The abbott field service representative (fsr) was dispatched to troubleshoot the issue.Upon inspection of the analyzer, the fsr observed that cuvette wash nozzle #5 had been leaking.The fsr replaced the nozzle, and then initiated a cuvette wash.There was no evidence of leaking or dripping after the nozzle was replaced.The fsr then recalibrated magnesium and performed a precision run to verify instrument performance.No additional erratic magnesium results were reported on architect (b)(4).A service ticket review for (b)(4) revealed no additional erratic or discrepant results, nor did it identify any service or complaint issues that may have contributed to this issue.Quality metrics were reviewed.No systemic issues or adverse trends for the issue associated with this ticket.(erratic / discrepant results) were identified.The architect system operations manual and the architect service and support manual provide adequate information, troubleshooting, and maintenance concerning erratic/ discrepant results.Based on the available information, a deficiency of the system was not identified.There is no evidence to reasonably suggest a malfunction occurred for the cuvette wash nozzles.The issue was resolved by the field service representative through standard troubleshooting procedures.
|