The customer stated that they had been having intermittent issues with low values for an unspecified number of patient samples tested for ise indirect k for gen.2 (potassium) on a cobas 6000 c (501) module - c501.The customer provided data for two patient samples that had questionable initial potassium results.Of these, one had an erroneous potassium result that was reported outside of the laboratory.The sample was repeated and the repeat result was believed to be correct.The sample initially resulted as 3.3 mmol/l and repeated as 4.0 mmol/l.The patient was not adversely affected.The potassium electrode lot number and expiration date were asked for, but not provided.The field service engineer determined there was a fluidic failure of the analyzer as the vacuum pressure was too low.The vacuum line at the vacuum vessel was not attached.He replaced and re-attached the vacuum line, replaced vacuum diaphragms.He replaced the ise sipper, sipper cover, and spring.He verified the fluidics dispense and probe alignment.A mechanism check was performed and this passed.The customer ran calibration and quality controls.Controls were near mean values.The instrument was determined to be performing within specifications.
|