| | Class 2 Device Recall Da Vinci 5 |  |
| Date Initiated by Firm | April 24, 2026 |
| Date Posted | June 04, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2310-2026 |
| Recall Event ID |
98569 |
| 510(K)Number | K232610 |
| Product Classification |
System, surgical, computer controlled instrument - Product Code NAY
|
| Product | Brand Name: Da Vinci 5
Product Name: ASSY,DV5 TOWER,IS5000
Model/Catalog Number: 380746
Software Version: N/A
Product Description: Preface:
This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system.
Product Description:
The NIR handheld camera system is compatible for use with the da Vinci 5 system to aid in
surgery. The components of the NIR handheld camera system include a camera control unit
and a light source, NIR handheld camera, and additional components that together, connect
to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from
the NIR handheld camera system is displayed on the Tower monitor.
Component: Yes, the SOCK is the component and integrated into the dV5 system |
| Code Information |
Lot Code: Part Number: 380746
UDI: 00886874119662
FDA Medical Device Listing Number: D531324
(U.S.)
Part Number: 470653-01
Serial Numbers:
1008007
1008008
1008009
1008010
1008011
1008012
1008013
1008015
1008016
1008017
1008020
1008021
1008022
1008024
1008100
1008101
1008102
1008104
1008105
1008107
1008108
1008109
1008112
1008113
1008114
1008115
1008116
1008117
1008997
1008998
1009902
1010029
1010033
1010034
1010036
1010272
1010296
1010297
1010318
1010319
1010449
1010450
1010451
1010496
1010497
1010498
1010499
1010501
1010581
1010618
1010619
1010620
1010621
1010633
1010634
1010635
1010636
1010638
1010640
1010642
1010643
1010644
1010645
1010646
1011333
1011334
1011335
1011336
1011337
1011338
1011340
1011362
1011363
1011364
1011366
1011370
1011371
1011372
1011374
1011377
1011378
1011379
1011380
1011381
1011382
1011383
1011384
1011386
1011388
1011389
1011390
1011391
1011392
1011393
1011394
1011395
1011396
1012056
1012057
1012058
1012059
1012060
1012062
1012063
1012064
1012065
1012066
1012068
1012069
1012070
1012071
1012073
1012075
1012076
1012078
1012079
1012080
1012081
1012082
1012084
1012085
1012284
1012998
1012999
1013001
1013002
1013003
1013004
1013005
1013007
1013008
1013009
1013010
1013011
1013012
1013013
1013015
1013017
1013019
1013108
1013109
1013110
1013111
1013112
1013113
1013114
1013115
1013116
1013117
1013118
1013119
1013120
1013121
1013122
1013124
1013125
1013126
1013127
1013129
1013130
1013132
1013491
1013492
1013493
1013494
1013496
1013497
1014027
1014028
1014029
1014030
1014031
1014032
1014034
1014036
1014037
1014038
1014039
1014140
1014143
1014146
1014147
1014149
1014150
1014151
1014152
1014153
1014154
1014155
1014156
1014158
1014159
1014160
1014163
1014164
1014165
1014166
1014167
1014168
1014169
1014170
1014171
1014173
1014174
1014175
1014177
1014178
1014574
1014575
1014576
1014577
1014578
1014580
1014581
1014582
1014583
1014584
1014585
1014586
1014587
1014589
1014590
1014591
1014592
1014594
1014595
1014596
1014597
1014598
1014599
1014600
1014601
1014602
1014603
1014604
1014605
1014606
1014607
1014608
1014609
1014610
1014611
1014612
1014613
1014615
1014616
1014618
1014619
1014622
1014623
1014624
1014721
1014722
1014723
1014724
1014725
1014726
1014727
1014728
1014729
1014730
1014731
1015247
1015248
1015250
1015251
1015252
1015253
1015254
1015255
1015256
1015257
1015258
1015259
1015260
1015261
1015262
1015263
1015264
1015266
1015854
1015855
1016110
1016111
1016112
1016113
1016114
1016115
1016117
1016118
1016119
1016120
1016122
1016123
1016124
1016125
1016126
1016127
1016128
1016129
1016130
1016136
1016137
1016138
1016139
1016140
1016141
1016142
1016143
1016144
1016150
1016151
1016152
1016153
1016154
1016155
1016475
1016478
1016479
1016480
1016481
1016482
1016483
1016484
1016601
1016602
1016603
1016604
1016605
1016606
