The customer reported that they received an erroneous result for one patient sample tested for total prostate-specific antigen (tpsa).The sample initially resulted as 0.715 ng/ml and this value was reported outside of the laboratory.The physician asked for the sample to be repeated since the patient had a prostatectomy in (b)(6) 2014.The sample was repeated in the original tube, resulting as < 0.007 ng/ml.The sample was also repeated in a sample cup, resulting as <0.007 ng/ml.The repeat result was believed to be correct.The patient was not adversely affected.The tpsa reagent lot number was 18677101, with an expiration date of 07/31/2016.The field service representative was not able to duplicate the issue.He inspected the fluidics system of the analyzer and no issues were seen.He replaced the measuring cell as a precaution.He performed a system volume check, a high voltage adjustment to the photomultiplier tube, a blank cell calibration, and performance testing.All results passed.
|