for China (GTIN 00195278548061) Venue Go R5 system (GTIN 00195278450319 / GTIN 00195278451774); System ID: VENGO000382 VGB002128 VGB003021 VGD000325 VGD002313 VGD000452 VGD000488 VGD002390 VGC006411 VGB002040 VGD000354 VGC000685 VGD001444 VGC006698 VGC007776 RVGC000034 RVGC000035 VGD002256 VGD002533 VGD002535 VGD002536 VGB001162 VGC007292 VGC007119 VGC006083 VGC006841 VGB001011 VGC000147 VGD001510 RVGC000559 RVGC000562 VGD000044 VGD001371 VGD001372 VGD001375 VGD001374 VGD001373 VGD001625 VGD001638 VGC000526 RVGD002184 VGD002513 VGB002239 VGD001705 VGD002307 VGC000746 VGC004373 VGC004378 VGC006707 RVGC007218 RVGC000591 VGD001899 RVGC004147 VGC004351 VGC007098 VGC007102 VGC007105 RVGC000024 RVGC004165 RVGD001551 RVGB000480 RVGB002675 VGC007113 VGD000091 RVGD001076 RVGB000467 VENGO000545 VGC007350 VGD001731 VGB002526 VGC000227 VGD001964 VGD002225 RVGC000303 RVGB000757 RVGB000797 RVGC000302 RVGC000554 VGC000799 RVGC006372 RVGC006771 VGC007879 VGC007880 VGD000524 RVGB000791 VGD001873 VGD002221 VGD002227 VGD002467 VGD000730 VGD001872 VGC007607 VGC007611 RVGC007204 VGC006763 VGD000731 VGD000843 RVGB000694 VGD000553 VGD000554 VGD000558 VGD001336 VGC000898 VGD000200 VGC006715 VGD002460 VGD002462 VGD002427 RVGB000032 RVGC000594 RVGC000617 VGD001738 VGD001747 RVGC006758 RVGD001055 VGB000796 VGB000901 RVGD001795 VGD001400 VGD001401 VGB002539 VGB002613 VGB001125 VGD001200 VGD001257 VGD001258 VGD001258 VGD001257 VGD001261 VGD001257 VGC000386 RVGD001071 VGD002275 VGD001701 VGD001704 VENGO000317 VGB002122 VGD001687 VGD001884 VGD001885 VGD001889 VGC006031 VGC006036 VGC004366 VGC004521 RVGD001797 VGB001150 VGB000766 VGC000947 VGC000948 VGC006790 VGD000142 VGD001602 VGD000814 VGC006847 VGD001463 VGD001475 VGD001584 VGD002463 VGB000701 VGC000109 VGD001673 VGB000737 VGC000379 VGC004001 VGD001058 VGD001923 VGD001982 VGD001984 VGC007456 VGC007643 VGC007644 VGD002283 VGD002285 VGB002527 VGB002531 VGD001569 VGD002397 VGD002506 VGB000720 VGB000722 VGB000723 VGD001090 VGB002102 VGC000435 VGC000951 VGC004160 VGC007136 VGB002574 VGC007639 VGD002403 VGD002407 VGD002411 RVGB000122 VGC000866 VGC006812 VGC006814 VGC000082 VGC000084 VGC000086 VGC000085 VGC006402 VGC006406 VGC006652 VGC007049 VGC007053 VGC006409 VGD000035 VGD002484 VGD001037 VGD001150 VGD001900 VGD002237 VGB000165 VGC000943 RVGB000184 VENGO000700 RVGC000185 RVGC007163 VGC000321 RVGB000265 VGB000674 RVGC006341 VGD000399 VGD002328 VGB001415 RVGC000488 14540 VGB001140 VGB001916 VGB002688 RVGB000226 RVGB000472 VGD000932 RVGD001065 RVGD001427 VGC000671 VGC006086 VGC000156 VGC006113 VGC006127 RVGB000425 RVGB000171 VGC006710 VGD000603 VGD002018 VGC000879 RVGB000073 VGD002487 VGD002375 VGD000969 VGD000970 VGC007104 VGB001132 14617 14280 VGD002377 VENGO000703 VGB000710 VGB000707 VGB002548 VGD001005 VGD002230 VGD001573 VGB001934 VGB002489 VGB002565 VGC006105 VGD001043 VGD001482 VGC000407 VGD002126 VGB001009 VGD002509 VGD000914 VGC006958 VGD000350 VGD000353 VGD002170 VGC000099 VGC000115 VGC000099 VGC000146 VGD001298 VGD001301 VGB002073 VGC000269 VGD001275 RVGB000420 VGD001933 VGB001168 VGD002234 VGC006278 VGC007636 VGD001719 VGC006661 VGB002734 VGC006294 VGC006296 VGB002469 VGD002189 VGC000718 VGD002362 VGC000521 VGB000466 VGC000324 VGC000319 VGC000322 VGC000322 VGC000324 VGC000327 RVGC000019 VGB000998 VGC000278 VGD000138 VGC006956 VGD001161 VGD001163 RVGB000437 VGB000911 VGB001876 VGD002280 RVGC000570 VGD000513 VGC006534 VGD000902 RVGC000041 VGC006995 VGD001390 RVGB002663 RVGB002674 VGC006562 VGC006563 VGC006567 VGC006569 VGC006570 VGC006574 VGD002329 RVGC000040 VGB002970 VGD002036 RVGB000116 RVGB000160 RVGB000832 VGD002466 RVGD000137 VGB003006 VGD000307 VGD002178 VGB004063 VGC004102 VGC004110 VGD001382 VGD001749 VGD001656 VGC006697 VGC007044 VGC006837 VGD001949 VGC000487 VGD002003 VGD002062 VGD000753 VGD000761 VGC006816 VGB002524 VGC000279 VGC006665 VGD002056 VGD001051 VGD001052 VGD001639 VGB000684 VENGO000686 VGB000688 VGC007135 VGB000712 VGD002516 VGB000865 VGC000130 VGC000480 VGC000703 VGC000704 VGC007107 VGD000536 VGB002772 VGC006340 VGC000346 VGD002370 VGC006702 VGD002017 RVGB000066 RVGB000793 RVGB002244 VGD001517 VGD001526 VGB001268 VGB002573 VGC000323 VGC000364 VGC000741 VGC000750 VGC000753 VGC006443 RVGC007168 VGC007421 VGC007422 VGC004000 VGC006634 VGC007326 VGD002350 VGC006672 VGD000306 VGC004205 VGC004206 VGD000176 VGD001514 VGB004017 VGB004019 VGC000081 VGC000083 VGC000090 VGC000089 VGC000092 VGC000091 VGC000087 VGC000093 VGC007032 VGC007037 VGC007032 VGC007053 VGC007037 VGC007032 VGC007048 VGC007051 VGC007052 VGD001713 VGD001718 VGC000086 VGD001897 VGC007741 VGD001194 VGD002172 VGC007455 VGD002346 VGC007023 RVGB000383 VGD000541 VGD000546XX VGD000955 VGD001377 VGD002233 VGC000094 VGD000297 RVGB000808 VGD000777 RVGD001755 VGB002547 VGD001023 VGC006506 VGC006508 VGD000938 RVGC006368 RVGC006370 VGD000901 VGB000992 VGB001020 VGB001246 VGB001999 VGB002696 VGB002281 VGB002554 VGB002555 VGB002556 VGB002557 VGB002559 VGB002561 VGC000911 VGC000970 VGD002132 VGD002134 VGD002273 RVGC006682 VGC000680 VGC006650 VGC006669 VGD000424 VGD000445 VGB000666 VGB002462 RVGB002692 VGB002781 VGC004144 VGD001694 VGD001695 VGB002098 RVGD001798 VGD002340 VGB000699 VGB002117 VGC004161 VGC004263 VGC004266 VGC007387 VGD002082 VGD002181 VGC006093 VGD000235 VGD001217 VGB000756 VGB002870 RVGC000045 VGC004458 VGD002041 VGD002051 VGD002041 VGD000247 RVGD001450 RVGB000470 VGD001645 VGD001847 VGC000760 VGC000763 VGC000785 VGC007041 VGD000391 VGC007400 VGC006723 VGD000040 RVGC000168 VGC004038 VGD000045 RVGB000427 RVGD001537 VGB002100 VGB002104 VGB002106 VGB002108 VGC006103 VGC006678 VGC007039 VGB001414 VGD002205 VGD002279 VGD002315 VGC007060 VGC004141 VGC000072 VENGO000293 VGD000310 VGD000317 VGD000310 VGC000461 VGC006070 VGC007596 VGD000985 VGC004078 RVGC000585 RVGC004182 VGD002473 RVGC000001 RVGC006764 VGB000655 VGD000343 VGB000649 VGD001830 VGC000936 VGC007431 VGC000004 VGC004463 VGC004512 VGC006575 VGD000538 VGD000555 VGC000004 VGC000098 VGC000097 VGC000106 VGC000105 VGC004513 VGC004514 VGC007064 VGC007069 VGC007073 VGD000234 VGB001004 VGB002039 VGC006675 VGD000342 VGB002170 VGD001710 VGC004497 VGC006062 VGD001857 VGD002033 VGC000742 VGC000756 VGC000903 VGC004116 VGC004118 VGC004506 VGC009011 VGD002337 VGD002338 VGC000143 RVGB001984 VGC006052 VGB002004 VGC000688 VGC000701 VGC000720 VGC007210 VGD002300 VGC000670 VGD000925 VGB001015 RVGB000252 VGC000417 VGB000692 VGB001480 VGB001491 VGB002016 VGB002052 VGB002123 VGC000533 VGC000921 VGD000715 VGD001031 VGD001337 VGD001337 VGD001337 VGD001337 VGD001611 VGD001953 VGD001955 RVGB000354 VGC000530 VGC000200 VGC000219 VGC000213 VGC000219 VGC000219 VGC000219 VGC000219 VGC004107 VGC004490 VGC006338 VGD000790 VGD001160 VGD001782 VGB000693 RVGC006684 VGC007562 VGC006759 VGD002209 VGD000884 VGD001481 VGC000602 VGD001457 VGD001468 VGB002311 VGC000481 VGC000489 VGC006041 VGC006043 VGC006094 VGC006124 VGC006126 VGC006913 VGC007631 VGC007635 VGD002402 VGD002405 VGD002408 VGD001118 VGD002376 VGC000645 VGC007671 VGC006794 VGC006169 VGC004086 RVGC004150 VGC004403 VGD000155 VGD001737 VGD001986 VGD002005 VGD002016 VGD002025 VGD002030 VGD002034 VENGO000403 VGB000503 VGB000504 VGB000510 VGC004516 VGC006378 VGC007027 VGD002507 VENGO000339 VGB002157 VGB000805 VGC007707 VGD000917 VGB000805 VGB000671 VGB000680 VGB000685 VGC000138 VGC000140 VGC000142 VGC000151 VGC000152 VGC004518 VGC004518 VGC000103 VGD002323 VGD002326 VGD002327 VGD002392 VGB001982 VGC006720 VGD001486 VGB001917 VGC007733 VGC007233 VGD000057 RVGB000835 VGD000302 RVGB000199 VENGO000385 VGC000077 VGC000759 VGC000762 VGC006030 VGC006035 VGD000926 VGB002643 VGD002086 VGD002086 VGC004479 VGC004461 VGC004027 VENGO000459 VGB002383 VGD002013 VGD001388 VGD001918 VGC004030 RVGC007198 VGD001085 VGD001266 VGD001278 VGB001269 VGC000403 VGC004488 VGC004493 VGB002029 VGC000385 VGC006029 VGC006034 VGC006444 VGC006450 VGC006967 VGC006988 VGC000485 VGC007781 VGB002129 VGB002130 VGC000144 VGC000934 VGC000935 VGC004026 VGC004508 VGC004526 VGC007063 VGC000953 VGC004515 VGC006071 VGB000667 VGC004509 VGC000950 VGB002101 VGC000912 VGC000932 VGC000965 VGC007450 VGD002414 VGD000935 VGD002434 VGD002450 VGB002431 VGC004197 VGC004199 VGC004201 VGC004203 VGB000873 VGB001002 VGB001994 VGD001004 VGB002520 VGC000421 VGC000413 VGD002530 VGD002538 VGB002030 VGD000832 VGD001197 VGD001378 VGB000673 VGD002265 VGC000522 VGC006243 VGC006252 VGC000897 VGB002140 VGB002141 VGB002492 VGB002512 VGB002626 VGB002641 VGC000073 VGC000074 VGC000074 VGC000478 VGC000479 VGC000539 VGC000895 VGC004143 VGC006084 VGC006097 VGC006139 VGC007075 VGC007594 VGC007601 VGC007637 VGD000797 VGD001459 VGD001460 VGD001540 VGD001823 VGD002341 VGD002511 VGC004034 VGC006140 VGC007002 VGB002532 VGC004023 VGC004517 RVGC007169 VGC000463 RVGB000402 VGC000924 RVGC006848 RVGC007194 VGD000590 VGC004088 VGC007618 RVGC000066 RVGD000338 RVGD001066 RVGD001067 RVGD001068 RVGD001070 RVGD001466 RVGD001506 RVGD001550 RVGD001622 RVGD001624 RVGD001789 RVGD001801 VGD000341 VGD000362 RVGC000012 VGD000311 VGD001032 VGD001034 VGD001958 VGD001483 VGB002103 VGC004559 VGC007835 VGD002011 VGD002021 VGB001104 VGC000929 VGD002413 VGC007630 VGC000392 VGD001493 VGB001174 VGC000532 VGB002125 VGB002533 VGC006419 VGC007395 VGD001120 VGD001121 VGD001151 VGD001152 VGB002267 VENGO000493 RVGC006687 VGC004066 RVGC007160 VGC006593 RVGB002253 RVGC004124 3061702 VENGO000444 RVGC006608 RVGB000130 VGC007121 VGB002025 RVGC000592 VGD002260 VGB001183 VGB001187 VGB001189 VGB002706 VGC000916 VGC006653 VGC006664 VGD001667 VGC000515 VGD000345 VGC006818 VGB000813 VENGO000528 VGB000842 VGC000484 VGC000890 VGC004481 VGD000330 VGD002004 VGD002503 VGD002508 VGD000336 VGC000958 VGD001836 VGC004248 VGC004255 RVGC006352 RVGC006774 RVGC006775 RVGC007154 RVGC007155 RVGC007166 RVGC007192 VGD000111 VGD000113 VGD000116 VGD001429 VGD001518 VGD002136 VGD002254 VENGO000687 VENGO000695 VGD001740 VGD001741 VGD002075 VGC004467 VGC004487 VGC007725 VGD001089 VGC000308 VGC000312 VGC000314 VGD000229 RVGD000081 VGD001515 3137004 VGD001686 VGC007599 VGB000390 VGB001021 RVGC000044 RVGC000046 RVGC000047 RVGC000062 RVGC000064 RVGC000248 RVGC000043 VGD000765 VGC007006 VGD001564 VENGO000422 VENGOB000327 VGB002082 VGB002378 VGB002391 VGB002681 VGB002715 VGC004472 VGC004473 VGD000998 VGD001008 VGD001008 VGD001396 VGD001541 VGD002147 VGD002168 VGD001226 VGD000899 RVGB000047 VENGO000439 VGD000117 VGD001729 VENGO000245 RVGC000015 VGC007031 VGC007035 VGC007036 RVGC006769 VGC000289 VGC000382 VGC004480 VGC004485 VGD001812 VGC007641 VGD001643 VGC004094 VGC004560 VGD000054 VGD000055 VGD000049 VGB002673 VGD002053 VGB000060 VGB000658 RVGC004135 RVGC006765 VENGO000394 VGD001972 VGD001478 VGD001479 VGD000272 VGB002174 RVGB000369 VGB001184 VGB001191 VGB001901 RVGB002261 RVGB002679 VGC000110 RVGC000616 RVGC006545 RVGD001387 VGD001806 VGD002336 VGD001967 RVGB002788 VGD002475 RVGC007157 VGD002173 VGD002175 VENGO000389 VGB001899 VGC000436 VGC000080 RVGB000765 VGC007056 VGD001576 VGC006996 VGC006999 VGB001903 VGC006358 VGC006830 VGD000038 VGD000115 VGD000149 VGD001489 VGD001501 VGD001502 VGD001505 VGD001508 VGB000817 VGB001568 VGD002259 VGD000344 VGB002568 VGC004507 VGC004520 VGD001030 VGB001897 RVGB000768 RVGC000576 VGD000561 VGB002523 VGC007443 RVGB002097 VGC000408 VGC000416 VGC006044 VGD000539 VGD000545 VGD002187 VGD000324 VGC000154 VGC000636 RVGB000743 VGD001756 VGD002135 VGC004064 VGC006641 VGD000332 VGD000333 VGD000421 VGD000436 VGD000443 VGD001934 VGD002295 VGD002330 VGD002332 VGD002334 VGC007271 VGC007275 VGC007277 VGC007271 VGC006683 VGC006727 VGD000359 VGB002525 VGB002731 VGC007663 VGD002373 VGD000741 VGC006783 VGD001536 VGC004238 VGD001745 VGB000406 VGC000277 RVGC007164 VGC007389 VGB002120 VGB002133 VGB002138 RVGB000173 VGD001808 VGD001948 VGD001608 VGD001610 VGD001613 VGB002486 VGC000376 VGD001079 VGD000752 VGD000754 VGC006668 VGC000078 RVGB002271 1000000926 VGD000997 VGD002235 VGC006696 VGC006700 VGC006704 VGC006957 1000003449 VGC006122 VGB002543 VGC006074 VGD001978 VGC000171 VGC000171 VGB003051 VGC007348 VGC007363 VGB002766 VGD000245 VGD001050 VGC000055 VGB004004 VGB004005 VGB001870 RVGB000809 VGC006114 VGD000232 RVGC000030 RVGD001666 VGC006260 VGC007132 RVGB000150 RVGC000587 VENGO000076LO VENGO0002407LO VENGO000241LO VENGO001981LO VENGO001992LO VENGO000233LO VENGO000720LO VENGO004236LO VENGO006099LO VENGO007263LO VENGO000559LO VENGO000907LO VENGO001750LO VENGO002398LO VENGO000981LO VENGO002054LO VENGO000085LO VENGO002100LO VENGO004196LO VENGO000084LO VGD001708 VGD002252 VGB000807 VGD001981 VGC000519 VGD000892 VGD002065 RVGB000818 RVGD000017 VGC006525 VGC006528 VGD000608 VGD000721 VGD000722 VGD001080 VGD001084 VGD002400 VGC007597 VGD001653 VGD002144 VGC000906 VGC004494 VGC004496 VGC006321 VGD002268 VGC004264 VGC004268 VGC007775 VGD000610 VGB002066 VGC004350 VGC000558 VGC006231 VGD001285 RVGB002260 VGC006908 VGC007114 VGD001419 RVGB000170 VGB000855 VGB002331 VGB002332 VGB002335 VGC000909 VGC000967 VGC000969 VGD001240 VGB002365 VGC004037 RVGC0000521 RVGC007213 VGD000058 VENGO000290 RVGB000177 VGB002208 VGD002048 VGC007533 RVGB002668 VGD002133 VGC007001 RVGC000586 RVGC004114 RVGC004136 RVGC004146 VENGO000381 VGD001547 VGC000491 VGC007110 VGB002118 VGB002121 VGB001885 RVGC000228 VGD001316 RVGB002273 VGC000520 RVGB002249 RVGC000028 VGC000184 VGC000187 VGD000355 VGD002420 VGD000606 VGD000607 VGD002179 VGC000358 VGD001714 RVGB002270 RVGC000226 RVGC000238 VGC006785 RVGB000178 RVGB000186 VGC000766 VGC006531 VGC006537 RVGB000108 RVGB000167 RVGB000029 RVGB000125 RVGB000137 RVGB000426 RVGB000144 VGB001288 RVGC004134 VGD002076 VGD002130 VGC004005 RVGC000249 VGD001669 VGD002289 VGD002299 VENGO000642 VGC007632 RVGB000198 RVGB000828 RVGB002258 VGD001635 VGB000871 VGB001926 VENGO000278 VENGO000281 VENGO000286 VENGO000289 VENGO000296 RVGB000117 VGD000923 VGD001392 RVGB000138 RVGB000142 RVGC006777 VGC000167 RVGB000143 VGD001439 VGD001535 VGC006147 VGC006148 VGC006149 VGC006150 VGD000540 VGD001413 VGD002476 VGD001243 VGB000732 VGB000754 RVGC000013 VGC007138 RVGC007179 RVGB002256 RVGB002275 RVGB002661 RVGB002662 VGC000281 VGC000281 VGC000422 VGC000500 VGC000499 VGC006911 VGD000473 VGD002412 VGD002421 VGC000499 VGC006337 VGB001262 RVGD000013 RVGD000016 RVGD000019 RVGC006826 VGC000764 VGC000771 VGC000774 VGC006631 VGD002384 RVGB002664 RVGC007180 VGD002180 VGC006067 VGB000647 VGC000161 VGC000161 VGD002514 VGD000072 VGB000856 VGB000868 VGB001001 RVGD001410 VGD001961 VGD001965 RVGC004130 VGD001854 VGD001041 VGD001255 VGD002393 VGC000772 VGC006660 VGC006660 VGD001297 VGD002023 VGD000885 VGC000937 VGC000393 