The customer contacted a siemens customer care center (ccc) specialist.The customer accepted remote connection from the ccc.A review of the quality control (qc) data showed that it was in range at the time of the event.The ccc auto-aligned reagent arm 5 and sample probe 3.The ccc ran a sodium hydroxide clean on reagent 5.The ccc homed all modules and reset the instrument.A siemens customer service engineer (cse) was dispatched to the customer's site.The cse ran a diagnostics for mixers and cleaned drains.The cse found sample probe 3 was not properly aligned.The cse realigned the probe.The cse ran alignments and the customer ran calibration, resulting in range.The cause of the discordant, falsely depressed magnesium result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely depressed magnesium result was obtained on a patient sample on a dimension vista 1500 instrument.The initial result was not reported out to the physician(s).The customer repeated a new sample on an alternate dimension vista instrument, resulting higher.The repeat result was released to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed magnesium result.
|