1016607
1016608
1016791
1017181
1017182
1017327
1017328
1017329
1017330
1017331
1017332
1017333
1017334
1017336
1017337
1017338
1017340
1017341
1017343
1017345
1017346
1017347
1017348
1017545
1017547
1017549
1017550
1017552
1017553
1017555
1017556
1017557
1017558
1017560
1017561
1017562
1017565
1017566
1017569
1017570
1017571
1017572
1017573
1017867
1017871
1017873
1017874
1017875
1017878
1017879
1017880
1017882
1017884
1017887
1017888
1017890
1017891
1017893
1017894
1017895
1017896
1017898
1017899
1017900
1017902
1017903
1017904
1018075
1018076
1018108
1018109
1018110
1018111
1018112
1018114
1018146
1018148
1018150
1018151
1018152
1018153
1018194
1018195
1018196
1018197
1018198
1018199
1018256
1025415
1025416
1025417
1025418
1025419
1025420
1025421
1025427
1025433
1025434
1025435
1025436
1025437
1025438
1025439
1025440
1025442
1025443
1025444
1025716
1025717
1025718
1025719
1025720
1025721
1025722
1025723
1025724
1025725
1025726
1025727
1025728
1025729
1025730
1025731
1025732
1025733
1025734
1026046
1026047
1026048
1026049
1026051
1026052
1026770
1026771
1026772
1026773
1026774
1026775
1026776
1026779
1026780
1026781
1026782
1026783
1026784
1026786
1026787
1026788
1027960
1027961
1027962
1027963
1027964
1027965
1027966
1027967
1028043
1028044
1028045
1028046
1028047
1028048
1028049
1028050
1028051
1028052
1028053
1028054
1028057
1028078
1028080
1030040
1030042
1030043
1030044
1030045
1030046
1030048
1030069
1030090
1030091
1030092
1030093
1030095
1030096
1030098
1030099
1030102
1030103
1030104
1030105
1030106
1030108
1031386
1031387
1031388
1031389
1031390
1031391
1031392
1031393
1031394
1031395
1031396
1031397
1031398
1031399
1031400
1031401
1031402
1031403
1031404
1031406
1033055
1033056
1033057
1033058
1033059
1033060
1033061
1033062
1033063
1033064
1033066
1033067
1033068
1033069
1033070
1033071
1033072
1033073
1033074
1033075
1033076
1033077
1033078
1033079
1033080
1033081
1033083
1033084
1033085
1033086
1033087
1033088
1033089
1033090
1034577
1034578
1034579
1034580
1034581
1034582
1034583
1034585
1034586
1034587
1034588
1034589
1034590
1034593
1034594
1034595
1034596
1036621
1036622
1036625
1036626
1036630
1036631
1036632
1036633
1036634
1036635
1036636
1036637
1036638
1036639
1036640
1036641
1036642
1036643
1036644
1036646
1036647
1036648
1036649
1036650
1036661
1036662
1036663
1036664
1036665
1037612
1037614
1037615
1037616
1037617
1037618
1037619
1037620
1037621
1037622
1037623
1037624
1037625
1037636
1037637
1037638
1037639
1037651
1037652
1037656
1037657
1037658
1037659
1037660
1037661
1037663
1037664
1039145
1039146
1039147
1040231
1040232
1040233
1040234
1040235
1040246
1040247
1040248
1040249
1040250
1040251
1040252
1040253
1040254
1040255
1040256
1040257
1040258
1040259
1041370
1041371
1041372
1041373
1041374
1041376
1041377
1041378
1041379
1041875
1041876
1041877
1041879
1041880
1041881
1041882
1041883
1041884
1041885
1042157
1042158
1042159
1042170
1042171
1042172
1042173
1042945
1042946
1042947
1042948
1043796
1043797
1043798
1043803
1043807
1043809
1043811
1043812
1043814
1043816
1043817
1043818
1043819
1043820
1043823
1043884
1043885
1043887
1043893
1043894
1043899
1043900
1043901
1043902
1043903
1043904
1043905
1045944
1045945
1045946
1045947
1045948
1045949
1045950
1045951
1045952
1045953
1045954
1045955
1045957
1045958
1045959
1045960
1045961
1045962
1045963
1045964
1045965
1045966
1045977
1045983
1045994
1045995
1045996
1045997
1046020
1046021
1046042
1046043
1046044
1046045
1046046
1046047
1046048
1046049
1046050
1046051
1046053
1047635
1047636
1047638
1047689
1047690
1047692
1047693
1047714
1047716
1047718
1047719
1047720
1047721
1047722
1047724
1047725
1047726
1047727
1047728
1047729
1047730
1047731
1047733
1047734
1047735
1049196
1049282
1049283
1049284
1049286
1049287
1049288
1049289
1049290
1049291
1049292
1049293
1049294
1049295
