A customer reported a lower than expected vitros glucose (glu) result obtained from a vitros performance verifier (pv) processed using vitros chemistry products glu slides in combination with a vitros 350 chemistry system.Vitros pv ii lot j6978 vitros glu result of <20 mg/dl versus an expected result of 288.9 mg/dl.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The lower than expected vitros glu result was from a quality control fluid.However, the investigation cannot conclude that patient samples would not be affected if the event were to recur undetected.There was no allegation of patient harm as a result of this event.This report corresponds to ortho clinical diagnostics inc.Complaint number (b)(4).
|