The customer stated that they received an erroneous result for one patient sample tested for ferritin on the e602 analyzer.The sample initially resulted as 33.29 ug/l and this value was reported outside of the laboratory.The result was questioned by the clinician since it did not fit with previous results for the patient.The initial result was also said to not fit a subsequent result.The sample was repeated with an automatic dilution, resulting as 6177 ug/l.The patient was not adversely affected.The ferritin reagent lot number and expiration date were asked for, but not provided.The field service engineer replaced pinch valve tubings and a valve.He also performed preventive maintenance on the analyzer, including running a fishing line through the sipper flow path, reagent probe, and tubing.He performed a "system volume".Controls were run for all assays and these were ok.He found that a fuse was blown and he replaced this, but it blew again.He replaced a printed circuit board within the analyzer.
|