The customer stated that they received an erroneous result for one patient sample tested for elecsys total psa immunoassay (tpsa) on an e602 analyzer.The sample initially resulted as 0.024 ng/ml and this value was reported outside of the laboratory to a physician.The sample was repeated on (b)(6) 2016, resulting as 8.01 ng/ml.The patient was not adversely affected.The tpsa reagent lot number was 12019800.The reagent expiration date was asked for, but not provided.
|
A specific root cause could not be determined based on the provided information.Additional information required for the investigation was not provided.Investigations determined that there was no issue with calibration before the date of the event and that controls were within range on the day of the event.Alarm data and performance testing did not indicate any specific issues.The root cause may be related to bubbles/foam on the sample or reagent surfaces, use of a non-specified tube type, non-use of recommended rack adapters for tubes, sample quality, or insufficient maintenance.
|