• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616S
Device Problems Fire (1245); Sparking (2595); Mechanics Altered (2984); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the powermax hand control started to grind and was shaving slowly.The attached dyonics power ii control unit also beeped, shut off and turned back on and started grinding.The physician removed it from the patient and the shaver handpiece started shaving without being turned on.Sparks and flames were also reported coming out of the end of the shaver.No patient or operating room staff impact as a result.
 
Manufacturer Narrative
Device investigation narrative - the reported complaint has been confirmed.Unit gives ¿motor stall¿ error message intermittently.Motor is drawing high current and is excessively noisy.Unit requires non-warranty repair.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key5804469
MDR Text Key50626506
Report Number1643264-2016-00120
Device Sequence Number0
Product Code NBH
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2016
Initial Date FDA Received07/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/23/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-