The product evaluation was performed by reviewing the complaint text, instrument service history, tracking and trending metrics, and the current labeling for the architect system operations manual.A review of the complaint text indicated troubleshooting was performed by laboratory staff, which included recalibration of the sample pipettor and replacement of a dirty and discolored cuvette dry tip.A cuvette wash was also performed and the calcium assay was recalibrated.The cuvette dry tip was considered to be the likely cause of the elevated calcium results.The trending data for the cuvette dry tip did not identify a product issue; furthermore, the erratic result rates were within expected limits for the architect c4000 system.A review of the product labeling concluded that the issue is sufficiently addressed.Taken together, no systemic issue or product deficiency was identified.
|
A product correction letter was issued to all architect clinical chemistry customers notifying them to adhere to the labeling instructions listed in the architect operations manual, section 10, to resolve error codes 3580 to 3585 or 5381 before transitioning the architect to the running status.The product correction letter provides guidance on how the customer can ensure the pressure monitoring system is functioning and also recommends following the architect systems operations manual and the assay-specific documentation to ensure all samples and reagents meet described requirements before being placed on the architect.The investigation found that when the pressure monitor board communication on the architect clinical chemistry analyzer is not successfully established during initial instrument boot up, and the following error codes 3580 to 3585 or 5381 are not resolved by the user prior to transitioning the architect to the running status, there is the potential to generate incorrect results as the pressure monitoring system is inactive.The architect system software will be updated in a future version to prevent the architect analyzer from transitioning to the running status when the pressure monitoring system is in an error state.
|