VGC000523 RVGC006840 VGD000724 VGD000043 VGD000061 VGC000190 RVGD000130 VGD000557 VGC006538 VGD000289 VGD000394 RVGB000462 RVGB000468 VGD002430 RVGB000128 RVGB000133 RVGB000145 RVGB000152 RVGB000203 RVGB000231 RVGC006828 VGD002203 RVGB000159 VGD000766 VGC000957 RVGB000069 VGB000772 VGB000774 VGD000592 VGB001006 VGB001006 VGB002784 VGB002288 VGB001176 VGC007076 VGC007082 VGC007080 VGC007079 VGC000893 VGC000930 VGD001664 RVGB000039 VGD000047 RVGD001072 VGD001319 VGD001318 VGD002290 RVGD002537 VGD002198 VGD001696 RVGB000833 VGC007071 VGB002671 VGC006192 VGC006202 VGD000075 VGD000078 VGD000079 VGD000082 VGD000886 RVGC007186 VGC006123 VGB000755 VGD001578 RVGD000444 VGC000601 VGC007736 VGC006050 VGC006064 VGC006080 VGD000060 VGC006265 VGD001868 RVGB000482 RVGB000488 VGD000908 VGD000915 VGD000922 VGD000928 VGD000929 RVGD001527 VGD001630 VGB004012 VGC007480 VGD000062 VGC004379 VGC007068 VGD002363 RVGC000571 VGD002019 VGC000163 VGC000896 VGC007541 VGC006850 VGD002143 VGD002312 VGC000390 VGB002425 VGC000258 VGC007115 VGC000765 VGD000357 RVGC004140 VGC006861 VGC006863 VGD001790 VGB002310 VGC004455 VGC007025 VGC000945 VGC006673 VGB001197 VGB001951 VGB002139 VGB002538 VGD000986 VENGO000646 RVGB002124 VGD001671 RVGB002657 VGD002009 VGB002546 VGB000999 RVGD000241 VGB000690 RVGC006548 VGD000204 VGC007434 VGB002224 VGB002230 VGD000323 VGC000889 VGC000891 VGC006676 VGC006679 VGC007600 VGC000514 VGC006693 RVGD001406 RVGB000428 VGD002357 VGB001998 VGB002041 VGC000401 VGD002495 VGD002182 RVGB000080 RVGB000110 VGB002597 VGC006566 VGD001511 RVGD000153 VGC000116 VGC000117 VGC000119 VGD001484 RVGB000798 RVGB002220 VGD001628 RVGC000020 VGD001647 VGC000702 VGD000221 VENGO000239 VGD000150 VGD002154 RVGB000135 RVGB000277 RVGB000831 VGB002250 VGC000952 VGC004055 RVGB000230 VGC006383 RVGC000573 RVGB002450 VGC000976 VGC004010 VGC006796 VGC006799 VGC006912 VGC006621 VGC006699 VGD002417 VGC000053 VGC000057 VGC000441 VGC007500 VGC007502 VGC007503 VGC007591 VGD000820 VGD001585 VGD002468 VGD002470 VGD002472 VENGO000536 VGB000672 RVGB000799 VGC000173 VGC000352 VGD001748 VGD001954 VGC006807 RVGC000501 VGD001286 VGD001519 VGB002698 VGC000527 VGC006335 VGC007546 VGD001881 VGD000143 RVGC007167 VGD001938 VGD002345 RVGD001440 VENGO000434 RVGC000229 RVGC000307 RVGD001391 VGD001492 VGC004142 VGC007623 VGD002429 RVGD000015 RVGD000135 RVGD001432 RVGB000773 VGC006659 VGD002152 VGC007409 VGC000207 VGC007837 VGD001902 VGD002064 VGB003022 VGD002499 VGD002497 VGD002498 VGD002490 VGC007805 RVGB000263 RVGC000584 VGD000304 VGD000304 VGD000320 VGC007061 VGB000664 VGB000981 VGB000982 VGB001017 VGB001019 VGC004498 VGC006032 RVGC000425 VGC004008 RVGC000578 VGD001204 RVGC000310 RVGB000164 VGD002293 VGD002415 VGB001564 VGC007806 VGC006401 RVGC000006 RVGC000016 RVGC000026 RVGC000021 RVGC000025 RVGC000027 RVGC000029 RVGB000779 VGC007134 RVGB000096 RVGB000124 RVGB000126 RVGB000136 VGB001932 VGB000691 VGD000063 RVGC004133 VGC009042 VGC009044 VGC009045 VGB000702 VGC006501 VGC007783 VGD001256 VGD001264 VGC000885 VENGO000811 VGD001919 VGD001921 VGD001925 VGD001929 VGD001930 VGD001951 VGC006038 VGD001996 VGD000789 VGC006624 VGD001343 VGD001344 VGD001345 VGD001346 VGD001347 VGD001349 VGD001352 VENGO000324 VGB002294 VGB000834 VGB000866 VGD001179 VGD001187 VGD001190 VGD001191 VGD002081 VGD002091 VGD001189 RVGC000070 VGC007687 VGC007687 VENGO000717 VENGO000718 VENGO000719 VENGO000721 VGC000137 VENGO000282 VENGO000283 VENGO000287 VENGO000288 VGB002356 VGD002459 VGD002433 VGC000177 VGC000178 VGC000177 VGC004002 VGC004002 VGD000817 VGD000819 VGC007126 VGD000738 VGD000727 VGB002725 VGC007638 RVGC006344 VGD002320 VGD002322 VGD002494 VGC004019 VGD001730 VGC006987 VGC006805 VGD001366 VGD002319 RVGC000398 VGB002362 VGB002519 VGC004348 VGC006620 VGC006671 VGC006817 VGD000896 VGD001716 VGD001891 VGB002038 VGB002042 VGD002032 VGB002303 VGB002303 VGB000713 VGB000753 RVGC000577 VGD002387 VGD002389 VGD001828 VGD002202 VGC009058 RVGC000023 VGD002007 VGC007429 VGD001332 VGD001783 VGC006039 VGD001294 VGC007211 VENGO000393 VENGO000404 RVGB000075 VENGO000386 VGC000483 VGC006511 VGC006757 VGD001454 VGD002059 VGC006435 RVGB000049 RVGB001900 RVGD001528 RVGB000031 VGC006512 VGB000689 VGB000714 VGC000715 VGC007660 VGC007670 VGD001906 VGD001909 VGD001910 VGD001914 VGC007669 VGB001441 VGB001570 RVGB000424 VGC009030 RVGC000049 RVGD000014 RVGD000124 VGB000854 VGC000315 VGC000925 VENGO000285 VGB002773 VGC007000 VGC007059 VGC000974 VGD002453 VGB002172 VGC000412 VGC000513 VGC000513 VGC000528 VGC000902 VGD001487 VGC000949 RVGB002272 RVGB002262 RVGC000556 VGC007122 RVGB000824 VGC007481 VGB002001 RVGB000284 RVGC006680 VGC006142 VGD001011 VGD000052 VGD002523 VGD000596 VGB000677 VGB000678 VGB002521 VGC000767 VGC004183 VGC004344 VGC000108 VGC007721 VGB002158 VGC004251 VGC004257 VGD002291 VGD002304 VGD002305 RVGB000786 RVGB000821 VGB001986 VGC009013 VGC004438 VGD002368 VGD001467 VGB001962 VGB000683 VGD001183 VGD002047 RVGC000575 VGC000325 VGC000326 VGD002258 VGC006129 VGC007452 VGC006712 VGC006138 VGC004080 VGC006106 VGC006136 VGC006146 VGB000665 VGB000670 VGD000046 VGC000265 RVGD001420 VGB000668 VENGO000733 VGB002359 VGB002361 VGD001621 RVGD001600 RVGB000041 VGC006786 VGD002491 RVGB000792 RVGB002269 VGC000972 VGD000199 VGD000393 VGC000420 VGC006918 VGD001874 VGD001875 VGC000170 VGC006824 VGC006825 VGD001959 VGC007442 VGC007810 VGB001991 VGB001997 VGB002490 VGC000113 VGC000357 VGC000357 VGC007123 VGD001047 VGB001192 VGD001640 VGC000328 VGC007077 VGC007077 VGC007100 VGC007100 VGD001650 VENGO000295 RVGB002264 VGC004245 VGC004249 VGC004525 VGC004527 VGD000909 VGD000910 VENGOB000337 VGB001166 RVGC007158 VGD001042 RVGB000156 RVGB000421 VGB000452 VGB000456 VGB000474 VGC000359 VGC004016 VGC007070 VGC007140 VGC007330 VGC007338 VGC007339 VGC007340 VGC007745 VGC007764 VGC007769 VGD000944 VGD001678 VGD001980 VGC007074 VGC007081 VGC007336 VGD001173 VGD001198 RVGB000123 RVGB000193 VGC000939 RVGD001056 RVGD001525 RVGD001543 VGD000801 VGD001581 VGD001768 VGD001353 VGD001016 RVGC000036 VGC006045 VGC006058 VGC006057 VGC006058 VGD001123 VGD001534 VGD002114 VGD001751 VGD002115 VGD002108 VGD002109 VGD002115 VGD002109 VGD002106 VGD002114 VGD002115 VGD000846 VGD001832 VGC000975 VGC004180 VGC006834 VGD000583 VGD000584 VGD000888 VGD000903 VGD000988 VGD001835 VGD001837 VGB001118 VGD000795 VGD001572 VGC006061 VGD001876 RVGC000071 VGC006557 VGC006565 VGC006568 VGC006584 VGC009006 VGC009005 VGC009008 VGC009004 VGD002084 VGC000928 VGC000923 VGC000910 VGC000918 VGC007150 VGC007152 VGD000159 VGD000882 VGD000883 VGD000937 VGD001102 VGD001108 VGD001815 VGD001817 VGD001818 VGD001819 VGD001826 VGD001826 VGD001819 VGD001831 VGC007072 VGD002372 VGD002378 VGC004542 VGC006477 VGC006477 VGC006481 VGC006498 VGD000879 VGD000880 VGD000881 VGD000912 VGC004541 VGD000351 VGC004408 VGD000591 VGC006130 VGC006141 VGC006448 VGC006452 VGC006452 VGC006791 VGC006795 VGD000126 VGD000127 VGD000126 VGD000818 VGD000821 VGD000859 VGD000861 VGD000862 VGD000865 VGD000990 VGD001081 VGD001101 VGD001125 VGD001339 VGD001341 VGD001341 VGD002212 VGD002218 VGD002219 VGD002212 VGD002232 VGD002218 VGD002232 VGD002218 VGB001902 RVGC000581 VGC006654 RVGC000305 VGB001005 VGB002142 VGB002147 VGC004454 VGC004511 VGC006717 VGD001878 RVGC006835 RVGC006369 RVGC006677 VGC006651 VGD000887 VGD000889 VGD000889 VGD001003 VGD001279 VGD001283 VGD002211 VGD002216 VGD002211 VGD002211 VGC007765 VGD000594 VGC006657 VGD001376 VGD001469 VGD001470 VGD002303 VGD002324 VGD001851 VGC007272 VGC007814 VGD001252 VGC006947 VGD002122 VGD002123 RVGB000036 RVGB000046 VGB000654 VGC007127 VGC007231 VGD000601 VGD000609 VGD001549 VGD001632 VGD001800 VGB002530 VGD000890 VGD001024 VGD001317 VGD001335 VGD001869 RVGB000129 VGC007603 VGD000595 VGD000597 VGD000598 VGD000599 VGD000602 VGD001883 VGD000597 VGD002171 RVGB000155 VGD000600 VGC006120 VGC006125 VGC006133 VGC006135 VGD002145 VGD000957 VGD000960 VGD000961 VGD000964 VGB001022 VGC006096 VGC006109 VGC006112 VGC006131 VGC006360 VGC006361 VGB002534 VGC006527 VGD001082 VGD000292 VGD000314 VGD001009 VGD001015 VGD001098 VGD001099 VGD001103 VGD001106 VGD001110 VGD001111 VGD001119 VGD001122 VGD001271 VGD001288 VGD001289 VGD001293 VGD001342 VGD002022 VGD002057 VGD002093 VGD002094 VGD002096 VGD002097 VGD002098 VGD002101 VGD002102 VGD002103 VGD002104 VGD001879 VGD001882 VGD001888 VGD001890 VGD001894 RVGB000146 RVGB000147 RVGB000148 RVGB000166 RVGB000172 RVGB000151 RVGC006345 VGD001155 VGD001615 RVGB000481 RVGC000022 VGD001464 VGC000744 VGC000745 VGC000748 VGC000749 VGC000752 VGC000754 VGC007449 VGD002185 VGD001557 VGC007619 VGC007629 VENGO000297 RVGC006773 RVGC007212 VENGO000266 RVGC000418 RVGC006689 RVGD001195 RVGC000039 VGD000329 RVGB000140 RVGC007224 VGD002015 RVGB002665 RVGB002669 VGC000913 VGC006709 VGC007029 VGC007030 VGC007033 VGC007034 VGC007038 VGC007040 VGC007045 VGD000924 VGD001046 VGD001054 VGD001046 VGD001231 VGD001296 VGD001303 VGD001310 VGD001307 VGD001304 VGD001313 RVGB002268 VGC004462 VGD002479 VGD000231 VGD000231 VGD001892 VGD002186 VGD002190 VGD000965 RVGB000027 RVGB000054 RVGB000056 RVGB000059 RVGB000062 VGB000067 RVGB000154 RVGB000183 RVGB000229 VGB000514 VGB000661 RVGB000750 RVGC000051 RVGC000056 RVGC000059 RVGD000348 RVGD001078 RVGD001178 RVGD001399 RVGD001415 RVGD001425 RVGD001430 RVGD001431 RVGD001438 RVGD001441 RVGD001442 RVGD001521 RVGD001522 RVGD001532 RVGD001788 RVGD001792 RVGD001794 RVGD001799 RVGD001803 RVGD001810 RVGD002192 RVGD002194 RVGB000157 VGB000715 VENGO000770 RVGB000051 VGD000782 VGB000679 RVGB000068 VGC007768 RVGD001520 VGD001636 VGC000509 VGC000770 RVGC006829 VGD002382 VGD002383 VGD002382 RVGB000781 RVGB000781 RVGC000242 VGC004386 VGC006326 VGB000566 VGB000675 VGB001950 RVGB000052 RVGB000153 VGB002131 VGB002137 RVGC000593 RVGC000598 RVGC000611 VGC000946 VGC006662 RVGC006770 VGC004369 VGD000833 VGD000528 RVGC000239 VGB000711 RVGC000607 VGD002269 VGD001253 VGD002239 RVGC004121 VGC007062 VGD002469 VENGO000345 RVGC006968 VGC006299 RVGC0000541 VGC006069 VGC006072 VGC006073 VGC006085 VGC006076 VGC006078 VGC006079 VGC006076 VGC006075 VGC006098 VGC006100 VGC006101 VGC006110 VGC006639 VGC007129 VGC007130 VGC007521 VGC007538 VGC007539 VGC007540 VGD000313 VGD000315 VGD000319 VGD000905 VGD000418 RVGD001074 RVGC000017 VGB000104 VGC007197 VGC006821 RVGB000175 RVGB000763 RVGC000003 VGD000904 VGD002210 VENGO000508 RVGB000376 VGC000220 RVGB000127 VGB001185 VGB002360 VGC007228 VGC007388 RVGD001732 RVGC000608 RVGC004115 VGD002505 VGC004185 RVGB000174 RVGB000761 RVGB000761 VENGO000262LO VENGO000446 RVGC007172 VGD002471 VGC006572 VGC006718 VGC007278 VGC007808 VGC007809 VGB002062 VGB002064 RVGD001416 VGC006648 VGC004553 VGD000321 VGB002544 VGB002583 VGB002672 VGB002960 VGC004014 VGD001320 VGD002451 VGB002126 VGB002854 VGC006077 VGD001735 VGD001833 VENGO000228 VGD000823 VGD001654 VGD001674 VGD001676 VGD001677 VGD001781 VGC006820 RVGD001421 RVGB000457 VGC000400 RVGB000822 VGD002072 VGB001877 VGC007374 VGD001616 VGC000507 VGC006729 VGD000831 VGD001471 RVGD001796 VGC000920 VGB002542 VGB001113 VGC000268 RVGB000120 VGD001370 RVGB000826 VGB001178 VGD001991 VENGO000380 RVGC000564 VGD001988 RVGC006365 RVGB000196 VGD001273 VGD001603 VGB002055 VGD000331 VGD000337 VGD001985 VGD001993 RVGC006367 RVGC007165 RVGC007171 VGD001230 VGD001232 VGD001233 VGD001235 VGD001239 VGD001233 VGD001244 VGD002519 RVGC007184 VGD002464 VGD002517 VGC006509 VGD000449 VGD000587 VGD000588 VGD000605 RVGD001791 VGD001895 VGC006091 VGC004240 VGC004241 VGC004156 RVGB000070 RVGB000072 RVGB002670 VGB001270 VGB000657 VGD001631 VGD000023 VGD002481 VGD002481 VENGO000461 RVGC006741 VGD002014 VGC004009 VGB001890 VGC000329 VGD002354 VGB002145 VGB002146 VGD002454 VGD001963 RVGC009064 VGB002511 VGD000546 VGC004452 VGB002536 VGB002541 VGB004102 RVGB000819 VGC006355 VGB000708 VGC004331 VGC006688 VGD001029 VENGO000407 VGC007141 VGC007139 VGC007131 VGC007143 VGC007145 VGC007146 VGC007151 VGC007779 VGD000074 VGD000076 VGD000919 VGD000920 VGD000927 VGD000943 VGD001292 VGD001689 VGD001689 VGD001707 VGD001709 VGD000039 VGD000911 VGD001451 VGC000397 VGD000982 VGC000919 VGD000589 VGD001952 VGD002174 VGD002406 RVGC007195 VGC007288 VGC006815 VGC006980 VGB001919 VGB002299 VGB002367 VGB002370 VGB002371 VGC006629 VGD000571 VGB002373 RVGC000008 RVGC000009 VGC000658 VGC000658 VGC000661 VGC000662 VGC006111 VGC000506 VGD001461 VGB002560 VGB002563 VGC000747 VGC000955 VGC006092 VGD001966 VGD001968 VGD001969 VGD001968 VGD001990 VGD001992 VGD001994 VGD001995 VGD001966 VGD001969 VGC000188 VGC004370 VGC004375 VGC004435 VGD002366 VGD002416 VGB002591 RVGC004152 VENGO000541 VGC007042 VGD000048 VGD001839 VGD001840 RVGB000827 VGD001793 RVGD001646 VGC007133 VGC000561 RVGC004173 VGD001935 VGC007007 VGC007009 VGD000895 VGD000897 VGD000895 VGD001094 VGD001104 VGOVGC007882 VGOVGB002883 VGB000077 RVGB000025 VGC000095 VGC006618 VGB000506 RVGC007156 RVGC007161 VGD001725 VGB000659 RVGB000131 VGB001154 VGD001380 VGC006089 VENGO000233 VGC004464 VGC004495 VGD001485 VGD001488 RVGC006633 VGD001774 VGC006792 RVGB000163 RVGB000168 VGC006623 VGC006922 VGD000298 VGD002226 VGC004020 VGC007423 VGC006716 VENGO000048 VGC004105 VGC000668 VGD002243 VGB000975 VGC000595 VGC007393 VGB002372 VGC004261 VGC004466 VGC004470 VGC006137 VGC006946 VGD000404 VGC000894 VGD002396 VGD002214 VGC000695 VGD001597 VGD000767 VGB001018 VGC000971 VGC006417 VGC006674 VGC006761 VGC007604 VGD000776 VGD000780 VGD000784 VGD000785 VGD002127 VGD002125 VGD002127 VGB002399 VGD000284 VENG0000311 VENGO000325 VGC000768 VGC006063 VGC006068 VGD002215 VGD002253 VGD001225 VGD000269 VGD002482 VGC006691 RVGC007149 RVGC007182 VGD001049 VGD000303 VGD000305 VGD002318 VGB001995 VGD002465 VGD000768 VGD000770 VGC006760 VGD000544 VGD002297 VGC004006 VGD001105 VGC000334 VGC000331 RVGC006772 RVGC007181 VGC006128 VGC004387 VGB000663 VENGO000476 VGC000531 VGC000536 VGC000537 VGC000538 VGC000541 VGC000529 VGB000812 RVGB000200 RVGB002263 VENGO000492 VGD001648 VENGO000400 VENGO000405 VGC000096 VGC007112 RVGC000580 VGC006961 VGC000381 VGD002213 VGD001064 VGD001649 VGD002301 VGC006810 VGD000403 VGD001559 VGB001271 VGB001012 VGB001560 VGB002112 VGB002252 VGB002279 VGB002283 VGB002285 VGB002291 VGC000155 VGC000157 VGC000158 VGC000160 VGC000162 VGC004082 VGC000159 VGB002290 VGC000494 VENGO000705 VGD000318 VGD002401 VGB001149 VGB001153 VGB002056 VGB002057 VGB002058 VGB002059 VGD001561 VGD001567 VGD001583 VENGO000463 VENGO000464 VGC000512 VENGO000396 VGD000527 VGD001452 VGD002229 VGC007592 RVGC006685 VGB000698 VGD000556 VGD002356 VGD002207 VENGO000526 VENGO000565 VENGO000576 VGC004510 VGB002013 VGB002730 VGB002733 VGC004003 VGC004007 VGB000994 