The customer reported that they received an erroneous result for one patient sample tested for total prostate-specific antigen (tpsa).The sample initially resulted as 0.024 ng/ml and this value was reported outside of the laboratory.The patient complained about the result, so the sample was repeated.The sample was repeated on (b)(6) 2015, resulting as 9.58 ng/ml accompanied by a data flag.The patient was not adversely affected.The tpsa reagent lot number was 181675, with an expiration date of 01/30/2016.It was determined that liquid flow cleaning of the instrument was pending and the last time it was performed was on 09/25/2015.The laboratory staff was informed to follow the recommended timelines for maintenance of the instrument.The sample probe cleaning was also not being performed as recommended for part of daily maintenance.A specific root cause could not be determined.The issue is most likely related to a fibrin clot or gel particle.A general issue with the reagent or analyzer can be excluded.
|