BECKMAN COULTER AU480 CLINICAL CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
Catalog Number B12183 |
Device Problem
Low Test Results (2458)
|
Patient Problem
Injury (2348)
|
Event Date 11/20/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
A beckman coulter field service engineer (fse) found severe buildup on the inside of the ise electrode block.The fse replaced the ise electrode block and verified instrument operation.The primary failure mode is unknown; appears to be an isolated event.The redraw and repeat result of the original sample considered correct by the customer were obtained prior to the replacement of the ise electrode block.Patient demographics (weight, date of birth, gender, age) were not provided by the customer.
|
|
Event Description
|
The customer reported obtaining a false low potassium (k) result for one (1) patient, involving the au480 clinical chemistry analyzer (model au481-10e).The false low k result was reported outside the laboratory.The patient, who was in the emergency room for an unknown condition received two 25 meq oral potassium bicarbonate (k-effervescent) tablets as a result of the low potassium result.The patient was redrawn after treatment and the sample result was within the laboratory's established reference range (3.5-5.5 mmol/l).In addition, the customer repeated the original sample that had generated the low k result and obtained a result also within reference range.The patient was not admitted into the hospital and was discharged from the emergency room.
|
|
Search Alerts/Recalls
|
|
|