This patient was on cath lab table, and as we were about to start the procedure, x-ray didn't work.A soft and a hard reboot was done without any improvement.Phillips will be called for notification.Patient was transferred to another lab to perform the case.Replaced the dcps power supply in the b-cabinet - rechecked everything and system was stable.Philips service report: flexvision monitor went dark.The fd system, the x-ray generator, and all of the monitors in the control room remained operational.What actions were taken to repair the equipment? replaced the dcps power supply in the b-cabinet.The customer was advised that if the problem happens again, to please note the color of the led in the bottom right corner of the flexvision to help with diagnostics.Tested the system and the system checked ok.
|