VGB002326 VGC004486 VGC004492 VGD000521 VGD001880 VGD002195 VGC000942 VGD001552 VGC004499 RVGB000829 RVGC000596 RVGB000197 VGC000743 VGC006327 RVGB000162 RVGB000176 RVGD000012 VGB000609 VGB002549 VGB002551 VGC006701 VGC007147 VGB000507 VGB000530 VGD000569 VGC000812 VGD000295 VGD001348 VGC004402 VGD001403 VGD001407 VGC000954 VGC007486 VGD000243 VGC000360 VGC004139 VGC004262 VGC004265 RVGC006836 VGC007057 VGC007058 RVGD000132 VGD001228 VGD001908 VGD001912 VGD001942 VGD001809 VGD001814 VGD001822 VGD001887 VGD001907 VGC000493 VGB001010 VGC007109 VGC006766 VGD000066 VGD002422 VGC006277 VGB000986 VGB002562 VGC000497 VGC000938 VGC000956 VGC007111 VGC007451 VGD000334 VGC000211 VGD000972 VGD002037 VGC000405 RVGB000471 VGD001618 VENGO000270 VGC000134 VGC000502 VGC000516 VGC004282 VGD000549 VGC006802 VGD001306 VGD001311 VGD001312 VGD002141 VGB002598 VGD001208 VGD001945 VGD002050 VGB000836 VGC004225 VGC004272 VGC007743 VGD002002 VGC006046 VGC006059 VGD002409 VGC000107 VGC000933 VGC007024 VGD000036 VGD000051 VGD001920 VGD001922 VGD002044 VGC007748 VENGO001008 VGB002951 RVGD000134 RVGB001722 VGC000215 VGC006066 VGD001546 VGD002246 VENGO000391 VGC000525 VGC004523 VGD002267 VGC004090 VGC007351 RVGD000139 VGC006049 VGD001642 VGC000434 VGD002274 VGC000508 VGC006219 VGD002248 VGD002078 VGD002385 VGC004304 VGC007440 VGD002296 VGC007144 VGD001229 VGC004243 VGC004259 VGC004415 VGC000079 VGC000126 VGC000214 VGC004137 VGC006670 VGC006823 VGC007103 VGC007106 VGC007108 VGD000918 VGD000991 VGD000548 VGC007436 VGC007494 VGC007496 VGC007507 VGC007507 VGC000330 VGC000474 VGC007534 VGC000330 VGC000474 VGC006087 VGD001386 VGC004482 VGC004484 VGC006102 VGD000547 VGD002043 VGD002066 VGC000276 VGC006107 VGB001931 VGC007125 VGD001877 RVGB000045 VGD000519 VGD000523 VGD000979 VGD000984 RVGC000065 RVGC000583 VGD001898 RVGC000042 RVGD000018 VGD001087 VGC006643 RVGB000478 RVGB002340 RVGD002199 RVGD002199 RVGC006108 VGD002060 VGD002071 VGD002060 VGD002071 VGB001164 VGB001165 VENGO000995 VGB002835 VGC000917 VGC000968 VGD000398 VGD002118 VGC000231 VGC000232 VGC000234 VGD000771 VGD000771 VGC004154 VGD002240 VGD002245 VGD002302 VGC000063 VGC000048 VGD002029 RVGB000042 RVGB000236 VGC006564 VGC007028 VGD002128 VGD001999 VGC007018 VGB000990 VGD002069 RVGB000134 VGC006090 VGD001063 VGD001555 VGD002418 VGC006500 VGD000059 RVGC006332 VGC000061 VGC006343 VGC007800 VGC006989 VGD001381 RVGC000579 VGC000676 VGD000913 VGB000562 VGC007128 VGC006051 VGC006053 VGC006054 VGC000182 VGB000650 VGB000681 VGB000682 VGC007065 VGD001620 VGD001668 VGB002545 VGD001927 VGC007428 VGC000153 VGC004145 VGD001634 VGD001813 VGC000645 VGC000428 VGC004242 VGC000757 RVGB000780 VGB002636 VGB002637 VGD002353 VGB002566 VGB002572 VGB002578 VGB002580 VGC007099 VGD000233 VGD000236 VGD000237 VGD000236 VGD001829 VGD002355 VGD002358 VGC000973 VGC007120 VGB002225 VGC000679 VGD001712 VGC007142 VGC000482 VGC007117 VGD000798 VGD002020 VGD000230 VGD001617 VGD001633 VGB000489 VGB002697 RVGC007202 VGD002284 VGC000402 VGC006040 VGC006042 VGC006104 VGD001113 RVGC000566 VGC000905 VGC000907 VGC000915 VGC000922 VGC004364 VGD001116 14616 VGB002115 VGD000864 VGD001411 VGB002522 VGC000282 VGD002518 VGD002012 VGC007116 VGD001367 VENGO000696 VGC007484 VGD000838 VGD002424 VGD000110 VGD000112 VGD002063 VGD002116 RVGD001825 VGC007518 VGD001619 VGD002264 VGC000941 RVGB002693 RVGC007159 VGD001623 RVGB000071 RVGB000830 VGC007642 VGC007723 VGC007771 VGC007801 VGC007807 VGD001295 VGD001601 VGD002287 VGD002529 VGD002534 VGC009043 VGD002183 VGD002270 VGC004100 VGC004132 VGD001937 VGD002321 VENGO000408 VGC004187 VGD001402 VGC000959 VGC006065 VGD002039 VGC007124 VGB000876 6070180WX0 VGD001644 RVGB000418 VGD001333 VGB001561 VGD001448 VGD001662 VGD001663 VGD001670 RVGB000423 RVGB000065 VGC007148 VGD001742 VGD001276 VGD000227 VGC000524 RVGC000419 VGC000464 VGB002552 VGD000729 VGD000735 VGB002352 VGC007004 VGD000327 VGC000510 VGC000496 VGC006706 VGC006731 VGD002077 VGC004471 VGC004489 VGD001928 VGB001126 VGB001128 VGB001133 VGB001134 VGB001138 VGB001145 VGB002207 VGC006725 VGD002288 VGB002550 VENGO000782 VENGO000783 VENGO000790 VGC006095 VGD001153 VGD002339 VGD002381 VGD001753 VGB000785 RVGC006622 VGC000194 VGD000339 VGC006986 VGC006919 VGD000328 VGD000383 VGD000543 VGC000145 VGC007761 VGD002120 VGD002121 VGD002119 VGD002395 RVGB002266 VGD001626 VGD001976 VGD001977 VGD002263 VGD001983 VGC006948 RVGB002241 RVGB002248 VGC000318 VGC006647 VGC006649 VGC006798 VGD001926 VGD002520 VGC000129 VGC000131 VGC000132 VGC000135 VGD001028 VGC000383 VGD000349 VGD001020 VGC007722 VGC000899 VGC000964 VENGO000441 VGC000535 VGC004012 VGC000259 VGC000139 VGC006787 VGC006839 RVGC007162 RVGC000590 VGC000164 VGC000165 VGC000244 VGC004011 VGD001660 VGD001957 VGD001962 VGD002035 VGD002435 VGC000769 VGD001802 VGD000815 VGD000816 VGD001870 VGD002073 RVGB000801 VGB002259 RVGC000842 VGC006949 RVGB000055 VGC000114 VGC000122 RVGC000247 VGC000270 VGC000271 VGC000272 VGC000781 VGC007564 VGC007566 VGD001605 VENGO000271LO VGC004081 VGD002138 VGB000704 VGB001013 VGB001014 VGB001180 VGB001190 VGD000517 VGD001384 VGD001389 VGD001553 VGD002052 VGB002079 RVGB000806 VGC000120 VGC000123 VGC000124 VGD001845 VGD002124 VGB002000 VGD001896 14281 14282 VGC004244 VGC004258 VENGO000372 VENGO000377 VENGO000429 VENGO000435 VGD000978 VGC000267 VGC007405 VGC006797 VGB000676 VGC007230 VGC004157 VGC006037 VGC006047 VGD001913 VGD002419 VGD002049 RVGB000053 RVGB000053 VGC000445 VGC000449 VGD000293 VGD000300 RVGB000776 VGD000525 VGC006849 VGC000111 VGD000941 VGD002200 VGD001434 VGD001435 VGD000825 VENGO000399 VGB002722 VGC000183 VGC004382 VGD000526 VGD001733 VGD001743 VGC000169 VGC000374 VGC000377 VGC000374 RVGD000011 RVGD001538 VGB001003 VGC006645 VGC006646 VGC006658 VGC006663 VGB002155 VGB002540 VGD002410 RVGC007153 VGC000666 VENGOB000322 VENGOB000323 VGB002235 VGD001592 VGD001824 VGD002139 VGC007506 VGC007509 VGB002553 VGC000504 VGC006132 VGD000056 VGB002444 VGD000238 VGD001998 VGC004524 VGC000940 VGC006285 VGB002529 VGC000362 VGD001820 RVGC000011 VGC000018 VGC004087 VGB002455 VGC006088 RVGB000169 RVGB000487 RVGB002265 VGB004016 VGC004564 VGC007026 VGC007645 VGC007803 VGB002076 VGB002086 VGB002087 VGB002088 VGB002089 VGB002090 VGB002091 VGB002092 VGB002094 VGB002095 VGB002096 VGB002156 VGB002514 VGB002063 VGB002069 VGC007528 VGC006284 VGB002084 VGB002085 VGB002153 VGD000294 VGC007022 VGB002026 VGD001395 VGD001353 VGC000892 VGC000926 VGC006082 VGC006134 VGC006144 VGC006292 VGD001675 VGD000968 RVGC007203 VGD000750 VGD002000 VGD002008 VGC006728 VGD001940 RVGB000061 RVGB000037 RVGB000043 RVGB000125 RVGB000023 RVGB000161 RVGB000248 RVGB000251 RVGB000419 RVGB000760 RVGB000778 RVGB000775 MBVGB000800 RVGB000803 RVGB000825 RVGB002274 RVGC000031 RVGC000010 RVGC000223 RVGC000589 VGD000322 VGD000347 RVGC006832 RVGD001558 RVGD001607 RVGD001805 VGD000244 VGD000020 VGD002223 VGD000346 VGD000356 VGD002316 RVGC000588 RVGC000604 VGD002286 RVGC004177 VGB000752 VGC006420 RVGC006838 VGD000242 VGD002204 VGD002206 VGD002208 VGD002224 VGC007005 RVGB000767 RVGC000582 VGC007185 VGD001069 VGD001394 VGD001424 VGD001412 VGD002196 VGD002201 VGD002282 RVGB000119 RVGB000139 VGD000973 VGB002776 VGC000286 VGD000352 VGD000184 VGD000185 VGD002001 VGB002245 VGC000914 VGD001045 RVGC000038 RVGC004131 RVGC000007 RVGC006505 RVGB001266 VGD001871 VGB002166 VGC007762 VGD000891 VGD000948 VGD000958 VGD002092 VGB000991 VGB001000 VGC006952 RVGB000473 VGD002070 VENGO000401 VGB002496 VGB002517 VGD000158 VGD000178 VGC004179 VGD000877 VGD002188 VGD002193 VGD002088 VGD002089 VGD002477 RVGB002254 RVGB000034 RVGB000040 RVGB000040 VGB002151 83206053931 83206053931 AUULS23214222 AUULS23214223 83UL0002021 83205869155 83UL0000036 83UL0000477 83UL0000920 AUULS24438113 83UL0001271 83UL0001349 83UL0001954 83UL0001134 83200509918 83213439592 83203688553 83206054252 83206054265 83207929334 83UL0000102 83213479185 AUULS24277095 AUULS24277094 83UL0001267 83UL0001268 83UL0001739 83UL0002090 83193415284 83205432323 AUULS23300383 83UL0000372 83208327868 83208327809 83215273657 83206054358 83UL0000660 83UL0000661 83UL0000662 83UL0000663 83UL0000659 83206054462 83206054472 83206281789 83206281425 83206527607 83206281749 83206737903 83206737699 83206737809 83206738177 83UL0001866 83UL0001261 83UL0001777 83UL0002128 83UL0002235 83UL0001170 83208327801 83193415088 83205432309 83214411871 83UL0000735 83UL0000281 83UL0000282 83UL0002312 83215374552 83212301843 83206816701 83UL0001827 83206279197 83215237225 83UL0000159 AUULS23205408 83193415406 83UL0000766 83202591911 83UL0000037 83UL0000562 83UL0000398 83UL0000905 83UL0001663 83UL0001666 83UL0002201 83UL0002200 83UL0002199 83208327878 83214411846 83193414972 83193414864 83200509994 83207107548 83UL0002197 83205913347 83UL0000459 83203688407 83UL0000875 83200503817 83205913236 83205913648 83206314190 83206282706 83206282734 83206314387 83194634803 83UL0000801 83UL0000799 83UL0000800 83206279024 83UL0000163 83UL0000724 83UL0000371 83205432193 83UL0000366 83UL0001664 83UL0001617 83UL0000410 83UL0000919 83193415391 83UL000012 83197088391 83UL0001011 83UL0000374 AUULS23222352 83200509094 83212298782 83UL0000773 AUULS24500412 83UL0002011 83UL0001697 83205675117 83205675215 83205675457 83205675539 83UL0001854 83UL0001855 AUULS23292888 83UL0000777 83UL0002022 83UL0002239 83205432220 83206489131 83206411219 83206936004 83206659837 83193415161 AUULS24396630 83205432451 83206489446 83UL0001348 83204007254 83UL0001136 83204006132 83206054247 83206054388 83205651222 83204006828 83213254227 AUULS23300385 83197088193 83UL0000489 83208328429 83208328216 83UL000013 83197088192 83206816438 83206816234 83UL0002099 83UL0002108 83UL0002109 83UL0002110 83200509598 83206816219 83208328188 83UL0000465 83206054337 83207107572 83UL0001410 83UL0001012 83UL0001214 83207483363 83UL0001745 83208327872 83206281314 83UL0001753 83UL0000924 83UL0000847 83UL0000848 83UL0000846 83UL0000148 AUULS23207096 83205651461 83205651423 83205651494 83205651602 83193415239 AH5166US06 AH5090US13 AH4317US08 AH5131US10 AH1413US13 TP7883US08 TP7880U100 TP7883US13 TP7881US32 TP7880US99 TP7880US98 TP7880US97 TP7883US12 TP7880US91 TP7880US96 TP7880US93 TP7880US92 TP7888US16 TP7883US14 TP7883US15 TP7883US16 TP7880U101 TP7880U103 TP7880U106 TP7881US38 TP7880U121 AH5041US05 AH5368US05 AH0907US02 AH1923US26 AH4313US40 AH4313US39 AH4313US41 AH4313US44 AH5407US46 AH5407US47 AH5824US13 ATULS24104541 AH5824US13 AH1000US17 AH1000US18 AH1000US19 AH1000US26 AH1000US25 DEMOUS0538 AH1000US27 DEMOUS0574 AH0132US76 AH1000US29 ATULS23167837 ATULS24281698 AP8871US01 DEMOUS0386 AH0112US14 AH5806US29 AH5806US33 AH5806US34 AH5098US10 AH5098US11 AH5806US37 AH0127U121 AH0127U126 AH0127U127 AH0127U128 AH0127U132 AH0127U134 AH0127U135 AH0127U140 AH0127U139 AH0127U145 AH0127U146 AH0127U148 AH0127U149 AH3210US27 AH2907US11 AH4001U109 AH5135US11 AH4510US15 AH4510US14 AH1912US48 AH5050US26 AH4003US26 AP8978US01 ATULS22883795 AP8727US01 AH5139US03 DEMOUS0606 AH5365US16 AH8217US10 AH0209US07 AP7850US04 AP7850US05 AP7850US07 AH7718US02 ATULS22883825 AH8286US48 AH4660US71 AH1908US68 AH5848U106 VGC006931 VGC006440 VGD000611 VGD001941 VGD002236 VGD002244 VGD002261 VGD002262 BH1072US10 BH1072US11 BH1062US25 BH1062US22 BH1068US05 110019US52 130083US22 772332US39 596320US06 090070US32 090070US40 090070US43 090070US44 110002US07 080092US25 080092US26 030220US42 030020US32 030020US48 030036US26 020144US12 030041US11 050021US05 210036US31 090042US08 190257US15 190257U294 030022US09 030022US10 030022US11 110022US48 190226US26 030082US22 140020US08 080004US11 110040US25 190014US09 030419US23 BEULS24570023 150019US24 150019US25 220826US31 220826US32 220826US47 220826US48 FRULS21261668 030133US33 190747US35 030006US05 090001US07 100029US16 030091US12 160070US06 160070US08 030018US27 030018US30 030092US14 030077US18 030077US19 030259US03 030259US06 060801US19 060801US20 110059US21 190257U116 190033US14 080100US18 030800US42 030078US35 140020US08 110042US31 030028US29 030028US31 030028US34 140021US12 VGOVGB003009 VGOVGC009071 VGOVGB003053 VGC000648 VGOVGB003050 VGOVGC007373 VGOVGB002623 VGOVGB004027 VGOVGB002216 VGOVGB002624 VGOVGC007576 VGOVGC007544 VGOVGC007542 VGOVGC007560 VGOVGC007520 VGOVGC007522 VG0VGB002622 VENVGC007579 VGOVGB002061 VGOVGB002217 VGOVGB002800 6073UL1777 709776VENGO 709637VEN1 709643VENGO1 403452UVENGOR4 403220VENGO 403220VENGO1 403265VEN1 403265VEN1 604254VGO1 250381VENA 604254VGO2 250871VENG 250428VENGO 604682VENUE 250374VENUE1 250374VENUE2 250374VENUE3 250492VENUEUL 250492VENU 250624VENUE 250624VENU 250699VENU3 250985VEN 250960VENUE 604278VEN1 604736VEN1 604874VENG 604874VENG2 604806VEN 905858VENUEGO2 604631VENG1 604806VENGO 604806VENGO1 604875VENGR1 604875VENGR2 604875VGO2 604875VGO1 604875VENGO1 604255VEN 604322VENUE2 250416VENA 250370VEND 604806VENUE 250480VENA 250480VENB 250480VENC 902838VENUE 506452VENUE 418649VENUGOA 418525VENUGO1 418525VENUGO2 418691VENUGO1 418682VENUGO1 418656BVENUGO1 418656BVENUGO3 418656BVENUGO2 418842VENUGO1 418337VENUGO1 418775VENGO 418629VENGO 418724VENGO 418724VENGO1 418724VENGO2 5260319UL11 418724VENGO4 418724VENGO5 418868VENUGO2 418868VENUGO1 418868VENGO3 418868VENGO4 418248VENUGO1 819357VENUGO1 418835VENUGOA 418835VENUGOB 418835VENUGOD 418835VENUGOE 418380VENUGOA 418549VENUGO1 418549VENUGO2 418549VENUGO3 604869VEN 418748VENU 418748R3 819370VENGO 819697VENUGO1 819697VENUGO2 450680VEN 450668VENR3 450668VENU 450668VENU1 514934VENGO 514630VENR41 514630VENGOA 226988VR3 819895VEN 819683VENGO 819647VENGO 819843VENUE 819478VENUGO1 819478VENUGO2 450266VENGO 450266VENUE 450466VENUEG 450466VENGO 504212VENO 450466VENGO1 450759VENGO 450654VENGO 450654VENGOA 450432VENGO 450432VEN11 450432VENGOA 450562VENGO 450562VENGOA 819777VENGO 819623VENGO 819825VENGO 819825VENGOA 819825VENGOA 819764VEN 819764VENU 5291088UL11 604322VENUE 514346VEN1 604806VENUE1 250262VENGO VGD002080 VGD002085 U-HRH-VENGO1 U-HRH-VENGO2 U-HRH-VENGO3 416423ULR4 506623VENUE1 506623VENUE2 506648VENUE 506648VEN1 506648VENR4 519667VENR4 613969VENGO 514252VGO 450468VEN 819964VENGO 418835VENUGOC 705743VENGO 705743VENGO2 705743VENGO3 705743VENR4 416419VENGO 5153965UL11 5153965UL12 5274880UL11 416967VENUE 416480VENGO4 416967VENUE1 416480VENUG 416480VGO1 416480VENGO 416480VENUE 416480VENGO1 416480VENGOR3 416480VENUL 416480VEN2 416480VEN3 416480VEN4 416480VEN5 416480GOR3 416480VEN11 416480VENUR 416480VENUE2 5311944UL11 416480VENGO4 416459VENUE 416967VENGO 416480VENUE1 416226VENUE 416813VEN 416997VEN 416289VENUE 416284VENUE2 416284VENUE3 705362VGO1 807773VENGO1 902894VENUE1 819732VENGO 819732VENU 819732VEN 819287VEGO1 