The customer received questionable high ise indirect for gen.2 chloride results for multiple samples over two days.He changed the electrode, the calibrator solution, the reference electrode, and the mix tower but the results were still high.The customer sent samples to another laboratory for testing.Of the data provided for four patient samples, only the results for one patient were discrepant.The initial chloride result was 106 mmol/l and the repeat result was 96 mmol/l.The initial result was reported outside the laboratory.The repeat result was believed to be correct.The patient was not adversely affected.The electrode lot number and expiration date were requested but were not provided.The field service representative found a problem with the vacuum/ pressure and the mix /dry line from the pump was kinked and restricting flow.He replaced the tubing to the air pump, rerouted the lines, and wrapped them in place.He performed mix/dry adjustments to specification and the customer successfully ran calibration and controls with results near the mean values.
|