• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT BNP CONTROLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT BNP CONTROLS Back to Search Results
Catalog Number 08K28-11
Device Problems Nonstandard Device (1420); High Test Results (2457)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2019
Event Type  malfunction  
Manufacturer Narrative
Correction/removal number: 3002809144-06/4/19-006-r a product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
 
Event Description
The customer observed abbott controls shifted high while using architect bnp control lot 44k82718 and calibrator lot 44k82618.There was no report of incorrect patient results or impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT BNP CONTROLS
Type of Device
BNP CONTROLS
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key8741123
MDR Text Key151341566
Report Number1415939-2019-00107
Device Sequence Number0
Product Code JJX
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2019
Device Catalogue Number08K28-11
Device Lot Number44K82718
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/07/2019
Initial Date FDA Received06/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2018
Is the Device Single Use? No
Type of Device Usage Unknown
Removal/Correction Number3002809144-06/4/19-006R
Patient Sequence Number1
-
-