819964VGOA 819964VENGO1 450468VEGOB 450468VENGO 450463VENGO 819228VEGO1 514345VENGO 514345VENGOA 514345VENGOB 905450VENU 514253VENGO 934193VEN1 934193VEN 934193VEN2 514376VENGO 514376VENUEGO 514376VENUE 83UL0001882 514398VENGO 514398VENGOA 514890VENGO 514890VENGOB 514890VENGOA 514890VENGOD 514890VENGOC 514340VENGO 514413VGO 514413VENGOA 514252VEN 514252VEN1 514252VEN2 514252VENR3 514252SCVGO 514242VGOA 514252VGOB 514338VEN1 514338VEN2 514338VEN 5289722UL11 819855VENGO 418663VENUE 819419VENGO 450229VENGO 819326VENGO 450771VENGO 450771VENGOA 450771VENR4 450774VENR3 418268VENUGO1 819536VENGO 819536VENGO2 905858VENUEGO1 819821VENGO2 819821VENGO1 819346VENGO2 819346VENGO 819346VENR3 819674VENGR 819346VEN 819346VENGO1 819346VENUE 819346VENU 450468VEGOA 450746VENGO 450359VENGOB 450359VENGOA 450466VENR3 514362VEN 514362VENA 514362VENGO 780250VENUEGO 705245VEN 250626VG1 604736VEN1 250262VEN 250565VEN1 604795VENR4 604588VGO1 604588VGO2 604988UGOR4 5350168UL56 5350168UL57 5350168UL511 5350168UL52 5350168UL53 519973VENGOA 5350168UL54 5350168UL55 5350168UL58 5350168UL59 5350168UL510 905522VENUE 905522VENUE1 905522VG1 905522VNGR 867393VENUE 082437103224 082437102993 082437060955 082437181094 082437151306 082437151472 082437151471 082437240762 082437240764 082437240763 082437170371 082437142774 082437181019 082437111492 082437191638 082437163503 082437123705 082437123704 082437123706 082437071698 082437092685 082437071697 082437081070 082437251100 082437251149 082437290601 082437290602 082437290611 082437103066 082437103064 082437211606 082437123462 082437103021 082437103020 082437181054 082437251119 082437132669 082437132743 082437133000 082437133001 082437132660 082437132596 082437132603 082437132604 082437123756 082437132668 082437132733 082437132079 082437191741 082437163519 082437042411 082437042313 082437163532 082437163348 082437163349 082437163780 082437163768 082437163420 082437123777 082437240744 082437050566 082437050582 082437050564 082437020593 082437020642 082437221011 082437221000 082437221079 082437181033 082437270699 082437311217 082437311289 082437311348 082437280255 082437092789 082437092622 082437092664 082437092641 082437092642 82437071835 82437151549 82437163845 82437163846 82437163849 82437163808 82437132799 82437132794 82437132793 82437132808 82437132763 82437132764 82437092957 82437270751 82437270734 82437290688 82437280268 82437300268 82437060981 82437042653 82437042686 82437042627 82437042648 82437042629 82437042630 82437042628 82437060986 82437042688 82437123871 82437181156 82437181139 82437201917 82437111678 82436110529 82437111664 82437111665 82437060973 82437163763 82437163851 82437163857 82437163966 82437163707 82437232016 82437232021 82437232024 82437232047 82437232040 82437232044 82437232019 82437232045 82437232033 82437232031 82437232037 82437232035 82437232022 82437232015 82437232038 82437232018 82437232026 82437232030 82437232020 82437232036 82437232041 82437232023 82437232017 82437232032 82437232046 82437232034 82437232048 82437232039 82437232025 82437232029 82437232043 82437232042 82437232028 82437232027 82437123897 82437123902 82437123981 82437251275 82437251207 82437251282 82437251202 82437251208 82437251246 82437251239 82437251250 82437251278 82437211686 82437031589 82437031635 82437103273 82437103274 82437103275 82437103276 82437103272 82437103277 82437103278 82437151511 82437071779 82437071777 82437071764 82437071763 82437071841 82437071842 82437151550 82437151463 82437151490 82437151492 82437151497 82437123914 82437123913 82437280272 82437092956 82437221117 82437221125 82437221104 82437221103 82437221101 82437020652 82437020654 82437020656 82437020655 82437020653 82437191865 82437191867 82437191808 VGOVGD002155 VGOVGC009076 VGOVGC004074 VGDVGD002437 VGOVGC006164 VGCVGC007871 VGOVGC009066 VGOVGC006333 VGOVGC006033 VGOVGC007841 VGOVGC006342 VGOVGD002308 VGOVGC009080 VGOVGC004049 VGOVGB002967 VENVGC000691 VGOVGB002969 VGOVGC007556 VGOVGC007868 VGOVGC007777 VGD000575 HR4106US21 VGB000414 CZ6188US01 VGC006991 VG006992 VG006993 VGDW000786 VGDW000546 VGDW000594 VGDW000595 VGDW000288 VGDW000286 VGC006163 VGD001124 VGD001154 VGD001156 VGD001022 VGD000070 VGDW000880 VGDW000231 CZ6026US35 CZ6026US36 CZ6026US45 CZ6026US46 CZ4010US29 CZ4010US27 CZ4010US26 CZ4010US28 CZ4010US25 VGDW000259 VGDW000291 VGD001184 VGD001075 VGD001086 VGD001091 VGD001092 VGB002324 VGDW000580 VGDW000589 VGD000180 VGD000187 VGD000189 DK1083US96 DK1047U199 DK1047U198 DK1047U200 DK1047U201 DK1047U222 DK1047U256 DK1047U252 DK1047U253 DK1047U257 DK1047U258 DK1047U259 DKULS24157321 DKULS24157324 DK1065US41 DK1065US42 DK9999U184 DK9999U185 DK1065US49 DK1065US56 DK1030US72 DK9999U192 DK1031U117 DK1031U113 DK1031U112 DK1031U114 DK1031U116 DK1031U115 DK1077US58 DK1079US53 DK1079US51 HTN0340 DK1667US01 DK1043US60 DK1043US59 DK1043US76 DK1043US79 DK1043US86 DK1424US13 DK1008US09 DK1066US94 DK1296US28 DK1090US30 DK1090US29 DK1090US32_Swap DK1058US62 DK1058US63 VGOVGC009060 VGOVGC004565 VGOVGC006553 VGOVGC007713 VGOVGC007778 VGOVGC007825 VGOVGC007833 VGOVGC007844 VGOVGC009053 VGOVGC009069 VGOVGD001956 VGDVGD002439 VGOVGD002442 VGCVGC007878 VGCVGC007881 VGOVGC007482 VGOVGC007483 VGOVGC007488 VGOVGC007491 VGOVGC009032 VGDW000331 VGD000457 VGDW000346 85UL0000987 85UL0000986 85UL0000445 85UL0000446 85UL0000447 85UL0000448 85UL0001008 FI1545US02 FI1545US03 FI1630US01 FI1630US02 FI1688US01 FI1688US07 FI1688US02 FI1688US03 FI1280US20 FI1128US80 FI1128US81 FI1128US82 FI1604US01 FI1604US06 FI1604US04 FI1604US08 FI1604US05 FI1604US10 FI1064US18 FI1029US17 FI1029US26 FI1666US02 FI1075US92 FI1075U100 FI1075US85 FI1075US84 FI1075U114 FI1075U106 FI1075U113 FI1014US21 FI1014US19 FI1718US01 FI1015US81 FI1015US87 FI1015US92 FI1015US93 FI1015US94 FI1015US96 FI1015US97 FI1015US98 FI1015U104 FI1109US04 FI1033US23 FI1033US34 FI1001US16 FI1021US31 FI1021US32 FI1021US36 FI1021US35 FI1021US39 FI1034US22 FI1025US40 FI1025US37 FI1025US43 FI1126US13 FI1126US14 FIULS25035338 FI1017U103 FI1017US92 FI1017U104 FI1017US69 FI1017US70 FI1017US71 FI1017US76 FI1017US77 FI1017US95 FI1017US80 FI1003U155 FI1052US32 FI1052US41 FI1000US63 FI1006US17 FI1006US24 FI1545US08 FI1643US03 FI1002US21 VGC004273 FRULS21364448 IMT0044901 VGB001730 VGC007011 VGC007337 VGC007013 FRULS22107587 USM4463220 FRULS21325091 VGC006746 IMT1743406 VGC000859 USM4160417 IMT0002502 FRULS22309996 FRULS23214569 FRULS23247808 FRULS23204610 FRULS20323677 FRULS21302343 FRULS21350558R1 FRULS21327058 VGC004339 IMT3010402 VGC006158 VGC007066 FRULS22138746 U227818307 USM4477401 VGD000032 FRULS22214873 FRULS21327055 USM4164325 USM4164326 FRULS22155718 VGC006516 VGC006523 VGC007238 FRULS21332182 FRULS21757901 FRULS21760169 IMT0056701 IMT0056702 IMT0039302 IMT0004502 FRULS21258015 VGC006858 VGC006864 VGC006910 FRULS23230095 VGC007314 FRULS22106684 VGB002569 VGD000949 VGD000950 FRULS22881698 FRULS23204612 VGB000735 IMT0004604 IMT0052901 IMT0059201 IMT2761505 FRULS22106681 FRULS22106682 FRULS23210933 IMT0268301 IMT1086907 VGC004053 FRULS21239263 IMT0041901 FRULS21332184 FRULS21332185 FRULS22138640 VGC006638 VGC004217 VGC004250 IMT9524814 FRULS21318807 U120781418 US4171818 US4171819 VGC004340 VGC004301 FRULS22254377 FRULS21261237 FRULS21270180 FRULS21270181 FRULS21270183 FRULS21270184 FRULS21256977 FRULS21258030 FRULS21724934 FRULS23204613 FRULS23204611 VGB001714 VGC006155 VGC006353 FRULS22881697 U187868408 VGB000628 VGC000615 VGC007012 VGC004330 VGD000878 FRULS22287881 VGC004153 VGC007490 FRULS22290251 FRULS22290267 FRULS22290075 FRULS22170073 IMT0011002 IMT2844614 VGC006984 FRULS21260043 VGD000024 FRULS21784851 VGC006590 IMT8787603 VGC006247 VGC000603 IMT2847312 GP1000US07 GP1000US08 GP1000US09 GP1000US10 VGB002246 USM4464609 FRULS21314379 FRULS21332183 VGC000719 VGC004219 FRULS21772581 FRULS22189182 FRULS21236971 FRULS21236897 NC1040US07 VGC006872 VGC007659 VGC007661 VGC007666 VGC007667 FRULS22141810 FRULS22141811 USM4487523 VGC006156 U173693062 USM4191713 USM4191714 USM4191715 FRULS22292690 FRULS22293899 FRULS22292697 FRULS22292691 FRULS22293903 FRULS21309276 VGC000735 VGC000849 VGC004286 IMT9524824 FRULS22170666 FRULS22283379 IMT0007408 VGC000605 IMT2838404 IMT2838405 IMT1712309 VGC006871 VGC006899 IMT0051801 US4172620 FRULS21255146 FRULS21260035 FRULS21763508 FRULS22137473 FRULS22137474 VGC007524 FRULS23138413 IMT9190128 IMT9190124 IMT9190126 IMT9190125 IMT9190127 IMT9190130 IMT9190131 VGB000531 IMT0038901 VGC000836 FRULS22881699 FRULS22294763 IMT733415 USM4499109 VGC004267 VGC007300 FRULS22215818 VGC000261 FRULS22217519 FRULS22295087 VGC006845 VGC000712 FRULS21282144 U253977608 U253977606 U253977607 VGC004281 VGC006875 VGC007282 FRULS22218206 VGC006793 VGC006892 FRULS22106685 FRULS21255147 FRULS21250522 IMT9783114 VGC000713 VGC007008 FRULS21724951 IMT0000702 VGC000733 VGC004056 VGB002227 FRULS22106680 USM4360515 VGD000034 FRULS22125894 IMT0043301 FRULS22218127 VGC000706 FRULS21296303 FRULS21296305 GP1021US05 GP1021US06 FRULS21272459 FRULS21272464 FRULS21272466 FRULS21272467 FRULS21273897 FRULS21272460 FRULS21272461 FRULS21272462 FRULS21272463 FRULS21272465 IMT0036102 VGC004214 VGC004311 VGC004345 VGC004409 VGC004410 VGC004413 VGC006944 FRULS21305965 FRULS21306029 FRULS22214874 FRULS21268980 FRULS22213733 FRULS22213732 FRULS21774694 IMT0009201 US62353147 US62353148 IMT8689207 VGC006389 IMT6083605 FRULS21242102 FRULS21242042 FRULS22107032 IMT0050602 IMT0050601 US4461211 VGC004228 FRULS21255187 FRULS22106683 VGC007364 VGC004306 U4217625 VGB000261 VGC000609 FRULS21265758 FRULS21266636 FRULS21265757 IMT8353904 FRULS21174202 FRULS21314713 151057U026 1906476U80 1906476U77 1906476U82 1906476U83 1906476U86 1906476U84 1906476U85 1906476U87 1906477U03 1906477U05 1906477U04 110528U022 110185U052 110185U053 110185U054 110185U057 112001U012 112001U013 804142U028 138880U020 138880U021 1468190U08 1468190U16 110387U073 138209U008 DEULS22863327R1 1389001U02 1389001U05 101098U009 145653U003 110019U014 803415U006 145219U033 3989799U03 1294983U02 3989799U02 1294983U03 3989799U04 145133U012 145133U013 110459U085 110459U089 110459U087 110459U088 110459U086 1873956U10 110513U031 149823U004 149823U005 2584366U02 110478U004 109637U007 109996U036 110012U130 2592998U01 122179U027 800093U023 800093U024 800093U032 800093U039 110574U024 1655450U05 140163U001 104001U014 1810336U20 1810336U18 1810336U23 1810336U24 1810336U25 104001U015 1810336U26 2613170U04 2613170U07 2613170U06 2613170U03 2613170U02 2613170U05 2613170U01 104462U012 104462U013 1078247U21 112073U015 100004U010 110911U039 138403U006 138403U007 112010U043 112013U024 112013U027 112013U026 112013U025 112013U030 104004U008 109005U019 1826731U05 1826731U09 1826731U14 1826731U23 1826731U33 DEULS23227309 DEULS22863216 DEULS24303271 101829U019 120085U006 151145U034 109115U007 127089U014 127089U015 1714896U28 1714896U29 DEMOUS2015 109072U007 2214136U04 2179786U15 100371U015 100371U017 2235853U10 145135U040 145135U041 145135U043 DEULS22863332 145135U045 145135U048 145135U047 145135U046 109006U015 109006U017 123003U024 123003U025 110740U010 1557297U15 350745U015 110480U040 138047U044 138047U049 138047U052 112032U026 112032U029 109060U008 2024214U02 2801740U05 2801740U04 110335U009 110707U005 109007U015 109007U017 109007U021 109007U018 109007U019 109007U020 110032U017 145304U059 145304U064 145304U065 145304U068 145304U070 145304U071 145304U069 145304U073 112024U011 103135U079 103135U087 103135U088 103135U086 103135U095 103135U096 103135U097 103135U099 103135U100 103135U112 103135U113 DEULS23253646 3121932U03 1513970U35 100007U042 100007U043 100000U049 100000U055 DEULS25203866 100000U045 PL3823U408 QA1116US01 DEMOEEU679 DEMO_US294 DEMO_US293 RU9594US22 110450U008 2624092U07 110039U006 1836031U02 104631U022 104631U023 104631U024 104631U021 801048U009 801048U011 801048U013 110045U007 147660U009 147660U011 147660U014 3116196U02 139931U096 139931U094 139931U102 1979063U01 1892918U48 1961127U15 1961127U18 100011U040 100011U042 100011U043 100011U044 100011U046 TP0784U001 TP0040U003 TP0784U011 TP0784U002 110040U033 4576491U03 126631U005 126631U004 126631U006 126631U007 126631U008 122024U056 122024U058 122024U053 122024U060 122024U061 122024U061 122024U063 122024U062 DEULS22863333 122024U086 145881U011 120013U007 120013U008 120013U009 120013U013 120013U014 2524508U01 110644U019 3595063U03 101010U022 2350419U05 2350419U06 145884U016 2001162U01 109529U107 109529U108 109529U112 109529U109 109529U114 142464U039 142464U042 142464U047 145883U008 1567892U25 1567892U27 113023U085 2448139U08 2448139U09 120010U010 2596264U01 110814U010 112079U008 DEULS23223105 138032U009 6885323U02 TP0052U001 110052U056 110052U050 110052U049 110052U048 110052U055 110052U053 110052U051 110052U054 3376913U03 DEULS24813142 117054U015 117054U019 117054U022 2541206U01 138015U014 141114U093 DEULS23260281 122016U047 DEULS23231478 110050U033 110050U032 124008U003 TP0791U030 TP0791U029 TP0791U027 TP0791U005 TP0791U006 TP0791U004 145274U046 145274U048 145274U053 145274U054 138712U017 122038U061 138865U058 138865U054 138865U057 138865U055 2038908U38 138409U013 110320U007 2189240U06 8197422U01 109016U014 109016U015 109016U016 100014U016 142494U008 142494U009 142494U011 142494U010 145539U028 139920U005 139920U004 145287U030 801296U028 801296U029 106989U050 106989U060 104023U019 104023U018 100018U069 802980U010 100019U053 5020808U26 5020808U31 2795540U04 2795540U06 2092156U12 2092156U13 802786U196 802786U202 802786U210 802786U209 802786U230 802786U246 802786U232 802786U247 802786U245 802786U248 802786U253 802786U254 802786U269 DEULS23239241 110775U005 101013U018 101013U017 112099U024 120158U030 DEULS24484353 107006U024 105101U016 107006U040 107006U028 107006U041 107006U044 105101U019 105101U020 107006U047 105101U021 1038511U09 3998259U16 112307U085 112307U083 112307U082 DEULS23227170 1442850U29 1442850U35 1442850U36 1598220U19 1598220U21 1598220U24 1598220U22 1598220U25 1598220U28 110066U021 110066U026 149695U013 149695U015 DEULS22863334 110033U022 120017U029 120017U031 110284U029 110629U011 110629U016 110629U017 2355001U02 109115U007 109115U010 1607646U27 101077U005 101077U006 101077U007 101077U008 1468214U11 122006U147 EK00202U07 104096U007 104096U009 104096U010 104096U011 110230U007 110230U008 110230U010 145525U016 145525U018 145326U017 101019U008 138495U023 120022U014 120022U015 2004770U03 TP4113U023 110987U011 801784U006 801784U007 2583987U07 107041U017 109503U036 109503U037 109901U002 TP1103U039 100432U014 100022U013 2503310U01 138790U004 DEULS22863330 TP7009U001 TP7009U014 TP7009U015 TP7009U019 149535U004 149535U005 149535U006 TP0143U017 