Discordant, falsely low creatinine_2 (crea_2) results were obtained on one patient upon repeat testing on an advia 1800 instrument.The sample was initially tested using a different sample on the same instrument and on an alternate advia 1800 instrument and the results were numerically (b)(6).The customer obtained two new samples from the patient and tested them on the same instrument, resulting falsely low both times.The customer then obtained another new sample from the patient and tested it on the same instrument and the result was numerically (b)(6).The discordant results were not reported to the physician(s).The customer sent a third redraw sample to another hospital, which was tested on an alternate platform and the result was higher.The repeat result from the alternate platform was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low crea_2 results.
|