THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, OUS Back to Search Results
Catalog Number 201-10002
Device Problems Mechanical Problem (1384); Nonstandard Device (1420)
Patient Problem No Patient Involvement (2645)
Event Date 10/22/2014
Event Type  malfunction  
Manufacturer Narrative
No patient involvement.The device returned for analysis(if this is true, if not indicate that the device did not return).The complaint investigation determined the reported difficulty was the result of a design related wear issue.After review of this event and similar incidents, abbott has decided to initiate a voluntary field action for centrimag.Abbott performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.
 
Event Description
It was reported that a centrimag motor was connected to a console, the error message "motor disconnected" appeared and the motor would not start.When the problem first occurred, the system was being prepared for use.No patient was involved in this case.The customer sent the motor and pump to the hospital engineers for an investigation.This event was previously not reported due no patient involvement.Abbott made the decision on july 13, 2018 to initiate a voluntary recall, therefore this event is being upgraded to reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, OUS
Type of Device
MOTOR
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key7775289
MDR Text Key117016859
Report Number2916596-2018-03344
Device Sequence Number0
Product Code DWA
Reporter Country CodeSW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 08/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201-10002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2014
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/23/2014
Initial Date FDA Received08/11/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-