TP0143U020 TP0143U022 151308U014 107042U005 122009U139 122009U141 1798068U10 122009U146 122009U152 1798068U31 122009U154 1798068U29 1798068U21 1798068U23 1798068U24 1798068U34 122009U166 2119872U02 TP3007U045 TP3007U052 TP3007U060 5208316U01 5208316U05 2490607U01 3971205U06 120062U021 145840U019 105885U004 151257U006 151257U007 109628U009 109628U010 142491U009 DEULS22863199 DEULS22863200 117052U011 111022U013 117052U012 111022U014 110079U023 110079U026 110079U029 138875U016 4746819U01 110530U010 101057U025 101057U027 101057U029 103041U017 103041U023 103041U022 103041U024 105042U018 105042U019 2413751U07 2413751U05 1685347U63 1685347U58 1685347U65 1685347U66 112037U102 110174U049 110174U046 110174U053 110174U050 110174U048 110174U054 110174U047 110174U051 110174U052 105021U016 105021U014 1531492U06 105021U022 1531492U07 120084U027 112093U022 112093U024 112093U023 2568873U02 2568873U04 2568873U06 2568873U05 2537566U13 122012U093 122012U100 DEULS23293529 1188530U09 1188530U08 1188530U10 1188530U11 1188530U13 110289U006 110289U005 138560U021 138560U023 DEULS23227333 144238U060 120063U024 120063U027 120063U030 2505505U02 110551U025 110551U028 110551U031 110551U032 DEULS23227156 110089U022 138881U004 1847670U14 112040U012 112040U013 109026U009 110899U015 104035U031 104035U038 110091U009 107317U012 117873U012 117873U013 2590353U01 1920702U31 1920702U32 1920702U44 110602U063 110602U064 110602U062 138106U011 138106U007 138106U010 138106U012 138106U008 138106U009 138106U013 110602U080 110602U082 110602U089 110600U024 110600U023 110602U094 110602U095 110602U093 110602U097 110602U074 2038812U28 2038812U29 145694U002 110080U028 7637342U02 7637342U06 765445U004 803328U005 DEULS22863090 149657U083 149657U084 149657U087 149657U088 149657U089 149657U090 149657U092 149657U094 149657U095 149657U093 149657U098 149657U096 149657U099 149657U101 149657U102 149657U106 149657U107 149657U109 149657U112 149657U115 149657U127 149657U133 149657U135 149657U136 149657U137 149657U138 149657U139 149657U140 145222U019 145222U020 145222U021 110436U003 110436U004 149670U050 139941U011 1867796U03 105037U025 142453U064 142453U063 2572605U01 5236077U08 5236077U07 5236077U10 110181U005 110182U033 1698610U02 2703186U07 110604U070 4576491U03 110107U026 110177U023 110177U026 110177U028 120032U089 4081633U02 104037U012 120097U004 110526U020 138035U056 110408U018 104038U006 2139981U42 138330U029 138330U028 112045U009 112045U012 2025865U02 2025865U01 120034U028 120034U035 1937574U32 1937574U33 1937574U44 117034U012 117034U013 117034U014 145410U009 145410U010 9002505U03 145084U029 145084U030 145084U031 DEULS23223085 151010U010 151010U011 151010U012 110197U014 110197U015 110197U019 DEULS23227341 149687U029 140141U061 140141U078 140141U079 140141U080 140141U086 104077U048 104077U047 104077U051 104077U052 9000099U02 110596U027 109834U025 109834U026 112137U014 112137U013 112137U015 112137U018 2103790U35 2103790U46 2600779U01 802008U021 802008U022 138175U020 802501U040 802501U045 802501U046 2605665U01 1460796U21 111182U012 111182U013 111182U017 9000039U02 145423U017 142377U012 2448139U10 2448139U11 3860901U06 120099U007 DEULS22863328 145356U013 109039U008 109039U009 109039U011 109039U012 110499U047 110499U046 110499U058 DEULS24284554 DEULS24284553 145505U055 145505U066 112088U020 1403280U11 1403280U12 2036460U05 104045U015 104045U014 9003434U43 101034U018 DEMOUS1907 DEMOUS2104 DEMOUS2105 DEULS23234591 DEMOUS2169 DEMOUS2475 120041U046 120041U055 110591U007 142454U026 104046U022 104046U028 104046U042 110393U019 110393U020 110393U021 110393U024 112016U016 142478U023 120545U030 120545U029 120545U033 120545U032 104066U031 104066U033 104066U032 104066U037 104066U038 104066U040 104066U041 104066U043 104066U042 104066U045 DEULS23263762 110442U049 110442U054 110442U057 110442U058 110442U066 111000U016 111000U018 112191U047 112191U054 112191U055 117027U030 117027U031 9005880U01 104049U010 2050266U03 2193178U12 2193178U13 2193178U16 2193178U17 2193178U18 2193178U19 2193178U20 2193178U21 2193178U22 2193178U23 2193178U24 2193178U25 2193178U26 2193178U27 2193178U28 2193178U30 151208U009 2193178U31 151165U012 151208U013 151208U014 151208U015 DEULS23222340 1859642U10 104050U013 110162U033 1732389U01 TP9041U026 DEULS23263748 2565728U01 2222813U12 2222813U13 2222813U14 120050U080 DEULS22863111 109725U053 109725U056 109682U007 110520U011 138953U014 138953U017 145358U030 102014U008 109003U011 112058U027 109806U004 110203U035 110203U037 110203U036 110203U039 110203U038 110203U039 110203U042 110203U041 110203U043 110203U044 109045U060 2488204U01 110333U015 142456U016 105060U007 1715002U30 100033U027 100033U026 112600U033 1957871U14 1957871U15 1957871U16 1957871U32 DEULS22863335 4769478U10 1837222U03 110456U057 TP9042U007 TP9042U012 138792U048 138792U055 138792U056 RF2895US01 RU3901US06 RU1000US27 RU1934US36 RU2741US61 RU2741US62 H2006US24 H2075US18 H2009US83 H2026US26 H2077US68 H2077US66 H2086US14 H61141US03 H6114US07 VGD000533 0828371315 0828371286 0828370949 0828371205 0828371206 0828371312 0828371308 0828371320 0828370992 0828371181 0828371338 0828371217 0828371215 0828371170 0828371317 0828371336 828371108 828371203 828371216 828371274 828371277 HU1068US04 HU1069US11 HU00361016 VGC006261 HU1939US07 VGD000842 VGD000787 HU1098US06 VGC006268 HU1103US05 HU1103US04 HU1103US07 HU1103US06 HU1866US01 600016US01 HU1008US19 HU1008US20 HU1008US21 HU1008US22 HU1111US07 HU1008US27 HU1568US05 HU1595US02 600129US08 HU1116US09 HU1116US14 HU1116US15 HU1116US16 HU1221US15 HU1947US05 HU1042US13 HU1137US06 HU1137US07 HU1137US08 HU1137US10 HU1140US01 HU1677US07 HU1677US06 HU1677US03 HU1677US02 HU1018US07 HU1872US02 HU1158US06 HU1158US11 GRULS25005756 GRULS25163334 600010US17 NO4302US67 NO4302US68 NO4302US64 NO4302US69 NO4302U104 NO4302U109 VGDW000087 VGDW000089 VGDW000091 083037829036623 083037609989723 083037024151322 083037066836923 083037825409423 083037820575924 083037826837223 083037829724423 083037829724323 083037826837123 083037828073223 083037826872421 R83037863803723 D83037823082122 D83037824722123 D83037826692123 083037219144623 083037847558524 D83037824752123 D83037829512124 D83037828962124 D83037829522124 083037826872521 083037049412323 083037809627224 083037809498324 083037800908722 083037802445324 083037808505823 083037808213723 083037800603922 VGB001320 VGB001411 083037801225124 083037843337023 083037841929924 VGB001303 VGC007396 083037843842323 083037005562421 083037068674924 083037848686023 083037843833723 083307823001723 083037845573622 083037202254722 083037283175722 083037022980523 083037043721623 083037043940322 083037042464121 083037241075823 083037048824021 083037047462021 083037047462121 083037066745221 083037603532023 083037602465723 083037020183824 083037027299522 083037027299422 083037021836024 083037004280623 083037066857023 083037208958523 083037201534324 083037207858920 083037206608723 083037820556821 083037213111621 083037066807521 083037006335823 083037006335923 083037007431623 083037001540421 083037002407923 083037801843521 083037802064524 083037802195824 083037802064624 083037802195024 083037802195924 083037802195124 083037211846820 083037211847420 083037211847920 083037211847720 083037210679323 083037847038724 083037100687323 083037842359224 083037809571723 083037609300124 083037068804022 083037068804922 083037068804822 083037068804122 083037209631223 083037209631023 083037864199724 083037822301624 GIB_C_UVERSE_VGD002167 D83037821032121 7982UL3822 7982UL3823 7982UL0892 7982UL0967 7982UL2040 7982UL2039 7982UL2060 7982UL3055 7982UL3058 7982UL2911 7982UL4319 7982UL4318 7982UL4457 7982UL4618 7982UL4619 7982UL4617 7982UL4459 IDULS54212611 7982UL1083 VGC000037 VGD001816 VGD000962 VGD000463 IE4205US04 VGB001339 VGDW000163 VGDW000336 VGB001752 IE4282US01 VGDW000113 VGD000505 VGD000867 VGD001096 VGDW000092 VGDW000093 VGDW000102 VGDW000450 VGD000266 VGDW000190 VGC007200 VGC007201 IE4256US11 IE4256US10 IE4256US09 IE4256US08 IE4256US07 IE4256US06 VGC007199 IE4267US01 IE4174US09 VGD000301 VGDW000034 VGDW000389 VGDW000889 VGD001247 VGD000192 VGD000439 VGD000448 IE4173US09 IE4051US01 VGDW000322 VGDW000391 VGDW000627 IE4053US14 IE4053US13 IE4053US12 IE4053US11 VGD000360 IE4066US15 IE4066US13 VGDW000614 IE4283US07 VGD000872 VGD000874 VGDW000239 VGDW000242 VGDW000252 VGDW000266 VGDW000243 VGDW000226 VGDW000262 VGDW000238 VGDW000877 VGDW000251 VGDW000230 VGDW000295 VGDW000250 VGDW000232 VGDW000246 VGDW000249 VGDW000270 VGDW000227 VGDW000241 VGDW000261 VGDW000247 VGDW000244 VGDW000240 VGDW000878 VGDW000248 VGDW000235 VGDW000233 B5368644 B5368643 ITULS24491124 ITULS22258265 ITULS22258266 A5400807 A5117927 ITULS22264727 ITULS21459546 ITULS22216042 ITULS23247782 A215175012 A51816100 B5258026 A58122133 A5291613 A5291612 A239619401 A6400303 ITULS23238277 A5356279 A5300743 A5356278 A51014170 A5356284 ITULS22268480 ITULS22280612 A5645342 A5645345 A5304842 A5353850 A6233905 A5110193 A5624156 B5127447 ITULS24214312 A5189381 A121831225 A121831233 ITULS22208766 A5183679 B5353610 A202090814 A51026158 A5189750 A6215817 A6215818 A6215824 A6215823 A6215822 A5130634 A5354116 A5130633 A6215825 A5354120 B51051104 A51051163 B51051103 B51051105 A5104208 ITULS23221634 VNU000980 VNU001578 VNU000505 VNU000795 VNU001699 VNU001176 VNU001673 VNU000813 VNU000682 VNU000965 VNU000972 VNU000262 VNU001289 VNU001026 VNU001652 VNU001573 VNU001198 VNU001024 VNU000885 VNU000775 VNU000764 VNU000933 VNU001008 VNU001369 VNU001368 VNU000059 VNU000249 VNU000157 VNU001006 VNU001639 VNU000909 VNU001222 VNU001670 VNU000888 VNU001425 VNU001414 VNU000591 VNU001453 VNU001452 VNU000768 VNU000765 VNU000422 VNU000535 VNU001241 VNU001240 VNU001558 VNU000940 VNU000228 VNU001559 VNU000719 VNU000587 VNU000679 VNU001147 VNU000730 VNU001535 VNU000350 VNU001207 VNU001128 VNU001127 VNU000959 VNU000503 VNU000557 VNU000142 VNU001429 VNU000799 VNU000920 VNU000384 VNU000501 VNU000698 VNU000559 VNU000826 VNU001370 VNU001409 VNU000989 VNU001001 VNU001000 VNU001136 VNU000914 VNU001031 VNU001348 VNU000979 VNU000624 VNU000625 VNU000879 VNU001706 VNU000057 VNU001050 VNU000709 VNU001213 VNU000442 VNU000346 VNU001352 VNU000352 VNU000360 VNU000367 VNU001148 VNU000495 VNU001350 VNU000634 VNU001377 VNU000323 VNU001199 VNU001374 VNU001297 VNU001085 VNU000632 VNU000629 VNU000628 VNU001186 VNU000211 VNU000210 VNU001003 VNU000258 VNU001431 VNU000924 VNU001599 VNU001135 VNU000982 VNU000229 VNU000919 VNU000876 VNU000096 VNU000777 VNU001322 VNU000903 VNU001081 VNU001234 VNU001405 VNU001467 VNU001589 VNU000944 VNU000320 VNU001701 VNU001721 VNU001337 VNU000736 VNU001079 VNU000363 VNU000770 VNU000351 VNU000276 VNU001719 VNU000803 VNU001346 VNU000244 VNU001175 VNU001408 VNU000435 VNU000630 VNU000871 VNU001383 VNU000192 VNU000668 VNU001264 VNU000890 VNU000216 VNU000215 VNU000880 VNU001113 VNU000897 VNU000976 VNU000443 VNU001039 VNU001255 VNU000056 VNU001697 VNU000095 VNU000623 VNU001257 VNU001179 VNU001364 VNU001365 VNU001574 VNU000284 VNU001152 VNU000603 VNU001532 VNU000872 VNU001132 VNU001134 VNU001107 VNU001219 VNU001163 VNU001182 VNU000812 VNU000343 VNU000635 VNU001082 VNU000509 VNU000943 VNU001183 VNU000684 VNU000710 VNU001042 VNU000701 VNU000699 VNU000985 VNU001083 VNU000254 VNU000641 VNU000070 VNU001367 VNU000441 VNU001002 VNU001144 VNU000327 VNU001517 VNU000830 VNU000961 VNU001009 VNU001607 VNU001338 VNU000372 VNU001341 VNU000242 VNU000260 VNU000265 VNU000642 VNU000761 VNU000790 VNU000419 VNU000420 VNU000726 VNU001005 VNU000998 VNU001010 VNU001027 VNU001032 VNU001028 VNU001076 VNU000906 VNU000939 VNU000541 VNU000588 VNU001674 VNU001209 VNU000817 VNU000780 VNU000773 VNU000772 VNU000781 VNU000787 VNU001048 VNU001633 VNU000155 VNU001345 VNU001598 VNU000793 VNU000794 VNU001049 VNU000863 VNU000864 VNU000921 VNU000785 VNU001313 VNU000784 VNU000626 VNU001052 VNU001030 VNU001045 VNU000594 VNU000063 VNU000449 VNU001530 VNU001557 VNU000536 VNU000932 VNU001433 VNU000881 VNU001211 VNU000913 VNU000737 VNU000199 VNU000439 VNU000232 VNU000240 VNU001342 VNU001466 VNU000304 VNU001153 VNU001078 VNU001565 VNU000788 VNU001033 VNU001047 VNU000237 VNU001644 VNU000261 VNU000264 VNU000667 VNU001375 VNU001536 VNU001531 VNU001131 VNU001130 VNU000712 VNU001168 VNU001109 VNU001120 VNU001200 VNU001614 VNU001654 VNU000987 VNU001388 VNU001306 VNU000905 VNU000341 VNU000783 VNU001029 VNU001641 VNU000269 VNU001354 VNU000480 VNU000916 VNU001181 VNU001430 VNU001210 VNU000263 VNU000500 VNU000734 VNU000486 VNU000544 VNU000485 VNU000580 VNU000502 VNU000912 VNU000911 VNU001191 VNU001193 VNU001194 VNU001192 VNU000507 VNU000558 VNU000738 VNU000421 VNU000735 VNU001254 VNU001411 VNU000633 VNU000650 VNU001262 VNU001274 VNU000941 VNU000581 VNU001123 VNU001232 VNU001314 VNU001315 VNU000212 VNU000816 VNU000602 VNU000662 VNU000644 VNU000197 VNU001263 VNU000883 VNU000991 VNU000822 VNU001261 VNU000430 VNU000533 VNU001146 VNU000213 VNU000583 VNU000585 VNU000806 VNU000808 VNU000807 VNU001385 VNU000600 VNU000867 VNU001051 VNU000270 VNU000610 VNU000574 VNU001428 VNU001043 VNU000094 VNU000103 VNU000771 VNU000829 VNU001044 VNU001046 VNU001154 VNU000301 VNU001142 VNU001173 VNU000060 VNU001004 VNU000767 VNU001145 VNU000220 VNU001634 VNU000995 VNU000639 VNU000868 VNU000865 VNU001124 VNU001413 VNU001406 VNU001651 VNU001205 VNU000538 VNU000779 VNU000762 VNU001189 VNU000152 VNU000546 VNU000671 VNU000448 VNU001214 VNU001454 VNU000918 VNU000285 VNU001118 VNU000326 VNU001353 VNU001129 VNU000151 VNU001412 VNU000093 VNU001561 VNU000064 VNU000776 VNU001528 VNU000470 VNU000809 VNU000800 VNU000499 VNU000805 VNU000498 VNU001196 VNU001290 VNU001296 VNU000251 VNU000293 VNU000922 VNU000974 VNU000137 VNU000221 VNU000445 VNU000561 VNU000584 VNU000700 VNU000578 VNU000778 VNU000825 VNU001575 VNU001579 VNU001180 VNU001349 VNU001363 VNU001133 VNU000640 VNU000359 VNU000365 VNU000763 VNU000065 VNU001576 VNU001529 VNU000296 VNU001465 VNU000981 VNU000884 VNU001184 VNU001653 VNU001220 VNU001606 VNU000158 VNU000636 VNU001221 VNU000627 VNU001344 VNU001622 VNU000316 VNU000992 VNU000354 VNU001217 VNU001216 VNU001037 VNU001188 VNU001055 VNU001710 VNU000678 VNU000769 VNU000663 VNU000877 VNU001298 VNU001166 VNU001150 VNU000997 VNU001613 VNU000595 VNU000596 VNU001384 VNU001427 VNU000666 VNU000058 VNU001284 VNU001177 VNU001208 VNU000827 VNU001305 VNU001253 VNU001351 VNU001187 VNU001432 VNU000468 VNU000466 VNU000467 VNU000482 VNU000450 VNU000259 VNU000957 VNU000233 VNU000576 VNU000661 VNU001636 VNU001635 VNU001141 VNU001056 VNU001231 VNU001283 VNU001282 VNU000760 VNU001708 VNU001707 VNU000069 VNU000902 VNU000643 VNU001560 VNU000153 VNU000218 VNU000984 VNU001287 