The customer initially stated that electrodes and tubing were replaced by the field service engineer on 03/10/2017 as part of preventive maintenance for the cobas 6000 c (501) module - c501 analyzer.The customer stated that erroneous initial results were reported outside of the laboratory for 8 patient samples tested for ise indirect ci for gen.2 (chloride) on (b)(6) 2017.These samples were repeated on the same analyzer and the repeat results were believed to be correct.Refer to the attachment for all patient data.The patients were not adversely affected.The chloride electrode lot number and expiration date were asked for, but not provided.On (b)(6) 2017, the customer stated that they were having calibration issues with the ise tests on the c501 analyzer on (b)(6) 2017.The customer calibrated the analyzer since multiple patient samples were showing decreased anion gap results.The customer replaced the potassium chloride system reagent and re-calibrated.Calibrations looked better.Controls and patient results were testing well.The customer has a low throughput of testing on the analyzer.Internal standard and diluent system reagents had been on board the analyzer for a while and were getting low.The customer also stated that the humidity in their laboratory was low.The customer was advised that humidity outside of analyzer specifications can cause evaporation of system reagents.The customer was advised to replace the system reagent bottles, perform a prime, run an ise check 30 times, and then recalibrate.The customer stated that controls were drifting on (b)(6) 2017 and this was resolved by replacing the system reagents and re-calibrating.The customer declined a service visit.A specific root cause could not be determined based on the provided information.
|