• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C501
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/23/2017
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).Unique identifier (udi)#: (b)(4).
 
Event Description
The customer questioned low direct bilirubin (bild2) and bilt3 bilirubin total gen.3 (bilt3) results for 1 patient sample tested on a cobas 6000 c (501) module.Based on the data provided, the total bilirubin results were erroneous.The erroneous results were not reported outside of the laboratory.The initial bilt3 result from a hitachi cup was 16.8 (unit of measure not provided).The repeat result was 235.There was no allegation that an adverse event occurred.The bilt3 reagent lot number and expiration date were not provided.The customer cleans the sample probe every day with alcohol.A specific root cause was not identified.The system has run with no further deviations since this one occurrence.Based on the reaction curve provided, sample was missing from the reaction.Bilt3 is sensitive to disturbances due to a partly clogged sample probe.A review of the customer¿s service history did not suggest any pre-analytical issues.A possible root cause may be that a prior sample was not centrifuged long enough or a sample that tends to clot after centrifugation caused a micro clot, but this cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 6000 C (501) MODULE
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6864147
MDR Text Key87231003
Report Number1823260-2017-01973
Device Sequence Number0
Product Code MQM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K060373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC501
Device Catalogue Number04745914001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/23/2017
Initial Date FDA Received09/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-