VNU001373 VNU001260 VNU001723 VNU000593 VNU000592 VNU000651 VNU000374 VNU000506 VNU000733 VNU000935 VNU000234 VNU000288 VNU000523 VNU000721 VNU001167 VNU001169 VNU000732 VNU000504 VNU001426 VNU000556 VNU000714 VNU000328 VNU000329 VNU000340 VNU000349 VNU001038 VNU001258 VNU001259 VNU001464 VNU001468 VNU000548 VNU000717 VNU000718 VNU001256 VNU001356 VNU000647 VNU001299 VNU001415 VNU001424 VNU001422 VNU001423 VNU000983 VNU001386 VNU000067 VNU000792 VNU000786 VNU000348 VNU000539 VNU000547 VNU000268 VNU001054 VNU001053 VNU000575 VNU001533 VNU001698 VNU000915 VNU000927 VNU000638 VNU001215 VNU000821 VNU000196 VNU000214 VNU000300 VNU000579 VNU000573 VNU000834 VNU001035 VNU000774 VNU000936 VNU000928 VNU000713 VNU000882 VNU001119 VNU000937 VNU000938 VNU000564 VNU001580 VNU001581 VNU001174 VNU000294 VNU000542 VNU000731 VNU000356 VNU000878 VNU000465 VNU000782 VNU001608 VNU000361 VNU001265 VNU000549 VNU001281 VNU000241 VNU000277 VNU000766 VNU001206 VNU001577 VNU000958 VNU000238 VNU000487 VNU001058 VNU001212 VNU001347 VNU000236 VNU000969 VNU001285 VNU001286 VNU000832 VNU000831 VNU000970 VNU001084 VNU000315 VNU000342 VNU000138 VNU000219 VNU000373 VNU000434 VNU001143 VNU000545 VNU001204 VNU001366 VNU000198 VNU001463 VNU001034 VNU001266 VNU000431 VNU000956 VNU001534 VNU000543 VNU001629 VNU000297 VNU000689 VNU001672 VNU000810 VNU001040 VNU000489 VNU001273 VNU001218 VNU000429 VNU000696 VNU001572 VNU000444 VNU000697 VNU000366 VNU000904 VNU000866 VNU001022 VNU001185 VNU000789 VNU000669 VNU000665 VNU001077 VNU000245 VNU000250 VNU001267 VNU000426 VNU000681 VNU000129 VNU000488 VNU000148 VNU000870 VNU000209 VNU000568 VNU000570 VNU000071 VNU000988 VNU000266 VNU000061 VNU000122 VNU001057 VNU000702 VNU000121 VNU000563 VNU000062 VNU001025 VNU000589 VNU000066 VNU000990 VNU000149 VNU000124 VNU000123 VNU000130 VNU000540 VNU001371 VNU000322 VNU000508 VNU001410 VNU001407 VNU000267 VNU000331 VNU000804 VNU000802 VNU000973 VNU001233 VNU001696 VNU001162 VNU000582 VNU000923 VNU001355 VNU000966 VNU001080 VNU000645 VNU000469 VNU000999 VNU000217 VNU000572 VNU000364 VNU000357 VNU000801 VNU000910 VNU000931 VNU001230 VNU001720 VNU000874 VNU000873 VNU001190 VNU001588 VNU000156 VNU000425 VNU000510 VNU000200 VNU000996 VNU000330 VNU000869 VNU000820 VNU001621 VNU000154 VNU001041 VNU000227 VNU001202 VNU000683 VNU001671 VNU000986 VNU000796 VNU000239 VNU001711 VNU001712 VNU001714 VNU001716 VNU001715 VNU000646 VNU001700 VNU001201 VNU000637 VNU001178 VNU001516 VNU000798 VNU001007 VNU001023 VNU001709 VNU000934 VNU001339 VNU001340 VNU000942 VNU000308 VNU000306 VNU001372 VNU001387 VNU000105 VNU001203 VNU000551 VNU000550 VNU000447 VNU000907 VNU000235 VNU000472 VNU000483 VNU000484 VNU000068 VNU000355 VNU000097 VNU000303 VNU000791 VNU000275 VNU000347 VNU000358 VNU000715 VNU000716 VNU000720 VNU000797 VNU000586 VNU001288 VNU001195 VNU000889 VNU000128 VNU000321 VNU000362 VNU001036 VNU001108 VNU001556 VNU000908 VNU000305 VNU000971 VNU000926 VNU000670 VNU000150 VNU000104 VNU001515 VNU001615 VNU000590 VNU001597 VNU000195 VNU000481 VNU000685 VNU001112 VNU000302 0850379313 0850379573 0850377901 0850376670 0850378625 0850379351 0850400601 0850378276 0850377250 0850400075 0850377635 0850400272 0850379206 0850400175 0850400229 0850378928 0850377844 0850378812 0850379336 KRULS54508721 0850377822 0850378679 0850378841 0850402923 0850377709 0850400103 0850400619 0850377360 KRULS54839261 0850400295 0850379533 0850402461 0850375844 0850378816 0850377019 0850402391 0850377133 0850377755 0850378631 0850378854 0850379488 0850402474 0850378391 0850379234 0850378148 0850400123 0850403167 0850379318 0850376919 0850378998 0850402989 0850402712 0850378490 0850400501 0850378389 0850402765 0850403135 0850376595 0850377605 0850400420 0850402340 0850401068 0850402261 0850402694 0850403361 0850401139 0850377160 0850378102 0850401330 0850402776 0850377651 0850378803 0850379290 0850379523 0850400293 0850377684 0850401362 KRULS54270041 0850401863 0850401778 0850376725 0850378170 0850378182 0850378179 0850378181 0850378180 0850400434 0850402312 0850376218 0850379921 0850378851 0850376554 0850401042 0850401759 0850379359 0850403146 0850377915 0850401606 0850402240 0850379190 0850378704 KRULS54244211 0850379009 0850402212 KRULS55190451 0850401371 0850377079 0850378939 0850400153 0850377429 0850377482 0850401184 0850378968 0850379247 0850402313 0850402670 0850402882 0850378035 0850379242 0850377070 0850378591 0850377009 0850377812 0850379341 0850401820 0850377351 0850377648 0850378097 0850378509 0850400101 0850401426 0850401830 0850402889 0850377087 0850378912 0850402349 0850376893 VGC006801 KW1055US36 KW1055US39 VGD001539 KW1001US65 KW1001US64 KW1008U118 KW1164US02 VGD002146 VGB001245 KW1019US12 KW1019US13 KW1019US14 KW1019US15 KW1018US65 KW1018US66 VGD002153 VGB002828 VGB002897 VGC000473 LB1050US11 LB1311US02 LB1359US06 LB1018US06 VGD001720 VGD001734 VGD002333 VGD000828 080017US13 080017US30 080017US31 080017US38 080017US39 080017US32 080017US67 120095US12 120095US13 030013U104 030013U113 030013U105 110807US17 110807US16 110807US21 110807US25 1553UL1485 1553UL1811 1553UL1815 1553UL1814 1553UL2142 1553UL1727 1553UL2599 1553UL2311 1553UL2399 1553UL2343 1553UL2568 1553UL2281 1553UL2282 1553UL1953 1553UL1816 1553UL1812 1553UL1813 1553UL1979 1553UL1787 1553UL1802 1553UL2209 1553UL1955 1553UL2224 1553UL2108 1553UL1954 1553UL2235 1553UL2236 MYULS24516207 1553UL2210 1553UL2283 1553UL2250 1553UL2575 1553UL2398 1553UL2400 1553UL2141 1553UL2638 1553UL2550 1553UL2363 1553UL1817 1553UL1877 1553UL2011 1553UL2249 1553UL2248 1553UL2251 1553UL2306 1553UL2277 1553UL2279 1553UL2278 1553UL2401 1553UL2427 1553UL2512 1553UL2511 1553UL2513 1553UL2598 1553UL2560 1553UL2597 1553UL2596 1553UL2626 1553UL1514 1553UL1533 1553UL1534 1553UL1535 1553UL1709 1553UL1708 1553UL1980 1553UL1952 VGDW000309 VGOVGB004147 VGOVGB004136 VGOVGB004129 VGOVGB004133 VGOVGB004145 VGOVGB004071 VGOVGB004128 VGOVGB004074 VGOVGB004025 VGOVGB004132 VGOVGB002859 VGOVGB004149 VGOVGB004067 VGOVGB004140 VGOVGB004141 VGOVGB004077 VGOVGB004064 VGOVGB004146 VGOVGB004143 VGOVGB004142 VGOVGB004144 VGOVGB004101 VGOVGB004114 VGOVGB004094 VGOVGB004115 VGOVGB004070 VGOVGB003018 VGOVGB004130 VGOVGB003075 VGOVGB004112 VGOVGB004118 VGOVGB004072 VGOVGB004065 VGOVGB004066 VENGO002090LO VGOVGB004076 VGOVGB004069 VGOVGB004073 VGOVGB004119 VGOVGB004075 VGOVGB004137 ALK01415 ALK01416 ALK01417 DHD1449 DHD14400 AMF02400 AMF02401 AMS03463 AMS03492 AMS0643 APL02442 APL02447 APL02445 ZTP01410 ZTP01409 ZTP01408 APL02448 APL02446 APL02464 BVW01408 BVW01409 DHG09419 DHG09427 DHG2540 DHG2541 DHG2542 DHG2543 DHG2544 DHG2545 DVT01452 DVT01453 DVT01459 DVT01457 DVT01466 DRK01413 DRK01421 DRK01422 EMM0740 EMM01450 HDW02445 HDW02450 HDW02454 HTN0341 LDN014055 LDN014054 LDN014061 LDN014058 LDN014062 LDN014063 LDN014070 LSD06406 LSD06408 HOR0246 HOR0247 HOR0249 PRD0144 RRM01406 RTD4042 TLB0740 UTR04419 UTR04421 UTR3047 UTR3048 UTR05408 UTR01469 UTR05402 85209459696 85209460525 85209460010 85209459988 85UL0001176 85UL0001208 85195399066 85UL0002225 85UL0001034 85UL0001828 85215765546 85UL0000873 85UL0000734 NG1868US11 NG1868US12 NG1868US10 NG1868US09 NGULS21743948 NGULS21743949R1 NG1087US04 NGULS22271979 NG1012US13 NG1098US03 NG1114US07 VGD000089 NG1132US01 NO4064US40 NO4064US42 NO4730US02 NO4011U253 NO4011U194 NO4011U195 NO4011U196 NO4011U207 NO4011U302 NO4011U304 NO4011U300 NO4011U301 NO4011U303 NO4011U299 NO4070US93 NO4070U102 NO4070US73 NO4070U108 NO4586US01 NO4102US14 NO4102US13 NO4136US15 NO4136US16 NO4136US17 NO4004US73 NO4004US63 NO4406US18 NO4018US53 NO4018US54 NO4060US24 NO4060US33 NO4107US20 NO4107US23 NO4029US19 NO4087US34 NO4087US62 NO4087US52 NO4087US51 NO4087US53 NO4012US43 NO4012US55 NO4051US22 NO4051US24 NO4391US04 NO4014U103 NO4014US91 NO4037US54 NO4037US55 NO4016U133 NO4016U132 NO4016U135 NO4016U154 NO4016U152 NO4016U157 NO4016U156 NO4005U159 NO4005U164 NO4005U170 NO4005U169 NO4005U225 NO4005U224 NO4005U189 NO4005U176 NO5005U216 NO4005U179 NO4005U185 NO4005U188 NO4005U197 NO4005U198 NO4005U199 NO4005U200 NO4005U201 NO4016U125 NO4016U179 NO4016U130 NO4016U129 NO4016U175 NO4016U176 NO4740US04 VGB000584 VGB001506 VGB002213 VGC004426 VGC004437 NO4302US78 NO4009U110 NO4009U114 NO4009U121 NO4009U122 NO4009US92 NO4000US62 NO4000US63 NO4068U147 NO4068U123 NO4068U146 VGD001723 VGD001744 VGD001787 VGC007174 VGC007175 VGC007176 VGC007177 OM1007US61 VGC004202 VGC006356 VGB000298 VGOVGC007313 VGOVGC006205 1557UL2056 1557UL1029 1557UL1665 1557UL1751 1557UL1782 1557UL1872 1557UL1871 1557UL2068 1557UL2229 1557UL2292 1557UL2293 1557UL2291 1557UL1783 1557UL1254 1557UL1253 1557UL1576 1557UL1033 1557UL1646 1557UL1645 1557UL1750 1557UL1781 1557UL1238 1557UL1647 VGOVGD001861 VGOVGD001866 VGOVGD001867 PL1398US17 VGDW000300 VGDW000392 VGDW000396 PL0302US08 PL0302US07 PL8869US02 PL1656US01 PL8907US03 PL1494US05 VGD000774 VGDW000303 PL8973US03 VGDW000302 PL2031US08 VGDW000375 VGDW000384 VGDW000385 VGDW000428 PL9106US01 PL3823U258 PL3823U500 PL1147US04 VGD000769 PL3823U357 PL1139US04 VGDW000174 PL1316US03 PL1316US06 PL1056US07 600062US09 600062US11 VGD000053 600042US41 VGD000281 VGC000927 VGC004054 VGC004325 VGC004178 VGDW000401 VGC007648 PL9269US01 VGC007740 VGC000888 VGDW000783 VGD001213 PL3823U410 PER43262 PER18435 PER18436 PER09115 PER09117 PER18707 PER59110 PER28526 PER00777 PES05837 PER16919 PER16919 PER03013 PER03015 PER19033 PER05212 PER25636 PER34947 VGB001129 VGB001131 VGB001136 VGC000141 VGC000150 VGC000253 VGD001746 VGD001784 VGC004095 VGC004104 VGC004106 QA1145US09 QA1145US08 QA1145US07 QA1145US06 QA1145US05 QA1145US04 QA1145US03 QA1145US02 QA1145US01 VGD000476 VGDW000218 RO5551US01 VGDW000195 VGDW000292 VGD000183 RO4801US04 VGC004189 RO4826US04 RO4826US10 RO4826US11 VGD000027 RO5631US02 RO4030US19 RO4030US18 RO4030US21 RO4030US20 RO4030US22 RO4030US17 RO4932US03 VGD000947 VGD001175 RU6261US19 VGC006843 VGD001433 SA1113US03 SA1060US27 SA1280US09 SA1280US10 VGD002489 SA2161US37 VGD001936 SA1114US43 SA1114US44 SA1114US45 SA1114US46 SA1138VJ02 VGD001754 VGD001757 VGD001759 VGD001761 VGD001574 VGD001577 VGD001579 VGD001587 SA1002US19 VGD002028 SA2660US64 VGD000778 SA1031US29 SA2415US09 VGD001556 VGD001570 VGC004234 VGC004247 SA2191US58 SA2454US21 VGD002492 VGD001588 VGD001589 VGD001590 VGD001855 SA1056US69 VGD001717 VGC004246 SA1054US84 VGD001727 VGD001860 VGD001947 SA2860US11 VGB000305 6073UL1371 6073UL1754 6073UL0870 6073UL1328 6073UL1886 6073UL1755 6073UL1661 6073UL1683 6073UL1684 6073UL1375 6073UL1372 6073UL1373 6073UL1686 6073UL1687 6073UL1827 6073UL1374 6073UL1946 6073UL1762 6073UL1889 6073UL1912 6073UL1928 6073UL1761 6073UL1877 6073UL1878 6073UL1879 6073UL1881 6073UL1882 6073UL1880 VGD000408 VGDW000399 VGD000796 VGD000812 VGC006887 VGDW000378 SI4142US05 VGC007435 VGDW000747 SI4073US07 VGDW000761 SI4056US66 SI4056US67 SI4056US77 SI4056US76 SI4056US86 SI4056US85 SI4056US84 VGB002295 SI4055US33 SI4055US39 SI4043US09 SI4125US05 ZA3364US01 ZAULS22203171 VGC006880 ZAULS21261635 ZAULS21261640 ZAULS22203176 ZAULS22203170 ZAULS22271942 ZAULS22271943 ZAULS22271933 ZAULS22271931 VGC007572 VGC007649 ZAULS21261636 ZAULS22223105 ZAULS22223098 ZAULS22223099 ZAULS22223104 ZAULS22223101 VGC000883 ZA3371US01 ZA2958US09 VGC006234 ZAULS22271930 ZAULS21257155 VGD000397 ZAULS21261638 ZAULS20644146 ZA1937U106 VGC007209 VGC007394 VGC007407 ZAULS22271932 ZAULS21240650 ZAULS21768721 ZAULS22203172 ZAULS21261639 ZAULS21231643 ZAULS21231642 ZAULS21261637 ZAULS21327404 DE188455 DPN21725 DPN21728 FEG49201 NPV39112 CEZ04117 FPG30621 MPD29211 MPD29210 MPD27003 MPX18505 MPX82267 MPX82270 MPX30237 MPX30240 NP100029 MP119305 FED06400 FPG86489 FPG86490 FED15100 DPP06600 NPD04200 FED14700 DEM52911 DEM52916 FEB520232 DEN00840 DEN00841 DEN00856 UP506A49 FPD31900 FED02705 DEL264133 DEN06344 CPZ62700 CP223924 CP223929 FED06300 FEG84901 FEG68705 FP943121 FEC800354 FEC800353 DPN49408 ME5043179 FED14400 FED14500 FED14100 FED14800 FED14000 FED14200 FEG46163 FEG46164 FEG46175 FEG46189 FED14300 FE387828 FE387842 FE387843 FE387844 MEX70129 FED09500 FED09600 FE421182 FED15200 GPX40901 GEC21870 GPX36912 FED14600 MP839903 MPX83504 MPX83505 FPD10816 FPD10817 DPC79017 DP418876 DP418886 DP418888 C99999388 S99999762 DM000004 DM000005 DM000006 DM000007 DED00100 DED00101 DED00102 DED00103 DE492A70 DE492A98 DE492A104 DEH13595 DEH13598 DE6588168 DE6588167 DE6588166 DE6588174 DE6588177 DE6588178 DE6588179 DE6588180 DPA97314 FPG98442 FPA99440 GPX24132 GPX24135 FEF70570 FEE93839 FEE93838 FEE93840 FEE93835 FEE93836 FEE93837 FEE93834 DEN04527 FED55A70 NP532A36 MPX35409 CEG94665 CEG94667 MP599332 MPX39936 FEG44905 FPG23631 UPW17538 ESULS24347879 ESULS24347880 ESULS24347931 EE266747 EE238586 DEN01127 CE165374 DPN886110 FEG35066 FEG35067 FEG35063 FEG35065 FEG35062 FEG35061 NE748632 FED14900 FEG37515 MPX92514 MPX92515 CPZ70600 MEX34424 FED15000 MP842051 MP842055 MP842058 MP842059 GE448076 UE535A184 UE535A192 FP182A23 FPG30569 FE457A162 FE457A203 FE457A205 FE457A204 FEE084117 FEE084118 FEE084114 FEE084123 FEE084101 FEE084122 FEE084120 FEE084110 FEE084116 FEE084113 FEE084115 FEE084111 FEE084108 FEE084124 FEE084125 FEE084119 FEE084112 FEE084104 FEE084103 FEE084102 FEE08499 FEE084107 FEE084100 FEE084105 FEE084106 FEE084109 FEE084121 CEK53848 CEK53857 CEK53865 GPX51402 GPX51413 ESULS23967003 DEH10743 DPN92112 DPN11016 DPN11021 DPN11019 DPN11020 DPN11018 DPN11017 DEN04414 DEN04409 NE185083 NE185094 NEV439237 NEV439274 NPV15998 NE219669 CEE156156 CEZ00333 FEG35068 FEG35064 UP389041 CE671340 VGB002827 L007US16 U026US01 A004U109 A004U116 A004U117 A004U118 A004U127 A004U125 A004U126 A004U130 F003US57 F003US58 C003US10 BD007US20 BD007US26 X001US57 X001US53 X001US55 X001US54 X001US56 X001US59 X001US64 SE9999U250 O005U145 O009U162 O009U179 N001US97 L005US05 M007US46 M007US48 M007US58 M007US70 X006US27 X006US28 X022US01 F001US87 F001US88 K002US70 K002US94 K002US93 K002US97 S001US55 S001US57 D002US07 U002US07 U016US02 G013US01 SE9999U275 E001US134 E001US157 BD003US54 BD003US61 BD003US62 M001US64 M001US65 M001US72 M001US73 M001US74 M004US71 M004US80 M004US84 W003US14 E003US34 E003US35 E003US36 E003US37 A008US17 D003US31 T002US58 T002US60 T002US59 T002US69 T002US61 T002US62 T002US63 T002US76 T002US77 T002US80 Y001US20 SE9999U276 SEULS24137218 A005US10 SEULS24137137 A007U111 A007U110 SE9999U237 A007U161 A007U160 A202US02 A203US07 A203US08 A203US10 SEULS24137219 S003US08 C001U107 C001U108 C001U118 C001U147 F004US30 U001US54 G001US47 I001US28 I001US27 I001US26 M009US12 529010US76 529250US16 504020US16 504020US19 529020US68 168650US01 492861US06 437918US84 437918US87 547020US13 