The customer contacted the siemens customer care center (ccc).Ccc reviewed the instrument data and found no errors during the time the sample was run.Ccc performed special methods testing on the instrument and found that reagent probe 5 (r5) bottom of cuvette (boc) needed adjustment.A siemens customer service engineer (cse) was dispatched to the customer site.After evaluating the instrument, the cse cleaned the reagent tray cover over cuvette ring.The cse replaced r5 vertical and horizontal belts.The cse realigned sample probe 3 mixer (s3), r5 probes and verified boc position.The cse checked the mixer motion, ran quick check, service methods and quality controls, resulting satisfactory except for level 3.During follow-up visits, the cse ran precision on patient samples, quick check and qc, resulting within range except for level 3 qc.A reagent loader failure prevented loading of fresh qc.The cse replaced the s3 timing drive belts, and internal ultra violet lamp, realigned the probes and verified that boc were aligned.The cse ran quick check and service methods for s3 mixer.The cse ran qc, resulting within range.A siemens headquarter support center (hsc) specialist reviewed the instrument data and concluded that the cause of the discordant, falsely elevated mg result is attributed to a faulty s3 mixer and misalignment of the s3 probe which was resolved by the replacement of s3 mixer, alignment of the s3 probe and other service performed.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely elevated magnesium (mg) result was obtained on a patient sample on a dimension vista 1500 instrument.The discordant result was flagged "above panic range" and was auto re-run, resulting lower.The sample was then repeated on an alternate dimension vista instrument, resulting lower and confirming the auto repeated result.The alternate dimension vista instrument result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated mg result.
|