• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MANUFACTURING INC ARCHITECT C16000 SYSTEM; AUTOMATED CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MANUFACTURING INC ARCHITECT C16000 SYSTEM; AUTOMATED CHEMISTRY ANALYZER Back to Search Results
Catalog Number 03L77-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2017
Event Type  malfunction  
Manufacturer Narrative
No patient returns were available for investigation.Field service was dispatched to the account and replaced the reagent probe and the peristaltic head tubing to resolve the issue.Customer complaint data was reviewed and no adverse trends or systemic issues were identified.The architect system operations manual was reviewed and was found to adequately address the issue.Based on the available information no product deficiency of the architect c16000 system was identified.
 
Event Description
The customer stated that an elevated cc magnesium result of > 3.90 mmol/l was generated.The sample was repeated and a result of 0.52 mmol/l was generated.The sample was also repeated on a second analyzer and a result of 0.52 mmol/l was generated.The elevated result was not reported.There was no report of impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT C16000 SYSTEM
Type of Device
AUTOMATED CHEMISTRY ANALYZER
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6946970
MDR Text Key90312189
Report Number1628664-2017-00377
Device Sequence Number0
Product Code JGJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Type of Report Initial
Report Date 10/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03L77-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/19/2017
Initial Date FDA Received10/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-