MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. ALERE HCG CASSETTE; RADIOIMMUNOASSAY, HUMAN CHORIONIC GONADOTROPIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. ALERE HCG CASSETTE; RADIOIMMUNOASSAY, HUMAN CHORIONIC GONADOTROPIN Back to Search Results
Lot Number HCG7040129;HCG7070005
Device Problems False Negative Result (1225); False Positive Result (1227)
Patient Problem Abdominal Pain (1685)
Event Date 11/09/2017
Event Type  malfunction  
Event Description
Several incidents reported of inaccurate point-of-collection testing (poct) urine human chorionic gonadotropin (hcg) tests-false positives and negatives with poc pregnancy testing.An ed patient who had a false negative poct urine test approximately one week ago, underwent a ct scan of her abdomen with contrast and 2 ultrasound tests.Based on the ct scan/us findings, an hcg serum quantitative blood test was performed and results came back at positive level.The products were pulled off the shelf post event from all our facilities.Excerpt from emr ".Female.Presents to the emergency department with complaint of severe right-sided abdominal pain.The patient's original testing came back with a negative pregnancy test.Ultrasound showed what looks like possible corpus luteal cyst in the right side and possible iup* but would be very early if true.The patient had negative pregnancy test therefore ct scan was obtained to evaluate for an appendicitis given her continued pain.Ct scan was unremarkable.The patient was discussed with the obgyn physician who recommended a serum hcg because of the recent event of a negative hcg.They came out positive on blood.Blood test was ordered which came back with a positive pregnancy test with an hcg of 1312" * intrauterine pregnancy.
 
Event Description
Several incidents reported of inaccurate point-of-collection testing (poct) urine human chorionic gonadotropin (hcg) tests-false positives and negatives with poc pregnancy testing.An ed patient who had a false negative poct urine test approximately one week ago, underwent a ct scan of her abdomen with contrast and 2 ultrasound tests.Based on the ct scan/us findings, an hcg serum quantitative blood test was performed and results came back at positive level.The products were pulled off the shelf post event from all our facilities.Excerpt from emr ".Female.Presents to the emergency department with complaint of severe right-sided abdominal pain.The patient's original testing came back with a negative pregnancy test.Ultrasound showed what looks like possible corpus luteal cyst in the right side and possible iup* but would be very early if true.The patient had negative pregnancy test therefore ct scan was obtained to evaluate for an appendicitis given her continued pain.Ct scan was unremarkable.The patient was discussed with the obgyn physician who recommended a serum hcg because of the recent event of a negative hcg.They came out positive on blood.Blood test was ordered which came back with a positive pregnancy test with an hcg of 1312" * intrauterine pregnancy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE HCG CASSETTE
Type of Device
RADIOIMMUNOASSAY, HUMAN CHORIONIC GONADOTROPIN
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
MDR Report Key7057361
MDR Text Key92873286
Report Number7057361
Device Sequence Number0
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Lot NumberHCG7040129;HCG7070005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/15/2017
Event Location Hospital
Date Report to Manufacturer11/15/2017
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/24/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/12/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age33 YR
Patient Weight60
-
-