• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE, INC. FISHER PAYKEL HEATER; HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE, INC. FISHER PAYKEL HEATER; HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) Back to Search Results
Model Number 82552
Device Problems Device Emits Odor (1425); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2019
Event Type  malfunction  
Event Description
Strong electrical burning smell noticed coming from equipment on standby, plugged in.Smoke noticed coming from equipment.Equipment was unplugged and biomed notified.Equipment labeled and picked up by biomed.Biomed sent equipment to bmes for evaluation/repair.Main cable, heater plate and wire adapter were replaced - calibrated and returned to service.Equipment was not in use by patient at the time.No harm to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER PAYKEL HEATER
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE, INC.
173 technology drive
suite 100
irvine CA 92618
MDR Report Key10009088
MDR Text Key189323333
Report Number10009088
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Type of Report Initial
Report Date 04/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number82552
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/21/2020
Event Location Hospital
Date Report to Manufacturer04/29/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/29/2020
Type of Device Usage N
Patient Sequence Number1
-
-