547020US14 547020US15 502010US71 502010US70 502010US75 502010US84 317010US07 605010US39 547010US15 532063US03 475230US52 475230US54 568676US01 529050US48 529050US50 529050US53 730160US10 121383US31 121383US33 529060US22 121383US39 985462US04 475200US78 475200US80 475200US85 475200US86 DEMOUSU425 547050US30 541010US14 547090US23 547050US33 541010US16 547050US36 541010US19 547050US38 547090US34 547040US04 760813US02 760813US01 CHULS22858696 169655US01 168739US01 167513US01 697040US20 697040US21 697010US28 697010US27 482040US01 475070US18 722090US07 679352US04 562020US20 562020US28 847040US12 847040US13 121512US29 121512US35 730052US01 730050US19 475380US07 CHULS22858652 529140US59 482896US07 CHULS23227653 CHULS23409973 529150US94 529150U113 529150U114 529150U103 529150U104 529150U102 529150US99 529150U105 529150U106 529150U107 529150U120 529150U124 529160US14 529160US17 529230US27 CHULS23252078 560140US09 475480US12 475480US14 150080US04 150080US06 150080US05 150080US07 794210US44 794210US50 794210US51 541010US18 CHULS23226325 850020US86 900010U167 0853372631 0853373218 0853372468 0853372849 0853372556 0853372557 0853372561 0853372560 0853372637 0853372710 0853373219 0853372518 0853373212 0853372940 0853373070 0853372656 0853372489 0853372841 0853373188 0853372676 0853372496 0853372405 0853372711 0853373213 0853372467 0853373187 0853372392 0853372620 0853373016 0853372978 0853372519 0853372916 0853372915 0853372920 0853373038 0853373185 0853372954 0853372395 0853372746 0853372493 0853372670 0853372774 0853372749 0853372750 0853372751 0853372752 0853372909 0853373171 0853372966 0853373220 0853372867 0853373022 0853372827 0853372810 0853372616 0853372617 0853372680 0853372842 0853372979 0853372854 0853372967 0853372408 0853372663 0853372664 0853372767 0853372868 0853372419 0853372876 0853372844 0853372495 0853372501 1573UL3507 1573UL3929 1573UL3930 1573UL3930 1573UL3931 1573UL3929 1573UL3927 1573UL2518 1573UL3604 1573UL4320 1573UL4321 1573UL4322 1573UL4323 1573UL4324 1573UL4325 1573UL4326 1573UL4327 1573UL4328 1573UL4329 1573UL4330 1573UL4331 1573UL4332 1573UL4333 1573UL4334 1573UL3751 1573UL2843 1573UL4106 1573UL4406 1573UL2361 1573UL3067 1573UL3242 1573UL3755 1573UL3850 1573UL2339 1573UL2288 1573UL2351 1573UL3007 1573UL3506 1573UL3508 1573UL3509 1573UL3510 1573UL3065 1573UL3068 1573UL3522 1573UL3634 1573UL3394 1573UL2149 1573UL2397 1573UL3287 1573UL3396 1573UL2841 1573UL2517 1573UL3743 1573UL3257 1573UL2151 1573UL2877 1573UL3523 1573UL3511 1573UL3550 1573UL4405 1573UL3502 1573UL2879 1573UL3395 1573UL2338 1573UL3503 1573UL3549 1573UL3773 1573UL2399 1573UL3168 1573UL3169 1573UL3170 1573UL3171 1573UL3172 1573UL3521 1573UL4162 1573UL2817 1573UL2921 1573UL3547 1573UL3009 1573UL3669 1573UL2774 1573UL2715 1573UL2716 1573UL2948 1573UL2337 1573UL2349 1573UL2350 1573UL3854 1573UL3401 1573UL3635 THULS54307041 1573UL3246 1573UL2009 1573UL2857 1573UL2523 1573UL4055 1573UL2818 1573UL2348 1573UL2616 1573UL3243 1573UL2347 1573UL2519 1573UL3775 1573UL2285 1573UL2286 1573UL2858 1573UL3008 1573UL2287 1573UL3066 1573UL2723 1573UL3564 1573UL2856 1573UL2047 1573UL3548 1573UL3504 1573UL3987 1573UL2767 1573UL3858 1573UL2876 1573UL3774 1573UL3827 1573UL2766 1573UL2831 1573UL2613 1573UL3515 1573UL2051 1573UL2284 1573UL3174 1573UL3393 1573UL3605 THULS54307061 1573UL3985 1573UL4054 1573UL2546 1573UL3418 1573UL2148 1573UL2919 1573UL2920 1573UL3816 1573UL3501 1573UL2398 1573UL3135 1573UL2001 1573UL2520 1573UL3133 1573UL3184 1573UL3285 1573UL3571 1573UL3852 1573UL4160 1573UL3636 1573UL3637 1573UL3638 1573UL3670 1573UL3776 1573UL4407 1573UL4404 1573UL4431 1573UL4446 1573UL4453 1573UL4471 1573UL4461 1573UL4470 1573UL4554 THULS53873431B THULS53873431C THULS53873431A 1573UL4549 1573UL4550 1573UL3986 1573UL3892 1573UL3893 1573UL4080 1573UL4413 1573UL4447 1573UL4444 1573UL4452 1573UL4468 1573UL4610 1573UL4620 1573UL2569 1573UL3245 1573UL2827 1573UL3134 1573UL3244 1573UL3392 1573UL3821 TN1121US05 TN1500US04 34008ULT35 P34189UL31 26001ULT25 VGC007675 VGD002344 VGD002461 VGD002474 46003ULT25 01002ULT50 01002ULT53 34393ULT33 VGC000014 VGDW000298 VGDW000097 VGC006589 VGC007432 AE1046US10 AE1046US13 VGD001971 VGD001530 VGD001531 AE1139US15 AE1016US32 AE1016US38 AE1016US39 AE1016US40 VGD000108 VGC007737 AE1486US54 AE1682US08 AE1682US09 VGD000170 AE1405U191 AE1007US47 AE1007US45 AE1007US46 VGD000069 AE1405U204 AE1405U210 AE1015US81 AE1015US82 VGD001443 VGD002480 AE1741US03 VGD002399 AE1405U215 AE1445US09 AE1445US07 AE1445US06 AE1445US05 AE1445US08 AE1445US10 AE1235US11 VGD000423 VGD000414 AE1149US28 AE1149US29 AE1149US38 VGD001973 OM1042US38 LB1218US01 LB1377US10 AE1149US30 AE1015US82 00192ULT68 00192USX02 00192USX01 00192ULT82 00192ULT81 00192ULT99 GBULS23294140 11383USX13 05169ULT09 05169ULT10 GBULS22173793 GBULS24423349 GBULS23299526 GBULS24850745 10010ULT20 10010ULT22 10010ULT21 00309USX52 00309USX51 00309ULT60 GBULS22310779 00316ULT31 00316ULT32 00316ULT30 05452ULT30 00249USX10 00249USX11 00249USX13 00249USX12 00249ULT15 00249ULT16 00362USX21 00336ULT02 05228ULT03 00558ULT04 IE1502USX2 VGDW000156 VGDW000157 VGDW000158 VGDW000160 VGDW000165 VGDW000170 VGDW000803 IE6070US02 IE6070US03 10347ULT01 00473ULT58 00473ULT59 00271ULT52 GBULS22247784 11342ULT01 11714USX08 11714USX06 00573ULT03 00101ULT30 00101ULT33 00101ULT36 00101ULT37 00658ULT02 00439ULT46 00439ULT49 00439ULT56 00439ULT57 00329USX49 00329USX56 00329USX54 00329USX55 00479ULT20 00479ULT21 GBULS22203192 GBULS22203193 00105ULT37 00105ULT41 GBULS23292874 GBULS23201208 GBULS23206853 GBULS22308570 00487ULT37 00487ULT32 00631USX25 00631USX24 00631USX26 00631USX22 00631USX23 00631USX21 00631USX28 00211ULT57 00211ULT56 00211USX59 00211USX58 00060USX82 00211USX62 00211USX60 00211USX61 00211USX63 10319ULT25 GBULS21998071 00242USX19 00242USX20 00242USX21 00242ULT41 00242ULT52 05230ULT07 11967ULT01 00094USX30 00094USX31 00094ULT35 GBULS22258378 00001USX86 00001USX85 00001USX83 00001USX84 00001USX90 00001USX89 00001USX88 00001USX87 00001USX91 00001USX93 00001USX96 00001USX92 00001USX95 00001USX97 00001USX94 00001ULTD2 00001ULTD4 00001ULTD3 GBULS22182069 GBULS22284485 00417ULT04 GBULS23299603 GBULS22298048 00055USX30 00055USX31 00055ULT49 11042ULT03 10705ULTC2 10705ULTA3 10705ULTB2 10705ULTG4 00058USX45 00058USX47 00058USX46 GBULS24848214 21414ULT02 GBULS22140116 GBULS22140117 GBULS22140118 GBULS22140115 GBULS22140119 00358USX39 00358ULT45 IE1503US17 00361ULT17 00361ULT28 00440ULTC2 00440ULTE5 00440ULTE3 00440ULTE4 00440ULTF1 GBULS22174177 GBULS22224755 GBULS22224773 GBULS22224772 GBULS22224771 GBULS22158673 GBULS22153891 GBULS22206965 00366ULT23 00366ULT24 00366ULT31 20011ULT04 10671ULT03 00327ULT27 10351USX15 10351USX14 10351ULT21 00257ULTB5 00257ULTB6 GBULS22270014 00355ULT20 00190ULT07 00246ULT03 00468USX15 00468USX18 00545ULT20 11383USX15 00197ULT64 11383USX26 11383USX28 11383USX21 11383USX14 11383USX31 11557ULT01 GBULS22153114 00298ULT64 00298ULT63 00298ULT65 00914ULT07 00914USX01 00199USX01 11383USX07 00199ULT67 20491ULT02 11383USX08 00200ULT27 00006ULT23 00006ULT24 10223ULT03 00507USX20 00507USX27 00507ULT31 00507ULT32 11834ULT01 11351USX01 00231ULT20 00231ULT25 00231ULT33 00231ULT36 00091USX42 00091USX43 00091USX49 00091USX50 00091ULT44 00091ULT43 00091ULT48 GBULS22182729 00127USX11 00127USX10 00127ULT14 00127ULT11 GBULS22254721 GBULS22271959 00149ULT57 00149ULT56 00743ULT02 20180ULT07 20180ULT08 GBULS22143732 21242ULT40 21242ULT41 21242ULT42 21242ULT43 11383USX19 11383USX09 11383USX18 GBULS22258350 GBULS21997935 GBULS21997936 GBULS21997937 GBULS21997938 GBULS22199322 GBULS21997941 GBULS21997942 00365USX47 00365USX48 00365USX45 00365USX49 00365USX50 00365ULT55 00175USX16 00175USX14 00175USX15 GBULS22203191 GBULS23216594 10609ULT01 GBULS22139856 00182USX12 00182USX11 00182ULT14 00182ULT15 00182ULT17 00182ULT18 00182ULT16 00064USX25 00064USX33 00064USX40 00064USX30 00064USX23 00064USX27 00064USX39 00064USX26 00064USX43 00064USX44 00064USX45 00478ULT56 00478ULT58 00018USX76 00018USX77 00018USX78 00463ULT29 00463ULT30 11590ULT01 GBULS22258372 00422ULT08 00422USX10 11341ULT03 GBULS24843640 20099US888 GBULS22300433 GBULS22310780 00296USX14 00296USX15 11593ULT10 11593ULT11 00459USX29 GBULS23290138 GBULS22239992 00181ULT06 00130USX29 00130USX27 00130USX30 00130USX28 00130ULT34 00130ULT37 10329ULT05 00556ULT41 GBULS22258346 GBULS24512385 00560USX43 00560ULT46 00560ULT47 00560ULT48 00560ULT49 00560ULT54 00560ULT57 00560ULT55 00560ULT56 20619ULT04 00102ULT06 00102ULT11 05451USX54 05451ULT56 05451ULT61 05451ULT63 00195USX01 GBULS22206939 GBULS22206940 00039ULT72 00039ULT74 00929USX01 11383USX23 11383USX27 00929ULT37 00929ULT38 00165USX29 00165USX32 00165USX31 00165USX30 10627USX01 00132USX65 10434ULT06 00318ULT78 00318ULT79 00318USX75 00318USX74 00318USX73 VGDW000802 VGDW000814 GBULS22258320 GBULS21998070 GBULS23290852 GBULS23263283 00686ULT14 00686ULT17 00025USX35 00025USX38 11596ULT01 00329USX51 00329USX50 00329USX52 00329USX53 00329USX57 00329USX58 00329USX60 00329USX59 00174ULT61 00174ULT62 00174ULT63 11562ULT04 20410ULT09 11653ULT02 11653ULT01 10344ULT20 11662ULT01 01347ULT99 01347USXA8 01347ULTD3 01347ULTD4 01347ULTD5 01347ULTD6 01347ULTD2 01347ULTD7 01347ULTD8 01347ULTD9 00174ULT65 01347ULTE9 11560ULT01 11513ULT15 11513ULT16 GBULS24303300 10510ULT17 10510ULT18 10875ULT02 20099US886 21361USX28 21361ULT35 21361ULT37 21361ULT36 00060US062 00069ULT61 00069ULT60 00383USX68 11745ULT01 11854ULT01 21244ULT14 00085ULT18 10531ULT15 10531ULT11 00040USX01 00040USX02 00040ULTB2 00040ULTB3 00040ULTB4 00040ULTB8 21218ULT05 11340ULT16 11340ULT07 11340ULT08 11340ULT06 11340ULT31 11552ULT08 11552ULT10 00023USXB8 00023ULTB9 00095ULTD5 00095ULTD4 00095ULTD6 00095ULTD7 00095ULTD8 00095ULTH4 00095ULTH3 00095ULTI1 00383ULT56 00383USX66 00383ULT59 00383ULT66 00064USX38 00383USX67 00383ULT67 00383ULT70 00383ULT69 00383ULT68 00060US049 00060US050 00060US047 00060US048 00060US051 00060USX84 00060US066 00060US068 00060US070 00060US069 00060US063 00060US064 00060US067 00060US060 00060US061 00060USX73 00060USX81 00060USX78 00060USX72 00060USX75 00060USX77 00060USX74 00060USX79 00631USX27 00060USX83 00060USX76 00060USX80 00060ULTA2 00060ULTB3 GBULS22243877 00026USX75 00026USX74 00026ULT71 00026ULT72 00028USX99 00028USX91 00028USX90 20584ULT10 20584ULT13 20584ULT12 20584ULT23 GBULS23215207 VGB001588 VGB001596 VGB001602 GBULS22216963 GBULS22216964 GBULS22216978 GBULS21997934 GBULS21997933 00314ULT35 00314ULT41 GBULS22157016 GBULS22254732 GBULS22258341 GBULS22258332 GBULS22258347 GBULS22258345 GBULS22258370 GBULS22258331 GBULS22258336 GBULS22258371 GBULS22258381 GBULS22254718 GBULS22258343 GBULS22254719 GBULS22258368 GBULS22258355 GBULS22258338 GBULS22258340 GBULS22258316 GBULS22258317 GBULS22258318 GBULS22258342 GBULS22258380 GBULS22258330 GBULS22258329 GBULS22258363 GBULS22258334 GBULS22258322 GBULS22258323 GBULS22258362 GBULS22258361 GBULS22258328 GBULS22258367 GBULS22254733 GBULS22258369 GBULS22254720 GBULS22258366 GBULS22254722 GBULS22254727 GBULS22254728 GBULS22254726 GBULS22258376 GBULS22258377 GBULS22258384 GBULS22258351 GBULS22258360 GBULS22258379 GBULS22258353 GBULS22258358 GBULS22258344 GBULS22258357 GBULS22258374 GBULS22258375 GBULS22258326 GBULS22258308 GBULS22258306 GBULS22258321 GBULS22258333 GBULS22258327 GBULS22258383 GBULS22258325 GBULS22258354 GBULS22258335 GBULS22258319 GBULS22258324 00133ULT84 GBULS22221989 00229ULT24 00229ULT25 12004ULT01 12004ULT02 12004ULT03 12004ULT04 00282ULT41 GBULS22630173 21375ULT04 11202ULT02 00772ULT05 00443USX41 00443USX42 00443USX43 00443USX50 00443USX51 00443USX54 00443USX49 00443USX52 00443USX53 00443USX55 00443USX56 00443USX62 00443USX57 00443USX59 00443USX58 00443USX60 00443ULT76 00443ULT78 00443ULT80 00443ULT74 00443ULT75 00443ULT77 00443ULT79 GBULS23290853 GBULS23290851 GBULS23290854 00356ULT14 00356ULT15 00356ULT16 00356ULT18 00356USX21 00356USX22 00262ULT31 00262ULT32 00262ULT35 00262ULT36 00262USX43 00262ULT55 00262ULT63 00262ULT62 00262ULT64 GBULS22281304 21172USX10 21172USX09 21172ULT11 21172ULT12 21172ULT13 21172ULT16 21172ULT17 09088ULT02 11344ULT01 GBULS23287310 00431USX59 00431USX63 00431USX65 00431USX64 00431USX66 00431USX70 00431USX68 00431USX71 00431USX69 00431USX67 00431USX80 00431ULT75 00431ULTB5 00431ULTB7 00432USX65 00432USX64 00432USX63 11091ULT03 21375ULT05 20591USX08 20591USX10 20591USX09 20591USX11 20591USX12 GBULS22206959 00583ULT04 00583ULT07 00583ULT05 00583ULT09 00583ULT10 11905ULT01 00471ULT06 05461ULT08 10448USX11 10448US007 10448USX12 10448ULTA6 GBULS22140753 20309ULT10 11383USX32 11383USX03 00352ULT39 00216USX26 00216USX25 00216USX27 00216ULT33 00216ULT32 00216ULT31 GBULS22225525 GBULS23226333 GBULS23225750 GBULS23225751 GBULS23225749 GBULS22258348 IE6068US01 IE6068US02 00230ULT42 GBULS22258365 00100USX09 00100ULT98 00100ULT99 00100ULTB4 GBULS22161497 GBULS23268943 00137ULT31 00137ULT32 00137ULT43 00137USX41 00137USX40 00137ULT53 00137ULT42 00137ULT58 20229ULT05 00324ULT36 00324ULT32 00324ULT33 00324ULT34 00324ULT35 GBULS22153898 GBULS22258364 00506ULT47 00506ULT48 00506ULT49 00506ULT62 00506ULT61 00506ULT57 00506ULT58 00506ULT59 20180ULT05 20180ULT06 00116ULT19 00116ULT18 00116ULT20 05141ULT10 05141ULT11 00131ULT13 00131ULT14 00131ULT15 00131ULT16 00131ULT17 00131ULT18 00131ULT32 00131ULT31 00131ULT33 00224ULT21 00345ULT20 00345ULT25 00632USX12 GBULS22258352 GBULS22310771 01314ULT01 00311USX26 00311USX28 00311USX25 00311USX29 00311USX27 GBULS22187677 GBULS22187699 00147ULT03 00292ULT22 00292ULT20 00292ULT21 01236ULT07 01236ULT06 01089ULT04 00674ULT07 00674ULT06 00674ULT04 00320ULT01 10878ULT01 00371ULT04 00490USX60 00490ULT62 00490ULT63 00490ULT64 00490ULT65 00490ULT66 00490ULT67 00490ULT76 00490ULT75 21203ULT09 11383USX25 11383USX30 GBULS24526861 GBULS22131836 GBULS22223097 00169ULT29 00169ULT39 00169ULT42 00169ULT41 00169ULT40 00169ULT37 00169ULT37 00169ULT36 00169ULT44 00223USX36 00223USX37 10447ULT14 11618ULT18 11618ULT19 11618ULT20 11618ULT21 11618ULT17 11618ULT22 11618ULT23 11618ULT24 GBULS22271184 GBULS22271183 01367ULT29 01367ULT31 01367ULT30 00391ULT46 00391ULT47 11904ULT01 VGB002902 20248ULT02 21242ULT60 21242ULT61 01119ULT41 01119ULT42 GBULS22258373 GBULS22258359 05202ULT43 00140ULT77 00140ULT85 00140ULT86 00140ULT87 00140ULT88 00140ULT78 00140ULT79 00140ULT80 00140ULT81 00140ULT82 00140ULT83 00140ULT84 00424ULT38 00424ULT39 GBULS22263939 10507ULT06 00331USX24 00331ULT39 GBULS22239998 05202ULT44 01041USX18 01041USX19 01041ULT26 GBULS22258307 GBULS23138428 00084ULT60 00084USX56 00084USX57 00084USX61 00084USX62 00084USX67 00348ULT10 00348USX13 00348ULT11 00348ULT12 