During a surgical procedure, a ureteral laminating stent was plugged into light source and after 15 to 20 seconds after being plugged in, staff noted smoke coming out of the light source hole above where the stent was plugged into.Surgeon stated unable to see stent lighted and surgeon started to unplug it.After turning off, smoke came out of both holes where it was unplugged.Staff person reported visually saw bright red center at end of stent where it was plugged into light source.Biomed noted event as "unit smoked when user connected ureteral illuminating catheter set.Cook medical; lot #6998468; ref # (b)(4).Tech was able to duplicate the problem when he connected the catheter to the light source.Unit worked ok with other head set.Contacted manufacturer and reported the problem and a representative picked up the catheter for an investigation.In follow up to the investigation, cook reported that there was no indication that a design or process related failure mode contributed to the event a definitive root cause could not be established.
|