The customer stated that they received an erroneous result for one patient sample tested for the elecsys anti-tshr immunoassay (anti-tshr) on a cobas 6000 e 601 module (e601).It was asked, but it is not known if an erroneous result was reported outside of the laboratory.The sample initially resulted as > 40 iu/l.The sample was repeated twice, resulting as 0.4 iu/l and 0.8 iu/l.No adverse events were alleged to have occurred with the patient.The anti-tshr reagent lot number and expiration date were asked for, but not provided.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.Based on the provided information, a general reagent issue is not likely.The most likely root cause is related to possible foam/bubbles on the surface of the reagent.Bubbles/foam on the reagent surface could lead to a pipetting of an insufficient amount of reagent.Another possible root cause is related to sample quality.
|