The customer contacted the siemens customer care center (ccc).The customer stated that quality control (qc) started running low since after the last service maintenance.A siemens customer service engineer (cse) was dispatched to the customer site.During troubleshooting the instrument, the cse performed mix test on sample arm and found poor precision.The cse replaced the sample mixer, reagent 3 (r3) probe and manifold.The cse reran mix test, and the mean was acceptable.The cse ran qc, resulting within specification.The cause for the discordant, falsely depressed (mg) result is due to a faulty r3 mixer.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely depressed magnesium (mg) result was obtained on a patient sample (sample id (b)(6) ) on a dimension vista 1500 instrument.The discordant result was reported to the physician(s).Two additional falsely depressed mg were obtained on the same dimension vista 1500 instrument.The discordant results were not reported to the physician(s).The samples were repeated on an alternate dimension vista instrument, resulting higher.The repeated results were reported to the physician(s) for all three patients.There are no reports of patient intervention or adverse health consequences due to the discordant, falsely depressed mg results.
|