MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/08/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, an electrode from the mapping catheter tip was missing.The mapping catheter was replaced to resolve the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the mapping catheter, 990063-020 with lot number 219068908, was returned and analyzed.Visual inspection of the mapping catheter showed that the pebax tubing was ribbed at around 2.75 inches from the loop tip.All the eight electrodes were in their place and no issue was observed on the mapping catheter loop.In conclusion, the reported issue of electrode detachment was not confirmed.The mapping catheter did not fail the returned product inspection due to missing electrodes.If information is provided in the future, a supplemental report will be issued.
|
|