MEDTRONIC CRYOCATH LP ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Death (1802)
|
Event Date 01/01/2020 |
Event Type
Death
|
Manufacturer Narrative
|
Medtronic was made aware of this event through a search of literature publications.It was not possible to ascertain specific device information from the literature publication or to match the event with previously reported events.This information is based entirely on journal literature.This event occurred outside the us.Patient information is limited due to confidentiality concerns.The overall baseline gender characteristics is male; the age of the patients was approximately 60 years old.The exact cause and date of death is not available at the time of this report; as there is no indication of specific serial number for patient information.Of note, multiple patients were noted in the article; however, a one to one correlation could not be made with unique product lot/serial numbers.The model listed in the report is a representative of the model family, as there is no specific model listed.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: ¿comparison of outcomes using the first and second generation cryoballoon to treat atrial fibrillation.¿ heart lung circ.2020 mar;29(3):452-459.Doi: 10.1016/j.Hlc.2019.03.017.Epub 2019 apr 9.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The literature publication reported the following patient complication during the use of a mapping catheter.There was one patient who died of an unknown cause.Of note, multiple patients were noted in the article; however, a one to one correlation could not be made with unique product lot/serial numbers.The status/location of the mapping catheter is unknown.Further follow up did not yet yield any additional information.The cause of death has been requested, but not yet received.
|
|
Event Description
|
Additional information was obtained through follow up with the author/physician who indicated that the adverse events were related to the cryoballoon.The perforation of the "left atrial is a known complication of left-sided structural procedures." no further information was provided.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|