SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Temperature Problem (3022)
Patient Problems Burn(s) (1757); Partial thickness (Second Degree) Burn (2694)
Event Date 06/14/2017
Event Type  Injury  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that during the knee arthroscopy the surgeon reached for the device and immediately noticed it was hot.Since the device was on the patient the surgeon checked the patient and noticed a small burn.The surgeon also burned his hand while reaching for the device.The healthcare professional claimed the handle wasn't turned on while it wasn't in use.
 
Event Description
It was reported that during the knee arthroscopy the surgeon reached for the device and immediately noticed it was hot.Since the device was on the patient the surgeon checked the patient and noticed a small burn.The surgeon also burned his hand while reaching for the device.The healthcare professional claimed the handle wasn't turned on while it wasn't in use.The procedure was completed with a backup device.The patient was diagnosed with a level 2 burn and treated with granuflex.
 
Manufacturer Narrative
A functional evaluation was performed and presented a motor stall error and heating of the hand piece.Corrosion was observed on the cable assembly connection.The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion.The root cause of this event was identified to be associated with corrosion of the motor and gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w william cannon drive
austin, TX 78735
MDR Report Key6703135
MDR Text Key79678731
Report Number1643264-2017-00398
Device Sequence Number0
Product Code NBH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/06/2017
Initial Date FDA Received07/11/2017
Supplement Dates Manufacturer Received07/20/2017
08/03/2017
Supplement Dates FDA Received07/20/2017
08/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;