• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 UNKNOWN HIP FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problems Nonstandard Device (1420); Naturally Worn (2988)
Patient Problems Cyst(s) (1800); Foreign Body Reaction (1868); Pain (1994); Pocket Erosion (2013); Scarring (2061); Joint Disorder (2373); Osteolysis (2377); Test Result (2695); No Code Available (3191)
Event Date 12/19/2017
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr medical records received.After review of medical record, the patient was revised to address failed right total hip replacement involving articular surface wear, adverse local tissue response, and periprosthetic osteolysis, aseptic loosening of the acetabular component, acetabular erosion and pseudocyst of the thigh.Revision notes reported large osteolytic cyst.The posterior aspect of the thigh appeared to be filled with murky fluid which was stained black.Scar formation was noted.Patient's hip was quite stiff.There was notching of the anterior neck of the femoral component by approximately 2-3 mm, suggesting it was impinging on the anterior aspect of his metal on metal component.There was small osteolytic lesion in zone 7.Acetabular component was grossly loose and was clearly mobile.Clinical visits reported pain, elevated metal ions, and cystic mass of 10.5 cm.Doi: (b)(6) 2009; dor: (b)(6) 2017; (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> this hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key8804789
MDR Text Key151507049
Report Number1818910-2019-98730
Device Sequence Number0
Product Code JDI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/27/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/27/2019
Initial Date FDA Received07/18/2019
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received08/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
-
-