1049296
1049298
1049299
1049300
1049301
1049302
1049303
1049304
1132890
1132891
1132892
1132893
1132894
1132895
1132897
1132898
1132900
1132901
1132902
1132903
1132904
1132905
1132906
1132908
1135325
1135326
1135332
1135333
1135334
1135335
1135336
1135338
1135341
1135342
1135343
1135344
1135345
1135346
1137670
1137671
1137673
1137675
1137677
1137678
1137689
1137710
1137712
1137716
1137720
1137721
1137722
1137723
1137724
1137725
1137728
1137729
1137730
1137731
1137732
1137733
1137734
1137735
1137736
1137737
1137740
1137741
1137744
1137746
1140563
1140564
1140566
1140567
1140570
1140571
1140573
1140576
1140577
1140578
1140579
1140581
1140582
1140585
1140586
1140587
1140588
1140590
1140593
1140594
1140595
1140596
1140597
1140598
1140599
1140600
1140601
1140602
1145413
1145414
1145415
1145419
1145422
1145423
1145424
1145425
1145426
1145945
1145946
1145947
1145948
1145949
1145950
1145951
1145952
1145953
1145954
1145955
1145956
1145957
1145958
1145960
1145961
1145962
1145963
1148140
1148141
1148142
1148143
1148144
1148145
1148146
1148147
1148148
1148149
1148150
1148151
1148152
1148153
1148154
1148155
1148156
1148157
1148158
1148159
1148160
1148161
1148162
1148163
1148164
1148165
1148166
1148167
1148168
1148169
1149836
1149837
1149838
1149839
1149840
1149841
1149842
1149843
1149844
1149845
1149846
1149857
1149858
1149859
1149860
1149861
1149862
1149863
1149864
1149865
1149866
1149867
1149888
1149889
1149890
1149891
1149892
1149893
1149894
1149895
1152001
1152002
1152003
1152004
1152055
1152056
1152057
1152058
1152059
1152060
1152061
1152062
1152063
1152064
1152065
1152066
1152067
1152068
1152069
1152070
1152071
1152072
1152073
1152075
1152076
1152077
1152078
1153842
1153843
1153844
1153845
1153846
1153849
1153852
1153853
1153854
1153855
1153856
1153857
1153858
1153859
1153860
1153861
1153862
1153863
1153864
1153865
1153866
1153867
1153869
1156018
1156020
1156021
1156022
1156023
1157113
1157114
1157115
1157116
1157117
1157118
1157119
1157121
1157122
1157615
1157636
1157647
1157678
1157689
1157700
1157701
1157703
1157707
1157718
1157719
1157720
1157722
1157723
1157725
1157726
1157727
1157739
1157741
1157743
1157747
1157749
1157755
1186377
Part Number: 470653-03
Serial Numbers:
1186136
1186139
1157705
1157706
1157721
1161015
1161017
1161028
1161029
1161030
1161056
1161064
1161066
1161067
1161068
1161069
1161070
1161071
1161073
1161074
1161075
1161076
1161077
1161078
1161080
1161081
1161082
1161149
1161150
1164405
1164407
1164428
1164449
1164450
1164455
1164456
1164464
1164465
1164466
1164519
1164520
1164522
1164523
1164524
1164525
1164526
1164527
1164528
1164530
1164531
1164532
1164533
1164537
1164541
1164545
1164549
1164550
1166267
1166268
1166269
1166272
1166273
1166275
1166276
1166280
1166281
1166285
1166312
1166313
1166314
1166316
1166323
1166325
1166328
1166329
1166330
1166331
1166333
1166334
1166336
1168930
1168932
1168933
1168934
1168938
1168939
1168941
1168942
1168943
1168944
1168945
1168946
1168947
1168950
1168953
1168954
1168955
1168956
1168962
1168968
1171019
1171036
1171038
1171039
1171040
1171041
1171043
1171045
1171059
1171060
1171062
1171063
1171069
1171072
1171073
1176677
1176698
1176741
1176742
1176743
1176744
1176745
1176746
1176748
1176749
1176750
1176751
1176752
1178538
1178543
1178555
1178567
1179195
1179201
1179203
1179224
1179225
1179226
1179228
1179229
1179231
1179232
1179254
1179255
1179276
1179277
1179298
1179299
1179300
1179301
1179312
1179313
1179333
1179338
1179340
1179343
1181818
1181819
1181820
1181822
1181823
1181824
1181825
1181827
1181828
1181830
1181832
1181833
1181855
1181857
1181861
1184449
1186134
1186135
1186145
1186168
1186374
1186385
1187866
1188191
1188194
1188195
1188206
1188207
1188208
1188209
1188210
1188211
1188212
1188213
1188219
1188220
1188222
1190015
1190016
1190017
1190018
1190019
1190021
1190022
1190023
1190025
1190029
(O.U.S.)
