• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO FOOTSWITCH; UNIT, OPERATIVE DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO FOOTSWITCH; UNIT, OPERATIVE DENTAL Back to Search Results
Model Number 5100008000
Device Problem Unintended Power Up (1162)
Patient Problem No Patient Involvement (2645)
Event Date 12/31/2019
Event Type  malfunction  
Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 previously reported event was included under mfr report # 0001811755-2020-00227 but should be included under this report.Product return status: 1 device was received.Event confirmation status: 1 reported event was not confirmed.Evaluation results: 1 device was found to be affected by a damaged cable.
 
Event Description
This report summarizes 1 malfunction event in which the device ran without user activation.1 event had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTSWITCH
Type of Device
UNIT, OPERATIVE DENTAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zach baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key10011402
MDR Text Key189591062
Report Number0001811755-2020-01239
Device Sequence Number1
Product Code EIA
UDI-Device Identifier04546540971272
UDI-Public04546540971272
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number5100008000
Device Catalogue Number5100008000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/31/2019
Initial Date FDA Received04/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-