The customer stated that lymphocyte and platelet results for two patient samples measured on a sysmex xt 1800 analyzer did not appear correctly when observed in the cobas infinity core middleware system.For each sample, the first screen shot is from the sysmex xt 1800 analyzer and the second screen shot is from the cobas infinity system.For the first sample, the lymphocyte result from the sysmex xt 1800 was 286, however the result within the cobas infinity system was 1286.For the second sample, the platelet result from the sysmex xt 1800 was 187, however the result within the cobas infinity system was 2187.No adverse events were alleged to have occurred with the patients.No erroneous values were reported outside of the laboratory.
|