00478ULT27 00478ULT28 00478ULT54 00478ULT57 00478ULT59 GBULS22258337 GBULS22258339 00659ULT02 19488USX07 00478USX32 00488ULT30 00488ULT18 00488ULT17 GBULS22165108 00416USX22 11383USX12 00141USX04 GBULS23290140 00347ULT14 00423USX40 00423USX41 00442USX24 00442USX25 GBULS22258349 00171ULT49 00171USX28 00171USX29 00171USX27 00171USX30 00171USX33 00171USX32 00171USX41 00171USX40 GBULS22370327 VGOVGC007647 VGOVGC009082 5973UL2855 5973UL4281 5973UL2854 Mfg. Lot or Serial # VGE000167 VGE000162 VGE000147 VGE000320 VGE000321 VGE000330 VGE000200 VGE000198 VGE000424 VGE000169 VGE000444 VGE000364 VGE000326 VGE000327 VGE000172 VGE000438 VGE000352 VGE000428 VGE000360 VGE000289 VGE000450 VGE000377 VGE000379 VGE000291 VGE000333 VGE000164 VGE000366 VGE000367 VGE000223 VGE000224 VGE000228 VGE000340 VGE000381 VGE000380 VGE000317 VGE000431 VGE000432 VGE000433 VGE000434 VGE000214 VGE000415 VGE000196 VGE000184 VGE000195 VGE000378 VGE000391 VGE000370 VGE000441 VGE000183 VGE000189 VGE000215 VGE000305 VGE000307 VGE000124 VGE000129 VGE000429 VGE000206 VGE000209 VGE000001 VGE000057 VGE000056 VGE000058 VGE000065 VGE000073 VGE000069 VGE000109 VGE000111 VGE000112 VGE000113 VGE000418 VGE000226 VGE000386 VGE000385 VGE000302 VGE000264 VGE000176 VGE000369 VGE000342 VGE000343 VGE000348 VGE000362 VGE000361 VGE000347 VGE000145 VGE000180 VGE000160 VGE000123 VGE000384 VGE000387 VGE000218 VGE000389 VGE000353 VGE000116 VGE000118 VGE000130 VGE000281 VGE000205 VGE000298 VGE000357 VGE000282 VGE000285 VGE000286 VGE000411 VGE000413 VGE000324 VGE000328 VGE000323 VGE000325 VGE000436 VGE000199 VGE000203 VGE000201 VGE000212 VGE000372 VGE000290 VGE000293 VGE000292 VGE000295 VGE000114 VGE000128 VGE000122 VGE000127 VGE000126 VGE000136 VGE000139 VGE000137 VGE000138 VGE000140 VGE000177 VGE000171 VGE000316 VGE000174 VGE000427 VGE000279 VGE000371 VGE000288 VGE000268 VGE000270 VGE000271 VGE000272 VGE000273 VGE000211 VGE000157 VGE000168 VGE000299 VGE000210 VGE000222 VGE000350 VGE000351 VGE000244 VGE000401 VGE000420 VGE000421 VGE000422 VGE000423 VGE000425 VGE000322 VGE000178 VGE000455 VGE000390 VGE000365 VGE000375 VGE000283 VGE000287 VGE000175 VGE000301 VGE000125 VGE000159 VGE000158 VGE000331 VGE000216 VGE000296 VGE000303 VGE000345 VGE000204 VGE000152 VGE000115 VGE000153 VGE000221 VGE000220 VGE000294 VGE000148 VGE000408 VGE000338 VGE000225 VGE000227 VGE000236 VGE000358 VGE000297 VGE000275 VGE000280 VGE000402 VGE000405 VGE000185 VGE000202 VGE000397 VGE000398 VGE000406 VGE000407 VGE000410 VGE000409 VGE000154 VGE000376 VGE000277 VGE000166 VGE000373 VGE000442 VGE000451 VGE000454 VGE000181 VGE000344 VGE000416 VGE000426 VGE000173 VGE000276 VGE000341 VGE000414 VGE000315 VGE000337 VGE000329 VGE000334 VGE000311 VGE000374 VGE000197 VGE000278 VGE000359 VGE000412 VGE000192 VGE000194 VGE000435 VGE000335 VGE000332 VGE000151 VGE000213 VGE000142 VGE000430 VGE000217 VGE000319 VGE000300 VGE000219 VGE000170 VGE000182 VGE000284 VGE000188 VGE000191 VGE000193 VGE000346 VGE000306 VGE000313 VGE000314 VGE000265 VGE000267 VGE000266 VGE000269 VGE000308 VGE000263 VGE000121 VGE000349 VGE000392 VGE000396 VGE000368 VGE000336 VGE000339 VGE000440 VGE000312 VGE000187 VGE000186 VGE000190 VGE000254 VGE000356 VGE000363 VGE000355 VGE000006 VGE000045 VGE000029 VGE000042 VGE000044 VGE000052 VGE000050 VGE000054 VGE000067 VGE000208 VGE000229 VGE000232 VGE000233 VGE000237 VGE000238 VGE000242 VGE000234 VGE000248 VGE000235 VGE000246 VGE000230 VGE000231 VGE000245 VGE000239 VGE000240 VGE000251 VGE000243 VGE000247 VGE000249 VGE000256 VGE000253 VGE000255 VGE000257 VGE000252 VGE000250 VGE000261 VGE000260 VGE000258 VGE000259 VGE000262 VGE000274 VGE000007 VGE000008 VGE000009 VGE000010 VGE000011 VGE000012 VGE000013 VGE000015 VGE000014 VGE000017 VGE000019 VGE000026 VGE000030 VGE000034 VGE000036 VGE000043 VGE000018 VGE000020 VGE000022 VGE000024 VGE000032 VGE000033 VGE000037 VGE000038 VGE000046 VGE000047 VGE000023 VGE000027 VGE000039 VGE000040 VGE000051 VGE000028 VGE000031 VGE000035 VGE000016 VGE000021 VGE000025 VGE000041 VGE000048 VGE000053 VGE000049 VGE000055 VGE000061 VGE000066 VGE000060 VGE000059 VGE000064 VGE000062 VGE000063 VGE000075 VGE000087 VGE000082 VGE000068 VGE000072 VGE000076 VGE000089 VGE000070 VGE000071 VGE000074 VGE000077 VGE000078 VGE000083 VGE000084 VGE000085 VGE000088 VGE000080 VGE000086 VGE000090 VGE000092 VGE000094 VGE000093 VGE000097 VGE000099 VGE000091 VGE000096 VGE000098 VGE000102 VGE000106 VGE000108 VGE000100 VGE000103 VGE000104 VGE000105 VGE000107 VGE000155 VGE000163 VGE000393 VGE000394 VGE000382 VGE000179 VGE000417 VGE000354 VGE000383 VGE000388 VGE000395 VGE000161 VGE000310 VGE000309 VGE000002 VGE000004 VGE000005 VGE000318 VGE000304 VGE000003 VGE000144 VGE000141 VGE000143 VGEW000392 VGEW000495 VGEW000452 VGEW000281 VGEW000131 VGEW000380 VGEW000106 VGEW000134 VGEW000608 VGEW000655 VGEW000501 VGEW000019 VGEW000038 VGEW000033 VGEW000494 VGEW000499 VGEW000650 VGEW000680 VGEW000098 VGEW000217 VGEW000435 VGEW000478 VGEW000480 VGEW000483 VGEW000598 VGEW000599 VGEW000606 VGEW000607 VGEW000609 VGEW000610 VGEW000613 VGEW000614 VGEW000616 VGEW000618 VGEW000620 VGEW000623 VGEW000630 VGEW000632 VGEW000661 VGEW000664 VGEW000667 VGEW000668 VGEW000669 VGEW000672 VGEW000673 VGEW000674 VGEW000675 VGEW000676 VGEW000677 VGEW000678 VGEW000679 VGEW000681 VGEW000682 VGEW000683 VGEW000684 VGEW000685 VGEW000686 VGEW000687 VGEW000688 VGEW000689 VGEW000690 VGEW000691 VGEW000693 VGEW000694 VGEW000695 VGEW000696 VGEW000697 VGEW000698 VGEW000699 VGEW000700 VGEW000701 VGEW000702 VGEW000703 VGEW000704 VGEW000360 VGEW000189 VGEW000197 VGEW000040 VGEW000050 VGEW000625 VGEW000579 VGEW000155 VGEW000104 VGEW000456 VGEW000161 VGE000400 VGE000403 VGE000404 VGE000437 VGE000443 VGE000439 VGE000445 VGE000446 VGE000447 VGE000448 VGEW000176 VGEW000427 VGEW000454 VGEW000474 VGEW000479 VGEW000546 VGEW000634 VGEW000651 VGEW000083 VGEW000030 VGEW000013 VGEW000422 VGEW000462 VGEW000346 VGEW000535 VGEW000009 VGEW000014 VGEW000072 VGEW000081 VGEW000290 VGEW000055 VGEW000128 VGEW000521 VGEW000203 VGEW000041 VGEW000306 VGEW000576 VGEW000652 VGEW000659 VGEW000660 VGEW000663 VGEW000666 VGEW000027 VGEW000032 VGEW000036 VGEW000044 VGEW000367 VGEW000372 VGEW000373 VGEW000099 VGEW000387 VGEW000132 VGEW000364 VGEW000342 VGEW000069 VGEW000264 VGEW000139 VGEW000180 VGEW000054 VGEW000074 VGEW000294 VGEW000101 VGEW000331 VGEW000541 VGEW000564 VGEW000093 VGEW000171 VGEW000097 VGEW000444 VGEW000232 VGEW000192 VGEW000166 VGEW000056 VGEW000057 VGEW000276 VGEW000277 VGEW000471 VGEW000477 VGEW000637 VGEW000612 VGEW000552 VGEW000568 VGEW000586 VGEW000588 VGEW000591 VGEW000641 VGEW000642 VGEW000269 VGEW000230 VGEW000324 VGEW000280 VGEW000196 VGEW000352 VGEW000533 VGEW000486 VGEW000287 VGEW000048 VGEW000061 VGEW000469 VGEW000493 VGEW000298 VGEW000204 VGEW000003 VGEW000084 VGEW000146 VGEW000170 VGEW000295 VGEW000303 VGEW000354 VGEW000363 VGEW000407 VGEW000425 VGEW000432 VGEW000507 VGEW000368 VGEW000224 VGEW000262 VGEW000284 VGEW000511 VGEW000633 VGEW000433 VGEW000622 VGEW000178 VGEW000075 VGEW000491 VGEW000307 VGEW000548 VGEW000549 VGEW000278 VGEW000273 VGEW000639 VGEW000222 VGEW000406 VGEW000066 VGEW000086 VGEW000259 VGEW000265 VGEW000270 VGEW000274 VGEW000288 VGEW000503 VGEW000640 VGEW000497 VGEW000028 VGEW000502 VGEW000052 VGEW000506 VGEW000268 VGEW000470 VGEW000020 VGEW000025 VGEW000049 VGEW000053 VGEW000293 VGEW000347 VGEW000550 VGEW000631 VGEW000381 VGEW000063 VGEW000223 VGEW000234 VGEW000447 VGEW000213 VGEW000229 VGEW000282 VGEW000431 VGEW000525 VGEW000526 VGEW000532 VGEW000556 VGEW000566 VGEW000488 VGEW000135 VGEW000078 VGEW000570 VGEW000095 VGEW000557 VGEW000094 VGEW000370 VGEW000523 VGEW000199 VGEW000271 VGEW000351 VGEW000534 VGEW000670 VGEW000257 VGEW000258 VGEW000391 VGEW000296 VGEW000021 VGEW000289 VGEW000046 VGEW000200 VGEW000300 VGEW000316 VGEW000194 VGEW000071 VGEW000621 VGEW000560 VGEW000305 VGEW000302 VGEW000059 VGEW000457 VGEW000266 VGEW000275 VGEW000207 VGEW000215 VGEW000034 VGEW000077 VGEW000443 VGEW000458 VGEW000514 VGEW000563 VGEW000304 VGEW000475 VGEW000627 VGEW000628 VGEW000629 VGEW000285 VGEW000202 VGEW000375 VGEW000227 VGEW000528 VGEW000208 VGEW000151 VGEW000154 VGEW000157 VGEW000162 VGEW000164 VGEW000165 VGEW000167 VGEW000169 VGEW000172 VGEW000173 VGEW000179 VGEW000182 VGEW000183 VGEW000184 VGEW000185 VGEW000186 VGEW000191 VGEW000198 VGEW000211 VGEW000233 VGEW000236 VGEW000237 VGEW000239 VGEW000244 VGEW000245 VGEW000246 VGEW000247 VGEW000595 VGEW000024 VGEW000604 VGEW000647 VGEW000649 VGEW000476 VGEW000064 VGEW000581 VGEW000160 VGEW000201 VGEW000551 VGEW000554 VGEW000235 VGEW000542 VGEW000228 VGEW000045 VGEW000031 VGEW000007 VGEW000286 VGEW000362 VGEW000388 VGEW000590 VGEW000016 VGEW000460 VGEW000319 VGEW000129 VGEW000221 VGEW000468 VGEW000473 VGEW000231 VGEW000594 VGEW000336 VGEW000122 VGEW000118 VGEW000113 VGEW000593 VGEW000114 VGEW000597 VGEW000159 VGEW000218 VGEW000219 VGEW000242 VGEW000252 VGEW000643 VGEW000125 VGEW000337 VGEW000339 VGEW000410 VGEW000411 VGEW000412 VGEW000417 VGEW000438 VGEW000440 VGEW000442 VGEW000645 VGEW000656 VGEW000657 VGEW000658 VGEW000123 VGEW000644 VGEW000116 VGEW000596 VGEW000408 VGEW000121 VGEW000251 VGEW000220 VGEW000439 VGEW000601 VGEW000602 VGEW000603 VGEW000254 VGEW000112 VGEW000120 VGEW000413 VGEW000338 VGEW000253 VGEW000409 VGEW000108 VGEW000109 VGEW000110 VGEW000111 VGEW000115 VGEW000248 VGEW000249 VGEW000250 VGEW000255 VGEW000256 VGEW000344 VGEW000415 VGEW000416 VGEW000124 VGEW000414 VGEW000592 VGEW000158 VGEW000600 VGEW000126 VGEW000441 VGEW000119 VGEW000117 VGEW000107 VGEW000654 VGEW000060 VGEW000279 VGEW000451 VGEW000356 VGEW000482 VGEW000335 VGEW000085 VGEW000042 VGEW000374 VGEW000389 VGEW000079 VGE000117 VGEW000070 VGEW000080 VGEW000089 VGEW000291 VGEW000310 VGEW000496 VGEW000573 VGEW000102 VGEW000067 VGEW000315 VGEW000317 VGEW000318 VGEW000326 VGEW000328 VGEW000430 VGEW000529 VGEW000490 VGEW000073 VGEW000015 VGEW000011 VGEW000308 VGEW000648 VGEW000065 VGEW000485 VGEW000002 VGEW000005 VGEW000705 VGEW000692 VGE000149 VGE000150 VGE000207 VGEW000301 VGEW000076 VGEW000105 VGEW000136 VGEW000145 VGEW000148 VGEW000149 VGEW000152 VGEW000153 VGEW000051 VGEW000087 VGEW000327 VGEW000008 VGEW000037 VGEW000332 VGEW000355 VGEW000133 VGEW000141 VGEW000150 VGEW000187 VGEW000188 VGEW000283 VGEW000571 VGEW000464 VGEW000510 VGEW000500 VGEW000505 VGEW000508 VGEW000519 VGEW000504 VGEW000520 VGEW000498 VGEW000515 VGEW000453 VGEW000553 VGEW000100 VGEW000103 VGEW000190 VGEW000147 VGEW000397 VGEW000062 VGEW000012 VGEW000365 VGEW000366 VGEW000369 VGEW000378 VGEW000379 VGEW000383 VGEW000384 VGEW000386 VGEW000395 VGEW000396 VGEW000399 VGEW000400 VGEW000401 VGEW000403 VGEW000404 VGEW000322 VGEW000325 VGEW000329 VGEW000334 VGEW000434 VGEW000091 VGEW000421 VGEW000626 VGEW000635 VGEW000638 VGEW000671 VGEW000238 VGEW000226 VGEW000330 VGEW000418 VGEW000423 VGEW000446 VGEW000449 VGEW000459 VGEW000461 VGEW000463 VGEW000465 VGEW000466 VGEW000467 VGEW000437 VGEW000212 VGEW000225 VGEW000240 VGEW000029 VGEW000333 VGEW000405 VGEW000068 VGEW000297 VGEW000216 VGEW000047 VGEW000092 VGEW000130 VGEW000143 VGEW000096 VGEW000142 VGEW000039 VGEW000090 VGEW000455 VGEW000636 VGEW000653 VGEW000662 VGEW000665 VGEW000022 VGEW000023 VGEW000026 VGEW000558 VGEW000168 VGEW000393 VGEW000058 VGEW000292 VGEW000527 VGEW000210 VGEW000043 VGEW000181 VGEW000174 VGEW000241 VGEW000243 VGEW000261 VGEW000272 VGEW000341 VGEW000343 VGEW000340 VGEW000394 VGEW000522 VGEW000127 VGEW000144 VGEW000035 VGEW000371 VGEW000513 VGEW000555 VGEW000611 VGEW000624 VGEW000175 VGEW000402 VGEW000517 VGEW000538 VGEW000137 VGEW000138 VGEW000140 VGE000110 VGEW000267 VGEW000492 VGEW000004 VGEW000010 VGEW000017 VGEW000487 VGEW000489 VGEW000516 VGEW000531 VGEW000536 VGEW000545 VGEW000567 VGEW000574 VGEW000582 VGEW000583 VGEW000584 VGEW000589 VGEW000263 VGEW000559 VGEW000424 VGEW000429 VGEW000436 VGEW000214 VGEW000385 VGEW000524 VGEW000530 VGEW000313 VGEW000359 VGEW000561 VGEW000565 VGEW000646 VGEW000323 VGEW000314 VGEW000518 VGEW000163 VGEW000382 VGEW000398 VGEW000450 VGEW000562 VGEW000578 VGEW000018 VGEW000177 VGEW000537 VGEW000539 VGEW000543 VGEW000577 VGEW000312 VGEW000580 VGEW000358 VGEW000615 VGEW000309 VGEW000299 VGEW000088 VGEW000311 VGEW000512 VGEW000420 VGEW000350 VGEW000617 VGEW000006 VGEW000353 VGEW000357 VGEW000361 VGEW000544 VGEW000547 VGEW000195 VGEW000205 VGEW000206 VGEW000209 VGEW000472 VGEW000481 VGEW000484 VGEW000156 VGEW000260 VGEW000585 VGEW000509 VGEW000321 VGEW000193 VGEW000377 VGEW000605 VGEW000376 VGEW000619 VGEW000348 VGEW000349 VGEW000419 VGEW000426 VGEW000428 VGEW000445 VGEW000448 VGEW000320 VGEW000540 VGEW000587 VGEW000572 VGEW000575 VGEW000345 VGEW000390 VGEW000569
FEI Number 2126677
Recalling Firm/
Manufacturer
GE Medical Systems, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
Manufacturer Reason
for Recall
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/6/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Venue Go and Venue Fit system after following the actions below: 1) Turn the system Off and unplug the system from the wall socket. 2) Remove the system batteries by following instructions provided in the User Manual Chapter 3. These instructions are also provided in this letter in Appendix A for Venue Go and in Appendix B for Venue Fit. 3) Check the date of manufacture of the system batteries by looking at the label on the bottom of each battery (See Figure 1). The date of manufacture is displayed as YYYY-MM for year and month of manufacture. NOTE: Venue Go has two batteries and Venue Fit has one battery 4) Calculate the age of each of the system batteries. 5) If any of the batteries are more than 2 years old: a. With the system unplugged, re-insert batteries, turn the system On and leave the system On to discharge the batteries. b. Once the battery capacity indicator shows two segments or less (See Figure 2), turn the system Off and remove the batteries per step 2. c. Dispose of batteries according to local regulations. NOTE: After battery removal, you can continue to use the system with use of AC power cord. The system is fully functional without the battery. 6) If the batteries are less than 2 years old, you may continue to use the battery until the battery reaches 2 years old, at which point, repeat instructions in step 5. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.78101@gehealthcare.com. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. After the correction has been implemented, please be sure to destroy
Quantity in Commerce12,808 total devices
DistributionWorldwide - US Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = IYN
-
-