MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problems Material Protrusion/Extrusion (2979); Material Split, Cut or Torn (4008); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2019
Event Type  malfunction  
Manufacturer Narrative
A sample product was not returned for analysis.Product history records were reviewed and documentation indicates the product met release criteria.The iol product history record could not be reviewed because facility did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.There are 6 other complaints in the lot.Additional information was requested.(b)(4).
 
Event Description
A technician reported that during an intraocular lens (iol) implant procedure, the tip of the cartridge split as the lens was being ejected out.The surgeon was able to pull the cartridge out of the eye with the lens still partially sticking out.The iol was removed without any damage.The lens was reloaded into another cartridge and implanted successfully.Additional information was requested.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The cartridge was returned inside the opened pouch, that had been taped closed.Viscoelastic was observed dried in the cartridge.The cartridge nozzle was stressed.An aneurysm formed, after the parting line and extended into torn tip damage along the right side of the tip.The tip has stress.The cartridge has evidence of being placed into a handpiece.An iol product identification label had been affixed to the clear side of the pouch.This product was identified as the associated lens.Event date, patient initials, birthdate, doctor initial, and cartridge info was also handwritten on the pouch.The cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and documentation, indicates the product met release criteria.The customer indicated, the use of a qualified diopter lens, handpiece and viscoelastic.The product investigation could not identify a root cause for the reported complaint.Qualified associated products were used.The extreme damage to the tip was observed.This type of damage typically occurs, if the lens is not positioned correctly for advancement.If there is a lack of viscoelastic between the lens and the cartridge lumen.If the lens is advanced too rapidly or if the handpiece plunger is not positioned correctly at the trailing optic edge.If the handpiece plunger is not positioned at the trailing optic edge, it can allow the lens to fold around the plunger tip, making it too large to correctly advance through the narrow tip of the cartridge.Which could cause damage to the tip or the lens.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key8438218
MDR Text Key139536610
Report Number1119421-2019-00341
Device Sequence Number0
Product Code MSS
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 01/25/2021
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number8065977763
Device Lot Number32674268
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/27/2019
Initial Date FDA Received03/20/2019
Supplement Dates Manufacturer Received03/28/2019
01/22/2021
Supplement Dates FDA Received04/02/2019
01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MONARCH III HANDPIECE; SA60WF; VISCOAT
Patient Age81 YR
-
-