• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT DOWNLOADER RECHARGER; DOWNLOADER/RECHARGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT DOWNLOADER RECHARGER; DOWNLOADER/RECHARGER Back to Search Results
Catalog Number 04P73-04
Device Problems Device Inoperable (1663); Electrical Shorting (2926)
Patient Problem No Patient Involvement (2645)
Event Date 04/06/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2018, abbott point of care (apoc) was contacted by a customer reporting that downloader sn (b)(4) was shortening out and smoking.The customer confirmed they are using the correct cable that came with the downloader.The product was replaced at no charge and returning for investigation along with its cable and power supply.There are no injuries associated with this event.Per i-stat1 system manual: art: (b)(4), rev.Date: (b)(6) 2012: the downloader/recharger can recharge a rechargeable battery in the analyzer.If the analyzer contains a rechargeable battery, the battery begins recharging automatically as soon as the analyzer is placed in the downloader/recharger.The downloader/recharger also has a compartment for recharging a rechargeable battery outside the analyzer.Placing an analyzer in a downloader/recharger will automatically initiate recharging of the rechargeable battery.The indicator light on top of the downloader/recharger will be green (trickle charge), red (fast charge), or blinking red (fast charge pending) when an analyzer with a rechargeable battery is placed in the downloader/recharger.As well, placing a rechargeable battery into the recharging compartment will automatically initiate trickle recharging.The indicator light near the recharging compartment will be green when a rechargeable battery is placed in the compartment.
 
Event Description
Na.
 
Manufacturer Narrative
Apoc incident # (b)(4).The investigation was completed on 06/01/2018.The customer reported that downloader recharger (drc) s/n (b)(4) has shorted out and emitted smoke.A waxy light-yellow substance leaked out from the vents under the drc.The customer confirmed that they are using the correct cable that came with the drc.While the customer complaint was not confirmed in that no smoke was emitted during failure analysis, a defective component u12 was discovered which would have caused a low analyzer voltage output on the drc and potentially caused the drc to emit smoke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT DOWNLOADER RECHARGER
Type of Device
DOWNLOADER/RECHARGER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7425955
MDR Text Key105432783
Report Number2245578-2018-00132
Device Sequence Number0
Product Code JQP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/11/2018
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04P73-04
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2018
Initial Date FDA Received04/12/2018
Supplement Dates Manufacturer Received06/01/2018
Supplement Dates FDA Received06/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-