Model Number G136
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Vibration (1674)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g136 scaler, there was intermittent water and vibration and the handpiece was getting hot; no injury resulted.
 
Manufacturer Narrative
The handpiece cable is badly clogged causing no water flow.Debris build up in water hose and solenoid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON PLUS ULTRASONIC SCALER WITH TAP-ON TECHNOLOGY
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key8035606
MDR Text Key127561652
Report Number2424472-2018-00169
Device Sequence Number0
Product Code ELC
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG136
Device Catalogue NumberG136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2018
Initial Date Manufacturer Received 10/03/2018
Initial Date FDA Received11/02/2018
Supplement Dates Manufacturer Received11/05/2018
Supplement Dates FDA Received11/29/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-