MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CHEMISTRY XPT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CHEMISTRY XPT Back to Search Results
Model Number ADVIA CHEMISTRY XPT
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2017
Event Type  malfunction  
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc) and stated that quality control (qc) was within range on the day the event occurred.A siemens customer service engineer (cse) was dispatched to the customer site.After inspecting the instrument, the cse replaced the dilution tray (dtt), reaction tray (rrv) cuvette and lamp.The cse performed cell blank and register lamp.The cse ran a shutdown wash.The cse ran qc, which passed.The cause of the discordant, falsely low lip result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
A discordant, falsely low lipase (lip) result was obtained on a patient sample on an advia chemistry xpt instrument.The discordant result was not reported to the physician(s).The sample was repeated on the same instrument, resulting higher.The repeated result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant falsely low lip result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CHEMISTRY XPT
Type of Device
ADVIA CHEMISTRY XPT
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
JEOL LTD
registration number: 300363768
3-1-2 musashino akishima
tokyo 196-8558,
JA  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown, NY 10591
9145242687
MDR Report Key6835559
MDR Text Key84362407
Report Number2432235-2017-00491
Device Sequence Number0
Product Code CHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CHEMISTRY XPT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/01/2017
Initial Date FDA Received08/31/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-