• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX Back to Search Results
Catalog Number PMX110
Device Problems Bent (1059); Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/24/2017
Event Type  malfunction  
Manufacturer Narrative
Results: the power inlet on the back of the penumbra system aspiration pump max 110 (pump max) was damaged.Conclusions: evaluation of the returned device revealed that the pump max power inlet was loose.The pump max was plugged in and powered on and functioned without an issue.However, the power inlet on the back of the pump max was damaged and, therefore, the pump max was deemed nonfunctional.The reported issues of the power inlet being full of plastic, bent, and the power cord not being able to be plugged in could not be confirmed.Penumbra pump maxs are 100% functionally tested during incoming inspection by quality.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, the technician found that the bracket for the power inlet on the pump max was broken, full of plastic, and was bent; therefore the power cord was unable to be plugged in.The power inlet was cleaned out, and the procedure was continued using the pump max.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Please note that the following sections were incorrectly reported on the initial mfr report and are being corrected on this follow-up #01 mfr report: 1.Occupation.2.510(k).3.Product code.4.Common device name.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6812027
MDR Text Key83298752
Report Number3005168196-2017-01401
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,07/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberPMX110
Device Lot NumberF21355-26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2017
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received01/14/2005
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age38 YR
-
-