The customer contacted the siemens customer care center (ccc).The customer stated that quality control (qc) was running out of range low for level 1 and 3 after the discordant results were obtained.Ccc performed service method testing (smt) and found that sample arm 3 (s3) and the integrated multi-sensor technology (imt) mixers failed as well as the s3 alignment.A siemens customer service engineer (cse) was dispatched to the customer site.After evaluating the instrument, the cse replaced the s3 mixer and belts.The cse repeated smt for s3 mixer, resulting within range.The cse ran qc, resulting within range.The cause of the discordant, falsely low mg results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely low magnesium (mg) results were obtained on patient samples on a dimension vista 1500 instrument.The discordant results of patient samples (b)(6) and (b)(6) were reported to the physician(s).Patient (b)(6) was administered 2 doses of mg based on the erroneous result.The samples were repeated on the same instrument, resulting higher.Corrected reports were reported for patients id (b)(6) and (b)(6).The repeated results for all remaining patients were reported to the physician(s).There are no reports of adverse health consequences due to the discordant, falsely low mg results.
|