A discordant, falsely elevated creatinine result was obtained on a patient sample on a dimension vista 1500 instrument.The initial result was reported out to the physician(s), which was questioned.The customer tested a new sample on the same dimension vista instrument, resulting lower.The customer repeated the new sample on an alternate dimension vista instrument, resulting lower than the initial result.The customer issued a corrected report to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated creatinine result.
|
The initial mdr 2517506-2017-00237 was filed on march 08, 2017.Additional information (04/13/2017): the siemens headquarters support center (hsc) reviewed the data supplied.Hsc found no indication of reagent delivery issues.No further issues were reported, related to this issue.The cause of the discordant, falsely elevated creatinine result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|