The customer contacted a siemens customer care center (ccc) specialist.The ccc reviewed the data provided.The ccc found the customer's quality control (qc) was in range at the time of the event.The ccc reviewed the error log and found no issues.The ccc ran service method testing (smt), resulting out of range.A siemens customer service engineer (cse) was dispatched to the customer's site.The cse replaced the valve on the sample pump panel and sample probe.The cse aligned the sample probe.The cse ran air elimination, twice.The cse emptied the tank, primed the system and repeated smt.Smt was out of specifications.The cse replaced the pump assembly for sample probe 1 and ran a quick check, resulting within range.The cse ran all service methods for sample probe 1, resulting within range.The cse performed qc, resulting within range.The cause of the discordant, falsely depressed magnesium results is due to a malfunction of the pump assembly.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely depressed magnesium results were obtained on five patient samples on a dimension vista 500 instrument.The initial results were not reported out to the physician(s).The customer repeated the same sample on an alternate dimension vista instrument, resulting higher.The customer reported out the repeat results to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely depressed magnesium results.
|