The customer received questionable elecsys total psa immunoassay results for four patient samples from (b)(6) 2016.Of the data provided, only the results for one patient sample that occurred on (b)(6) 2016 were discrepant.The patient had a history of total psa results at approximately 8 ng/ml.On (b)(6) 2016, a sample from the patient was tested and the result was 0.061 ng/ml (<0.1 ng/ml).The sample was repeated and the result was 0.080 ng/ml (<0.1ng/ml).The results were reported to the doctor.On (b)(6) 2016, the doctor questioned the results and the same sample was thawed and repeated with a result of 7.70 ng/ml with a data flag.This result was believed to be correct.On (b)(6) 2016, the patient gave a new specimen and the result for that sample was 9.24 ng/ml.The patient was not adversely affected.The reagent lot number was 13598600 with an expiration date of 05/31/2017.The field service representative visited the site and collected data for further investigation.
|
A specific root cause could not be identified.Possible root causes include foam or air bubbles on the sample surface, a missing tube adapter, foam or air bubbles on the reagent pack, foam or air bubbles on the system reagent, or old pinch tubes being used on the system.Based on the provided calibration and qc data, a general reagent or system issue was not indicated.
|