Section g3: correction.Section d10, h3, h4: additional information.Manufacturer's investigation conclusion: the reported event of m4: motor alarms was confirmed; however, it was revealed that the centrimag console was unrelated to the cause of the event.The returned centrimag console was functionally tested on 28may2020 and was found to perform as intended.The serviced and tested console was returned to the customer site after passing all tests per procedure.It was revealed that the root cause of the reported event was related to the returned centrimag motor.The motor¿s evaluation along with the log file associated with this event have been addressed.Review of the device history record for centrimag motor showed the device was manufactured in accordance with manufacturing and qa specifications.No further information was provided.The manufacturer is closing the file on this event.
|