Part Number: 470653-01
Serial Numbers:
1014033
1014035
1014142
1014157
1014161
1014162
1026777
1026778
1026785
1028055
1028081
1028084
1043799
1043800
1043801
1043821
1043895
1043898
1045998
1046052
1047715
1047717
1137738
1137745
1140562
1140569
1140584
1140589
1140591
1140592
Part Number: 470653-03
Serial Numbers:
1179196
1161059
1168931
1171068
1159334
1157750
1157748
1166282
1157753
1171021
1168936
1161053
1157758
1164454
1164467
1159327
1159259
1159329
1157702
1157757
1159331
1164548
1157759
1161060
1186384
1171066
1161079
1157756
1159330
1164457
1159328
1168966
1159335
1164406
1157751
1168969
1166266
1161054
1168951
1159326
1159332
1159324
1157120
1166283
|
| FEI Number |
3001675293
|
Recalling Firm/ Manufacturer |
Intuitive Surgical, Inc.
1266 Kifer Rd
Sunnyvale
CA
94086-5304
|
| For Additional Information Contact | Shahbaz Khan 1-408-5232443 |
Manufacturer Reason for Recall | Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope. |
FDA Determined Cause 2 | Device Design |
| Action | On 04/24/2026, the firm emailed an "Urgent Medical Device Correction" Letter to customers informing them that Intuitive has identified a condition in which the NIR modes may remain active unintentionally when attempting to turn off visible white light illumination during normal clinical workflow transitions, when switching from the handheld camera to the robotic endoscope and may result in thermal injury if the distal end of the light guide (detached) or assembled laparoscope contacts the patient, user, or sterile drapes while NIR mode is active.
Customer Actions required for Continued Use of dV5 Systems Equipped with the NIR Handheld Camera:
"Prior to each use, ensure that the NIR Handheld Camera Light Source indicators remain visible to operating room staff so that the laser status can be confirmed. Position the dV5 Tower Monitor and other equipment so the Light Source indicators are not obstructed from view.
"The Tower Monitor should be positioned to avoid obstructing the Light Source indicators to maintain clear visibility of the CCU during use. Monitor positioning may vary based on Operating Room configuration. Refer to the Near Infrared Handheld Camera System User Manual (PN 557076-02) for additional setup guidance.
"The indicators include:
- LASER ON status indicator
-Laser symbol on the LCD display
-Green-colored button backlights
"Prior to placing the device on the patient or sterile drapes, confirm that the Light Source is in Standby Mode, as indicated by white-colored button backlights and Standby on the LCD display.
"Press and hold the green up arrow button on the Camera Head for approximately 2.5 seconds to deactivate the light source.
"Be aware of the following Cautions stated in the Handheld Camera Light Guide IFU (Schoelly s reference: TPA801-460-01; Intuitive s reference: PN 557072-01):
For questions/support, Clinical Sales Rep or Intuitive Customer Service at the numbers listed below:
"North and South America: (800) 876-1310, Option 3 (4 am |
| Quantity in Commerce | 1357 systems |
| Distribution | U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
O.U.S.: Australia, Belgium, France, Germany, Italy, Japan, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, South Korea, Spain, Switzerland, and United Kingdom |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = NAY
|
|
|
|