SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP FT4; RADIOIMMUNOASSAY, FREE THYROXINE
|
Back to Search Results |
|
Model Number ADVIA CENTAUR XP |
Device Problems
Mechanical Problem (1384); Nonstandard Device (1420); Calibration Problem (2890)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
A siemens customer service engineer (cse) was dispatched to the customer's site.The cse performed an acid and base decontamination and check the relative light units, resulting in range.The cse ran calibration, resulting in range.The cse ran quality control (qc) which resulting out of range.The cse ran a four day precision study and two patient samples, resulting within range.Qc resulted within range and one precision result was out of range.The cse replaced the reagent probe 1 diluter syringe.The customer moved the instrument and qc shifted in range and matches the alternate advia centaur instrument.The cause of the discordant, falsely depressed free thyroxine (ft4) results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
|
Event Description
|
Discordant, falsely depressed free thyroxine (ft4) results were obtained on two patient samples on an advia centaur xp instrument.The initial results were sent out to the physician(s), which were not questioned.The same samples were repeated on an alternate advia centaur instrument, resulting higher.Corrected reports were issued to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely depressed free thyroxine results being reported out.
|
|
Manufacturer Narrative
|
The initial mdr 2432235-2016-00802 was filed on december 22, 2017.Additional information (01/05/2017): siemens has performed an internal investigation and has confirmed a negative bias for the advia centaur ft4 when used with calibrator a kit lots ending in 90 on the advia centaur, advia centaur xp and advia centaur xpt systems.In addition, siemens healthcare diagnostics confirmed the potential for calibration failures due to above limit calibrator rlu %cvs when using calibrator a kit lots ending in 90 with the ft4 assay.Siemens issued ufsn cc 17-04.A.Ous (january 2017) and umdr cc 17-04.A.Us (january 2017) informing the customer of ft4 assay negative bias observed with calibrator a kit lots ending in 90.Instructions on actions to be taken are provided in the customer communication.
|
|
Search Alerts/Recalls
|
|
|