The customer questioned low direct bilirubin (bild2) and bilt3 bilirubin total gen.3 (bilt3) results for 1 patient sample tested on a cobas 6000 c (501) module.Based on the data provided, the total bilirubin results were erroneous.The erroneous results were not reported outside of the laboratory.The initial bilt3 result from a hitachi cup was 16.8 (unit of measure not provided).The repeat result was 235.There was no allegation that an adverse event occurred.The bilt3 reagent lot number and expiration date were not provided.The customer cleans the sample probe every day with alcohol.A specific root cause was not identified.The system has run with no further deviations since this one occurrence.Based on the reaction curve provided, sample was missing from the reaction.Bilt3 is sensitive to disturbances due to a partly clogged sample probe.A review of the customer¿s service history did not suggest any pre-analytical issues.A possible root cause may be that a prior sample was not centrifuged long enough or a sample that tends to clot after centrifugation caused a micro clot, but this cannot be confirmed.
|